Model | MDL4U200 | MDL4U252 | MDL4U302 | MDL4U305 | MDL4U400 | MDL4U505 | MDL4U600 | |
Input rating | ||||||||
Input Voltage | 0 to 80 V | 0 to 80 V | 0 to 80 V | 0 to 500 V | 0 to 80 V | 0 to 500 V | 0 to 80 V | |
Input Current | Low | 0 to 4 A | 0 to 3 A | 0 to 4.5 A | 0 to 3 A | 0 to 6 A | 0 to 3 A | 0 to 12 A |
High | 0 to 40 A | 0 to 20 A | 0 to 45 A | 0 to 20 A | 0 to 60 A | 0 to 30 A | 0 to 120 A | |
Input Power | 200 W | 250 W / 50 W (1) | 300 W / 300 W (1) | 300 W | 400 W | 500 W | 600 W | |
Channels | 1 | 2 | 2 | 1 | 1 | 1 | 1 | |
Minimum Operating Voltage | Low | 0.10 V at 4 A | 0.15 V at 3 A | 0.14 V at 4.5 A | 0.7 V at 3 A | 0.15 V at 6 A | 0.54 V at 3 A | 0.18 V at 12 A |
High | 1 V at 40 A | 1 V at 20 A | 1.4 V at 45 A | 4.5 V at 20 A | 1.5 V at 60 A | 5.4 V at 30 A | 1.8 V at 120 A | |
CV mode | ||||||||
Range | Low | 0 to 18 V | ||||||
High | 0 to 80 V | 0 to 500 V | 0 to 80 V | 0 to 500 V | 0 to 80 V | |||
Resolution | Low | 1 mV | ||||||
High | 10 mV | |||||||
Accuracy | Low | ± (0.05% + 0.02% F.S.) | ± (0.05 % + 0.025% F.S.) | ± (0.05% + 0.02% F.S.) | ||||
High | ± (0.05% + 0.025% F.S.) | |||||||
CC mode | ||||||||
Range | Low | 0 to 4 A | 0 to 3 A | 0 to 4.5 A | 0 to 3 A | 0 to 6 A | 0 to 3 A | 0 to 12 A |
High | 0 to 40 A | 0 to 20 A | 0 to 45 A | 0 to 20 A | 0 to 60 A | 0 to 30 A | 0 to 120 A | |
Resolution | Low | 0.1 mA | 1 mA | 0.1 mA | ||||
High | 1 mA | 10 mA | 1 mA | |||||
Accuracy | Low | ± (0.05% + 0.05% F.S.) | ± (0.05% + 0.1% F.S.) | |||||
High | ± (0.05% + 0.05% F.S.) | ± (0.1% + 0.1% F.S.) | ||||||
CR mode | ||||||||
Range | Low | 0.05 Ω to 10 Ω | 0.25 Ω to 10 Ω | 0.05 Ω to 10 Ω | 0.2 Ω to 10 Ω | |||
High | 10 Ω to 7.5 kΩ | |||||||
Resolution | 16-bit | |||||||
Accuracy | Low | 0.01% + 0.08 S | ||||||
High | 0.01% + 0.0008 S | |||||||
CW mode | ||||||||
Range | 200 W | 250 W | 300 W | 400 W | 500 W | 600 W | ||
Resolution | 10 mW | |||||||
Accuracy | ± (0.2% + 0.2% F.S.) | |||||||
Transient mode (CC mode) | ||||||||
T1&T2(2) | 20 μs to 3600 s / Res: 5 μs to 10 ms | |||||||
Accuracy | 5 μs + 100 ppm | |||||||
Slew Rate(3) | Low | 0.0001 to 0.25 A/μs | 0.0001 to 0.2 A/μs | 0.0001 to 0.25 A/μs | 0.0001 to 0.1 A/μs | 0.0001 to 0.25 A/μs | 0.0001 to 0.1 A/μs | 0.0001 to 0.25 A /μs |
High | 0.001 to 2.5 A/μs | 0.001 to 2 A/μs | 0.001 to 2.5 A/μs | 0.001 to 1 A/μs | 0.001 to 2.5 A/μs | 0.001 to 1 A/μs | 0.001 to 2.5 A/μs |
(1) MDL4U252: The user can allocate 250 W to either channel up to 300 W total (e.g. 50 W/250 W, 250 W/50 W, 150 W/150 W). MDL4U302: The user can allocate 300 W to either channel up to 600 W total (e.g. 300 W/300 W).
