We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
Part of the P3800 series of high pressure hydraulic deadweight testers, Fluke Calibration’s P3830 is specifically configured with a 30,000 PSIG pressure range. The P3830 comes equipped with all the necessary items for operation, such as a top quality hand pump, oil reservoir, test station and even adjustable feet. Even more, the deadweight tester ensures premium long term stability its tungsten carbide manufactured pistons.
The deadweight tester is designed to improve reliability, augment accuracy and simplify user operation, making the P3830 from Fluke Calibration a reliable and accurate solution for calibrating high pressure devices.
Pressure Range | |
500 to 30,000 psi 40 to 2,000 bar 4 to 200 MPa 40 to 2000 kgf/cm2 | |
Accuracy | |
Accuracy* | Standard accuracy is 0.02% of Reading1 |
*Accuracy is based on % of Reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10%, ± (accuracy class) x 10% of the piston range. | |
Materials of Construction | |
Standard weight material | Series 3 non-magnetic, austenitic, stainless steel |
Weight density | 7.8 g/cm3 |
Piston material | Tungsten carbide with nickel binder |
Cylinder material | Tungsten carbide with cobalt binder |
Thermal coefficient of expansion | 11 ppm/°C |
General | |
Test port adapters | 9/16 – 18 UNF (Autoclave); 3/8 BSP; 1/2 BSP and 3/4 BSP |
Instrument weight | 66 lbs. (30 Kg) Instrument Base only |
Instrument size | 18 x 13.5 x 19 in (455 x 340 x 478 mm) |
Reservoir volume | 235 cc (14.3 in3) |
Intensifier ratio | 7 to 1 |
Seal materials | Nitrile |
Operating fluid | Dioctyl Sebacate (DOS), our reference 55-600 |
Options | |
Presscal software | Windows based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Supports additional 12 pressure units (bar, mbar, MPa, kPa, psi, kg/cm2, atm, inH2O, mH2O, mmH2O, inHg, mmHg). |
Conversion weight sets | For applications that require nominal pressure increments in measurement units different than the main unit, conversion weight sets are available. Available measurement units are: psi, bar kgf/cm2 or MPa. |
- • Transducers
- • Gauges
- • Transmitters
- • Pressure Switches
- • Calibrators