Fluke Calibration 4180 Precision Infrared Calibrator

SKU
4180-156
Manufacturer:
Fluke Calibration
Fluke Calibration

Key Features:

  • Calibrated radiometrically for meaningful, consistent results
  • Accredited calibration included
  • Accurate, reliable performance from –15 °C to 500 °C
  • Large target size of 152 mm (6 in) is required for calibrating most thermometers
  • Light instrument weight of 8.6 kg (19 lbs) makes it easy to lift and carry
  • Intuitive, easy-to-read display that indicates when temperature is stable
Product Options:
$11,861.00

Secured and trusted checkout with

As part of their 4180 series of precision infrared calibrators, the 4180 from Fluke Calibration is both efficient and accurate.

The 4180 Precision Infrared Calibrator is equipped to reach temperatures from -15°C to 120°C. The calibrator highlights the importance of uniformity in infrared temperature calibration work, as an infrared thermos meter will “see” as much as the whole target when situated at the correct calibration distance.

The 4180 boasts an accuracy as superior as ±0.40°C at -15°C and can satisfy specifications without added emissivity-related corrections, allowing legitimate test uncertainty ratios (TUR) as good as 4:1.

 

Includes:

  • • 4180 IR Calibrator
  • • Report of Calibration (NVLAP Accredited)
  • • Power Cord
  • • User’s Guide
  • • Documentation CD
  • • Target Cover
  • • Serial Cable

Specifications

4180

4181

Temperature range (@ 23°C ambient, 0.95 emissivity)

–15°C to 120°C

 35°C to 500°C

Display accuracy 1

±0.40°C at –15°C
±0.40°C at 0°C
±0.50°C at 50°C
±0.50°C at 100°C
±0.55°C at 120°C

±0.35°C at 35°C
±0.50°C at 100°C
±0.70°C at 200°C
±1.20°C at 350°C
±1.60°C at 500°C

Stability

±0.10°C at –15°C
±0.05°C at 0°C
±0.10°C at 120°C

±0.05°C at 35°C
±0.20°C at 200°C
±0.40°C at 500°C

Uniformity (5.0 in dia of center of target) 2

±0.15°C at –15°C
±0.10°C at 0°C
±0.25°C at 120°C

±0.10°C at 35°C
±0.50°C at 250°C
±1.00°C at 500°C

Uniformity (2.0 in dia of center of target) 2

±0.10°C at -15°C
±0.10°C at 0°C
±0.20°C at 120°C

±0.10°C at 35°C
±0.25°C at 250°C
±0.50°C at 500°C

Heating time

15 min: –15°C to 120°C
14 min: 23°C to 120°C

20 min: 35°C to 500°C

Cooling time

15 min: 120°C to 23°C
20 min: 23°C to –15°C

100 min: 500°C to 35°C
40 min: 500°C to 100°C

Stabilization time

10 minutes

10 minutes

Nominal emissivity 3

0.95

0.95

Thermometer emissivity compensation

0.9 to 1.0

0.9 to 1.0

Target diameter

152.4 mm (6 in)

152.4 mm (6 in)

Computer interface

RS-232

RS-232

Power

115 V AC (±10%), 6.3 A,
50/60 Hz, 630 W
230 V AC (±10%), 3.15 A,
50/60 Hz, 630 W

115 V AC (±10%), 10 A,
50/60 Hz, 1000 W
230 V AC (±10%), 5 A,
50/60 Hz, 1000 W

Fuse(s)

115 V AC 6.3 A, 250 V, slow blow
230 V AC 3.15 A, 250 V, T

115 V AC 10 A, 250 V, fast blow
230 V AC 5 A, 250 V, F

Size (H x W x D)

356 x 241 x 216 mm
(14 x 9.5 x 8.5 in)

356 x 241 x 216 mm
(14 x 9.5 x 8.5 in)

Weight

9.1 kg (20 lb)

9.5 kg (21 lb)

Safety

EN 61010-1:2001, CAN/CSA C22.2 No.
61010.1-04

1. For 8 um to 14 um spectral band thermometers with emissivity set between 0.9 and 1.0
2. The uniformity specification refers to how IR thermometers with different spot sizes both focused at the center of the targer will measure the same temperature.
3. The target has a nominal emissivity of 0.95, however it is radiometrically calibrated to minimize emissivity related uncertainties.

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account

Products In This Series

As low as