GW Instek PFR-100LGL 50 Volt 10 Amp Fanless Multi-Range DC Power Supply with GPIB and LAN Interface

SKU
PFR-100LGL
Manufacturer:
GW Instek
GW Instek

Key Features:

  • Maximum output voltage of 50V (at 2A)
  • Constant Power Output for 5 Times Multi-range(V&I) Operation
  • Natural Convection Cooling Design (Fanless Structure)
  • Preset Memory Function
  • Output ON/OFF Delay Function
  • CV, CC Priority Mode
  • Adjustable Slew Rate For Voltage and Current
  • Bleeder Circuit Control
  • Protections: OVP, OCP, AC FAIL and OTP
  • Support Front and Rear Panel Output
  • LAN+GPIB Interface
  • Web Server Monitoring and Control
  • External Analog Control and Monitor Function
  • Remote Sensing Function
Product Options:
$1,043.00

Secured and trusted checkout with

The GW Instek PFR-100LGL DC power supply has a longer life cycle because of its fanless structure. The fanless structure allows users to focus on their experiments and tests in a quiet environment. The PFR-100LGL has a maximum output voltage of 50V (at 2A) with rated 100W. The PFR-100LGL also has a GPIB and LAN interface.
 ModelPFR-100LPFR-100M
OUTPUT RATINGRated Output Voltage
Rated Output Current
Rated Output Power
50V
10A
100W
250V
2A
100W
REGULATION(CV)Load Regulation (*2)
Line Regulation (*1)
10mV
3mV
33mV
5mV
REGULATION(CC)Load Regulation (*9)
Line Regulation (*1)
10mA
8mA
3.2mA
1.2mA
RIPPLE & NOISE (*3)Vp-p (*4)
Vr.m.s.(*5)
A r.m.s.
50mV
4mV
10mA
150mV
15mV
2mA
PROGRAMMING ACCURACYVoltage

Current
0.1% of setting +
0.2% of setting +
40mV

20mA
200mV

2mA
MEASUREMENT ACCURACYVoltage

Current
0.1% of reading +
0.2% of reading +
40mV

20mA
200mV

2mA
RESPONSE TIMERise Time (*6)
Fall Time (*7)

Transient Response Time (*8)
Rated load
Rated load
No load
50ms
100ms
500ms
1.5ms
200ms
300ms
3000ms
2ms
PROGRAMMING RESOLUTIONVoltage
Current
2mV
1mA
10mV
0.1mA
MEASUREMENT RESOLUTIONVoltage
Current
2mV
1mA
10mV
0.1mA
PROTECTION FUNCTIONOver Voltage Protection (OVP)
Over Current Protection (OCP)
Under Voltage Limit (UVL)
Over Temperature Protection (OTP)
Low AC Input Protection (AC-Fail)
Power Limit (Power Limit)
Setting range

Setting range

Setting range

Operation

Operation

Operation
5~55V

1~11A

0~52.5V

Turn the output off.
Turn the output off.

Turn the output off.
5~275V

0.2~2.2A


0~262.5V


Turn the output off.

Turn the output off.

Turn the output off.
FRONT PANEL DISPLAY
ACCURACY, 4 DIGITS
Voltage

Current
0.1% of reading +
0.2% of reading +
40mV

20mA
200mV

2mA
ENVIRONMENT CONDITIONOperaing Temperature
Storage Temperature
Operating Humidity
Storage Humidity
0°C to 40°C
-20°C to 70°C
20% to 80% RH; No condensation
20% to 85% RH; No condensation
READBACK TEMP. COEFFICIENT
(After A 30 Minute Warm-up)
Voltage
Current
100ppm/℃
200ppm/℃
OTHERAnalog Control
Interface
AC Input
Yes
LAN/GPIB
85~265VAC, 47~63Hz, single pahse
DIMENSIONS & WEIGHT70(W)x124(H)x300(D)mm; Approx.2.5kg

Notes:
*1: At 85 ~ 132Vac or 170 ~ 265Vac, constant load.

*2: From No-load to Full-load, constant input voltage. Measured at the sensing point in Remote Sense.
*3: Measure with JEITA RC-9131B (1:1) probe
*4: Measurement frequency bandwidth is 10Hz to 20MHz.
*5: Measurement frequency bandwidth is 5Hz to 1MHz.
*6: From 10%~90% of rated output voltage, with rated resistive load.
*7: From 90%~10% of rated output voltage, with rated resistive load.
*8: Time for output voltage to recover within 0.1% + 10mV of its rated output for a
load change from 50 to 100% of its rated output current.
*9: For load voltage change, equal to the unit voltage rating, constant input voltage.

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account

Products In This Series