The GW Instek GPT-9503 is a three-in-one multi-channel tester that only provides the setting of H or X of the 8 channels, making the GPT-9503 more suitable for general components such as passive components for high-voltage testing between two points.
Includes:
• Quick Start Guide x1, CD x1(completed user manual)
• Power cord x1
• Test lead GHT-115 x1, GHT-116B x1, GHT-116R x8
AC Withstanding | |
Output-Voltage Range | 0.050kV~5.000kV |
Output-Voltage Resolution | 1V |
Output-Voltage Accuracy | ± (1% of setting + 5V) [no load] |
Maximum Rated Load | 150 VA (5kV/30mA) |
Maximum Rated Current | 30mA 0.001mA ~ 10mA (0.05kV≦ V ≦0.5kV) 0.001mA ~ 30mA (0.5kV< V ≦5kV) |
Output-Voltage Waveform | Sine wave |
Voltage Regulation | ± (1% + 5V) [maximum rated load → no load] |
Output-Voltage Frequency | 50 Hz / 60 Hz selectable |
Voltmeter Accuracy | ± (1% of reading + 5V) |
Current Measurement Range | 0.001mA ~ 30.00mA |
Current Best Resolution | 1μA (0.001mA ~ 9.999mA) 10μA (10.00mA ~ 30.00mA) |
Current Measurement Accuracy | ±(1.5% of reading + 30μA) |
Current Offset | 80μA maximum |
ARC Detect | Yes |
RAMP TIME (Rise T ime) | 0.1s~999.9s |
FALL Time | OFF~999.9s |
WAIT T ime | OFF~999.9s |
TIMER (Test Time) | CONT2, 0.3s~999.9s |
TIMER Accuracy | ± (100ppm + 20ms) |
GND | ON/OFF |
DC Withstanding | |
Output-Voltage Range | 0.050kV~6.000kV |
Output-Voltage Resolution | 1V |
Output-Voltage Accuracy | ± (1% of setting + 5V) [no load] |
Maximum Rated Load | 50W (5kV/10mA) |
Maximum Rated Current | 10mA 0.001mA ~ 2mA (0.05kV≦ V ≦0.5kV) 0.001mA ~ 10mA (0.5kV< V ≦6kV) |
Voltmeter Accuracy | ± (1% of reading + 5V) |
Voltage Regulation | ± (1% + 5V) [maximum rated load → no load] |
Current Measurement Range | 0.001mA ~ 10.00mA |
Current Best Resolution | 0.1μA (0.1μA ~ 999.9μA) 1μA (1μA ~ 9.999mA) 10μA (10.00mA) |
Current Measurement Accuracy | ±(1% of reading + 1μA) when I Reading < 1mA ±(1% of reading + 10μA) when I Reading ≧1mA |
Current Offset | 5μA maximum |
ARC Detect | Yes |
RAMP TIME (Rise Time) | 0.1s~999.9s |
FALL Time | OFF~999.9s |
WAIT Time | OFF~999.9s |
TIMER (Test Time) | CONT2, 0.3s~999.9s |
TIMER Accuracy | ± (100ppm + 20ms) |
GND | ON/OFF |
Insulation Resistance | |
Output Voltage | 50V~1000V dc |
Output-Voltage Resolution | 1V |
Output-Voltage Accuracy | ± (1% of setting + 5V) [no load] |
Resistance Measurement | 0.1MΩ~ 10GΩ |
Test Voltage | Measurement Range / Accuracy |
50V≦ V<500V | 0.1MΩ~1MΩ:±(5% of reading + 3 count) 1 MΩ~50MΩ:±(5% of reading + 1 count) 51MΩ~2GΩ:±(10% of reading + 1 count) |
500V≦ V ≦1000V | 0.1MΩ~1MΩ:±(5% of reading + 3 count) 1 MΩ~500MΩ:±(5% of reading + 1 count) 501MΩ~10GΩ:±(10% of reading + 1 count) |
Voltage Regulation | ± (1% + 5V) [maximum rated load → no load] |
Voltmeter Accuracy | ± (1% of reading + 5V) |
Short-Circuit Current | 10mA max. |
Output Impedance | 2kΩ |
RAMP TIME (Rise T ime) | 0.1s~999.9s |
FALL Time | OFF~999.9s |
WAIT TIME | OFF~999.9s |
TIMER (Test Time)* | 0.3s~999.9s |
TIMER Accuracy | ± (100ppm + 20ms) |
GND | ON/OFF |
Continuity Test | |
Output-Current | 100mA dc |
Ohmmeter Measurement Accuracy | 1W ± 0.2W, ON/OFF |
Interface | |
Signal I/O | Standard |
RS-232C | Standard |
USB (device) | Standard |
USB (host) | Standard (for parameter / LCD hardcopy) |
Rear Output | Scanner |
DISPLAY | |
| 4.3” color LCD |
POWER SOURCE | |
| AC 100V~240V ± 10%, 50Hz/60Hz |
DIMENSION & WEIGHT | |
| 320(W) x 120(H) x 435(D) mm; Approx. 11kg |