We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
Fluke Calibration’s 7109A Portable Calibration Bath has all the capabilities of the 6109A, but is crafted with for a -25°C to 140°C range.
Click here to see the Fluke Calibration 6109A Portable Calibration Bath with a 35°C to 250°C range.
General specifications | |
Size | Height: 382 mm (15 in) Width: 242 mm (9.5 in) Depth: 400 mm (15.7 in) |
Weight | 6109A: 16 kg (35 lb) 7109A: 20 kg (44 lb) |
Fluid volume | 2.5 liters (0.66 gallons) |
Fluid working area | 75 mm x 75 mm (3 in x 3 in) |
Maximum fluid depth | 154 mm (6.1 in) |
Factory calibration | Traceable NVLAP-accredited calibration included |
Temperature specifications The temperature specifications describe the Absolute Instrumental Uncertainty at 95 % level of confidence (coverage factor k = 2) within one year from calibration. The specifications include environmental temperature effects from 13 °C to 33 °C. | ||
6109A | 7109A | |
Range (at 25 °C ambient temperature) | 35 °C to 250 °C | –25 °C to 140 °C (–15 °C cover off) |
Accuracy (maximum permissible error) | 0.1 °C | 0.1 °C |
Display resolution | 0.01 °, 0.001° | 0.01 °, 0.001° |
Stability | 0.015 °C | 0.01 °C |
Typical uniformity | 0.03 °C up to 200 °C 0.04 °C above 200 °C | 0.02 °C |
Repeatability | 0.04 °C | 0.04 °C |
Typical heating time | 35 °C to 100 °C: 25 minutes 100 °C to 250 °C: 45 minutes | –25 °C to 25 °C: 35 minutes 25 °C to 140 °C: 55 minutes |
Typical cooling time | 250 °C to 100 °C: 35 minutes 100 °C to 35 °C: 55 minutes | 140 °C to 25 °C: 45 minutes 25 °C to –25 °C: 75 minutes |
Typical settling time | 15 minutes | 10 minutes |
Process Input Module specifications (-P models) The Input Module specifications describe the Absolute Instrumental Uncertainty at 95 % level of confidence (coverage factor k = 2) within one year from calibration. The specifications include calibration uncertainty, linearity, repeatability, resolution, stability, and environmental temperature effects from 13 °C to 33 °C. | ||||
Function | Range | Accuracy (Maximum Permissible Error) | ||
Reference input resistance | 0 Ω to 42 Ω 42 Ω to 400 Ω | 0.0025 Ω 0.006 % | ||
Reference input temperature | -25 °C to 250 °C | 0.007 % + 0.015 °C | ||
Resistance sensing current | 1 mA | 8 % | ||
DUT 4-wire resistance | 0 Ω to 31 Ω 31 Ω to 400 Ω | 0.0025 Ω 0.008% | ||
DUT 3-wire resistance accuracy | 0 Ω to 400 Ω | 0.12 Ω | ||
Thermocouple mV | -10 mV to 100 mV | 0.025 % + 0.01 mV | ||
Reference junction temperature | 0 °C to 40 °C | 0.35 °C | ||
Thermocouple temperature | -25 °C to 250 °C | J: 0.44 °C E: 0.44 °C L: 0.42 °C | K: 0.49 °C N: 0.57 °C U: 0.48 °C | T: 0.53 °C M: 0.48 °C |
0 °C to 250 °C | R: 1.92 °C D: 1.12 °C | S: 1.88 °C G/W: 3.34 °C | C: 0.84 °C | |
mA range | 0 mA to 22 mA | 0.02 % + 0.002 mA | ||
Loop power | 24 V dc | ± 6 V |
Write Your Own Review
Products In This Series
As low as
Now $12,945.00As low as
Now $15,061.00