The GW Instek PFR-100LGL DC power supply has a longer life cycle because of its fanless structure. The fanless structure allows users to focus on their experiments and tests in a quiet environment. The PFR-100LGL has a maximum output voltage of 50V (at 2A) with rated 100W. The PFR-100LGL also has a GPIB and LAN interface.
Model | PFR-100L | PFR-100M | ||
OUTPUT RATING | Rated Output Voltage Rated Output Current Rated Output Power | 50V 10A 100W | 250V 2A 100W | |
REGULATION(CV) | Load Regulation (*2) Line Regulation (*1) | 10mV 3mV | 33mV 5mV | |
REGULATION(CC) | Load Regulation (*9) Line Regulation (*1) | 10mA 8mA | 3.2mA 1.2mA | |
RIPPLE & NOISE (*3) | Vp-p (*4) Vr.m.s.(*5) A r.m.s. | 50mV 4mV 10mA | 150mV 15mV 2mA | |
PROGRAMMING ACCURACY | Voltage Current | 0.1% of setting + 0.2% of setting + | 40mV 20mA | 200mV 2mA |
MEASUREMENT ACCURACY | Voltage Current | 0.1% of reading + 0.2% of reading + | 40mV 20mA | 200mV 2mA |
RESPONSE TIME | Rise Time (*6) Fall Time (*7) Transient Response Time (*8) | Rated load Rated load No load | 50ms 100ms 500ms 1.5ms | 200ms 300ms 3000ms 2ms |
PROGRAMMING RESOLUTION | Voltage Current | 2mV 1mA | 10mV 0.1mA | |
MEASUREMENT RESOLUTION | Voltage Current | 2mV 1mA | 10mV 0.1mA | |
PROTECTION FUNCTION | Over Voltage Protection (OVP) Over Current Protection (OCP) Under Voltage Limit (UVL) Over Temperature Protection (OTP) Low AC Input Protection (AC-Fail) Power Limit (Power Limit) | Setting range Setting range Setting range Operation Operation Operation | 5~55V 1~11A 0~52.5V Turn the output off. Turn the output off. Turn the output off. | 5~275V 0.2~2.2A 0~262.5V Turn the output off. Turn the output off. Turn the output off. |
FRONT PANEL DISPLAY ACCURACY, 4 DIGITS | Voltage Current | 0.1% of reading + 0.2% of reading + | 40mV 20mA | 200mV 2mA |
ENVIRONMENT CONDITION | Operaing Temperature Storage Temperature Operating Humidity Storage Humidity | 0°C to 40°C -20°C to 70°C 20% to 80% RH; No condensation 20% to 85% RH; No condensation | ||
READBACK TEMP. COEFFICIENT (After A 30 Minute Warm-up) | Voltage Current | 100ppm/℃ 200ppm/℃ | ||
OTHER | Analog Control Interface AC Input | Yes LAN/GPIB 85~265VAC, 47~63Hz, single pahse | ||
DIMENSIONS & WEIGHT | 70(W)x124(H)x300(D)mm; Approx.2.5kg |
Notes:
*1: At 85 ~ 132Vac or 170 ~ 265Vac, constant load.
*2: From No-load to Full-load, constant input voltage. Measured at the sensing point in Remote Sense.
*3: Measure with JEITA RC-9131B (1:1) probe
*4: Measurement frequency bandwidth is 10Hz to 20MHz.
*5: Measurement frequency bandwidth is 5Hz to 1MHz.
*6: From 10%~90% of rated output voltage, with rated resistive load.
*7: From 90%~10% of rated output voltage, with rated resistive load.
*8: Time for output voltage to recover within 0.1% + 10mV of its rated output for a
load change from 50 to 100% of its rated output current.
*9: For load voltage change, equal to the unit voltage rating, constant input voltage.