We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
Whether you’re using in-line or handheld infrared pyrometers, you need good calibration standards to verify their accuracy. Our portable IR calibrators provide stable blackbody targets for calibrating noncontact IR thermometers from –30°C to 500°C.
These units feature a temperature controlled measurement surface with a diameter of 2.25" (57 mm). The target temperature can be controlled in set-point increments of 0.1° from –30°C to 500°C and a well is located directly behind the blackbody surface for contact calibration.
For IR calibrations above normal ambient, the 9132 provides a stable measurement surface up to 500°C (932°F). Short heating and cooling times mean you won’t have to wait long to get your work done.
Specifications | ||
Temperature Range | 50 °C to 500 °C (122°F to 932°F) | |
Accuracy | ± 0.5 °C at 100 °C (± 0.9°F at 212°F) ± 0.8 °C at 500 °C (± 1.4 °F at 932 °F) | |
Stability | ± 0.1°C at 100°C (± 0.18°F at 212°F) ± 0.3°C at 500°C (± 0.54°F at 932°F) | |
Target Size | 57 mm (2.25 in) | |
Target Emissivity | 0.95 (± 0.02 from 8 to 14 mm) | |
Resolution | 0.1 ° | |
Heating Time | 30 minutes (50 °C to 500 °C) | |
Cooling Time | 30 minutes (500 °C to 100 °C) | |
Computer Interface | RS-232 I/O included with 9930 Interface-it software | |
Power | 115 V ac (± 10 %), 3 A or 230 V ac (± 10 %), 1.5 A, switchable, 50/60 Hz, 340 W | |
Size (HxWxD) | 102 x 152 x 178 mm (4 x 6 x 7 in ) | |
Weight | 1.8 kg (4 lb) | |
NIST-Traceable Contact Calibration | Data at 50°C, 100°C, 200°C, 250°C, 300°C, 400°C, and 500°C |