We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
In addition to all the features and functions of the 6109A, Fluke Calibration’s 6109A-P Portable Calibration Bath comes equipped with process electronics for the connection of an external reference probe.
General specifications | |
Size | Height: 382 mm (15 in) Width: 242 mm (9.5 in) Depth: 400 mm (15.7 in) |
Weight | 6109A: 16 kg (35 lb) 7109A: 20 kg (44 lb) |
Fluid volume | 2.5 liters (0.66 gallons) |
Fluid working area | 75 mm x 75 mm (3 in x 3 in) |
Maximum fluid depth | 154 mm (6.1 in) |
Factory calibration | Traceable NVLAP-accredited calibration included |
Temperature specifications The temperature specifications describe the Absolute Instrumental Uncertainty at 95 % level of confidence (coverage factor k = 2) within one year from calibration. The specifications include environmental temperature effects from 13 °C to 33 °C. | ||
6109A | 7109A | |
Range (at 25 °C ambient temperature) | 35 °C to 250 °C | –25 °C to 140 °C (–15 °C cover off) |
Accuracy (maximum permissible error) | 0.1 °C | 0.1 °C |
Display resolution | 0.01 °, 0.001° | 0.01 °, 0.001° |
Stability | 0.015 °C | 0.01 °C |
Typical uniformity | 0.03 °C up to 200 °C 0.04 °C above 200 °C | 0.02 °C |
Repeatability | 0.04 °C | 0.04 °C |
Typical heating time | 35 °C to 100 °C: 25 minutes 100 °C to 250 °C: 45 minutes | –25 °C to 25 °C: 35 minutes 25 °C to 140 °C: 55 minutes |
Typical cooling time | 250 °C to 100 °C: 35 minutes 100 °C to 35 °C: 55 minutes | 140 °C to 25 °C: 45 minutes 25 °C to –25 °C: 75 minutes |
Typical settling time | 15 minutes | 10 minutes |
Process Input Module specifications (-P models) The Input Module specifications describe the Absolute Instrumental Uncertainty at 95 % level of confidence (coverage factor k = 2) within one year from calibration. The specifications include calibration uncertainty, linearity, repeatability, resolution, stability, and environmental temperature effects from 13 °C to 33 °C. | ||||
Function | Range | Accuracy (Maximum Permissible Error) | ||
Reference input resistance | 0 Ω to 42 Ω 42 Ω to 400 Ω | 0.0025 Ω 0.006 % | ||
Reference input temperature | -25 °C to 250 °C | 0.007 % + 0.015 °C | ||
Resistance sensing current | 1 mA | 8 % | ||
DUT 4-wire resistance | 0 Ω to 31 Ω 31 Ω to 400 Ω | 0.0025 Ω 0.008% | ||
DUT 3-wire resistance accuracy | 0 Ω to 400 Ω | 0.12 Ω | ||
Thermocouple mV | -10 mV to 100 mV | 0.025 % + 0.01 mV | ||
Reference junction temperature | 0 °C to 40 °C | 0.35 °C | ||
Thermocouple temperature | -25 °C to 250 °C | J: 0.44 °C E: 0.44 °C L: 0.42 °C | K: 0.49 °C N: 0.57 °C U: 0.48 °C | T: 0.53 °C M: 0.48 °C |
0 °C to 250 °C | R: 1.92 °C D: 1.12 °C | S: 1.88 °C G/W: 3.34 °C | C: 0.84 °C | |
mA range | 0 mA to 22 mA | 0.02 % + 0.002 mA | ||
Loop power | 24 V dc | ± 6 V |