Fluke Calibration 7109A Portable Calibration Bath (-25 to 145 C)

SKU
7109A
Manufacturer:
Fluke Calibration
Fluke Calibration

Key Features:

  • Calibrate up to four tri-clamp sanitary sensors at the same time
  • Wide temperature ranges cover most process applications:
    - 7109A: –25 °C to 140 °C
  • Excellent display accuracy of ±0.1 °C provides 4:1 test uncertainty ratio (TUR) for critical applications
  • Easy to transport up stairs and across catwalks
  • Stainless steel casing withstands harsh sterilizing chemicals and is rust proof
  • Easy to use and maintain
Product Options:
$12,911.00

Secured and trusted checkout with

Fluke Calibration’s 7109A Portable Calibration Bath has all the capabilities of the 6109A, but is crafted with for a -25°C to 140°C range.

Click here to see the Fluke Calibration 6109A Portable Calibration Bath with a 35°C to 250°C range.

General specifications
SizeHeight: 382 mm (15 in)
Width: 242 mm (9.5 in)
Depth: 400 mm (15.7 in)
Weight6109A: 16 kg (35 lb)
7109A: 20 kg (44 lb)
Fluid volume2.5 liters (0.66 gallons)
Fluid working area75 mm x 75 mm (3 in x 3 in)
Maximum fluid depth154 mm (6.1 in)
Factory calibrationTraceable NVLAP-accredited calibration included
Temperature specifications
The temperature specifications describe the Absolute Instrumental Uncertainty at 95 % level of confidence (coverage factor k = 2) within one year from calibration. The specifications include environmental temperature effects from 13 °C to 33 °C.
 6109A7109A
Range (at 25 °C ambient temperature)35 °C to 250 °C–25 °C to 140 °C (–15 °C cover off)
Accuracy (maximum permissible error)0.1 °C0.1 °C
Display resolution0.01 °, 0.001°0.01 °, 0.001°
Stability0.015 °C0.01 °C
Typical uniformity0.03 °C up to 200 °C
0.04 °C above 200 °C
0.02 °C
Repeatability0.04 °C0.04 °C
Typical heating time35 °C to 100 °C: 25 minutes
100 °C to 250 °C: 45 minutes
–25 °C to 25 °C: 35 minutes
25 °C to 140 °C: 55 minutes
Typical cooling time250 °C to 100 °C: 35 minutes
100 °C to 35 °C: 55 minutes
140 °C to 25 °C: 45 minutes
25 °C to –25 °C: 75 minutes
Typical settling time15 minutes10 minutes
Process Input Module specifications (-P models)
The Input Module specifications describe the Absolute Instrumental Uncertainty at 95 % level of confidence (coverage factor k = 2) within one year from calibration. The specifications include calibration uncertainty, linearity, repeatability, resolution, stability, and environmental temperature effects from 13 °C to 33 °C.
FunctionRangeAccuracy (Maximum Permissible Error)
Reference input resistance0 Ω to 42 Ω
42 Ω to 400 Ω
0.0025 Ω
0.006 %
Reference input temperature-25 °C to 250 °C0.007 % + 0.015 °C
Resistance sensing current1 mA8 %
DUT 4-wire resistance0 Ω to 31 Ω
31 Ω to 400 Ω
0.0025 Ω
0.008%
DUT 3-wire resistance accuracy0 Ω to 400 Ω0.12 Ω
Thermocouple mV-10 mV to 100 mV0.025 % + 0.01 mV
Reference junction temperature0 °C to 40 °C0.35 °C
Thermocouple temperature-25 °C to 250 °CJ: 0.44 °C
E: 0.44 °C
L: 0.42 °C
K: 0.49 °C
N: 0.57 °C
U: 0.48 °C
T: 0.53 °C
M: 0.48 °C
0 °C to 250 °CR: 1.92 °C
D: 1.12 °C
S: 1.88 °C
G/W: 3.34 °C
C: 0.84 °C
mA range0 mA to 22 mA0.02 % + 0.002 mA
Loop power24 V dc± 6 V
Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account

Products In This Series

As low as

Now $10,793.00