(2) Fast pulse trains with large transitions may not be achievable.
(3) The slew rate specifications are not warranted, but are descriptions of typical performance. The actual transition time is defined as the time for the input to change from 10% to 90%, or vice versa, of the programmed current values. In case of very large load changes, e.g. from no load to full load, the actual transition time will be larger than the expected time. The load will automatically adjust the slew rate to fit within the range (high or low) that is closest to the programmed value.
Model | MDL4U200 | MDL4U252 | MDL4U302 | MDL4U305 | MDL4U400 | MDL4U505 | MDL4U600 | |
Readback voltage | ||||||||
Range | Low | 0 to 18 V | ||||||
High | 0 to 80 V | 0 to 500 V | 0 to 80 V | 0 to 500 V | 0 to 80 V | |||
Resolution | Low | 0.1 mV | 1 mV | 0.1 mV | 1 mV | 0.1 mV | ||
High | 1 mV | 10 mV | 1 mV | 10 mV | 1 mV | |||
Accuracy | ± (0.025% + 0.025% F.S.) | |||||||
Readback current | ||||||||
Range | Low | 0 to 4 A | 0 to 3 A | 0 to 4.5 A | 0 to 3 A | 0 to 6 A | 0 to 3 A | 0 to 12 A |
High | 0 to 40 A | 0 to 20 A | 0 to 45 A | 0 to 20 A | 0 to 60 A | 0 to 30 A | 0 to 120 A | |
Resolution | Low | 0.01 mA | 0.1 mA | 0.01 mA | 0.1 mA | |||
High | 0.1 mA | 1 mA | 0.1 mA | 1 mA | ||||
Accuracy | Low | ± (0.05% + 0.05% F.S.) | ± (0.05% + 0.1% F.S.) | |||||
High | ± (0.05% + 0.05% F.S.) | ± (0.1% + 0.1% F.S.) | ||||||
Readback power | ||||||||
Range | 200 W | 250 W | 300 W | 400 W | 500 W | 600 W | ||
Resolution | 10 mW | |||||||
Accuracy | ± (0.2% + 0.2% F.S.) | |||||||
Protection range (typical) | ||||||||
OPP | 200 W | 250 W | 310 W | 300 W | 400 W | 500 W | 600 W | |
OCP | Low | 4.4 A | 3.3 A | 5 A | 3.3 A | 6.6 A | 3.3 A | 13.2 A |
High | 44 A | 22 A | 50 A | 22 A | 66 A | 33 A | 132 A | |
OVP | 82 V | 510 V | 82 V | 510 V | 82 V | |||
OTP | 185 ºF (85 ºC) | |||||||
General (typical) | ||||||||
Short Circuit | ||||||||
Current (CC) | Low | 4 A | 3 A | 5 A | 3 A | 6 A | 3 A | 12 A |
High | 40 A | 30 A | 50 A | 20 A | 60 A | 30 A | 120 A | |
Voltage (CV) | 0 V | |||||||
Resistance (CR) | 25 mΩ | 50 mΩ | 30 mΩ | 220 mΩ | 25 mΩ | 180 mΩ | 15 mΩ | |
Input Terminal Impedance | 300 kΩ | 1 MΩ | 300 kΩ | 1 MΩ | 300 kΩ | |||
Safety | EN61010-1:2010, EU Low Voltage Directive (LVD) 2014/35/EU | |||||||
Electromagnetic Compatibility |
| |||||||
Warranty | 3 Years | |||||||
Dimensions | 3.2” x 6.7” x 22.6” (82 x 170.5 x 573 mm) | |||||||
Weight | 11 lbs (5 kg) |