We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The Fluke 87V features accuracy and diagnostic functions for maximum industrial productivity. The new Fluke 87V has improved measurement capabilities, troubleshooting features, resolution and accuracy to solve more problems in electronics, plant automation, power distribution, and electro-mechanical equipment. The 87-5 is the next generation of the ever popular Fluke 80 series DMM, but with more problem solving power, safety, convenience and durability.
Save when you buy the Fluke 87V/E2 Industrial Electrician Combo Kit
The Combo Pak Save Over $50 in useful accessories for your new Fluke 87-5.
The Pak Includes:
- 87V Multimeter
- TL224 silicone test leads
- AC220 long reach alligator clips
- TP238 test probes
- TPAK ToolPak™ Magnetic Meter Hanging Kit
- 80BK temperature probe
- C35 lightweight soft accessory case
- 9V battery (installed)
- Getting Started Manual
- CD with detailed manual
Add the Fluke TPAK Meter Hanging Kit
Now you can hang you meter almost anywhere with the new Fluke TPAK ToolPak accessories kit. ToolPak includes a universal hanger, a magnet, a 17-inch (23 cm) "hook-and-loop" hanger strap. The universal hanger is designed to fit into the special module on the back of the meter. The universal hanger allows you to hang your meter on a nail, hook or other object. You can add a piece of rope, flexible wire or cable to hang the meter in almost any position. The 17-inch (23 cm) hook-and-loop strap is used with Fluke's universal hanger to wrap around and connect to a pipe, beam or any other convenient support structure. And finally, the strong magnet accessory will attach to an electrical panel, a metal doorjamb or any other ferrous-type metal.
Specifications | ||
Voltage DC | Maximum voltage | 1000 V |
Accuracy | ±(0.05% + 1) | |
Maximum resolution | 10 µV | |
Voltage AC | Maximum voltage | 1000 V |
Accuracy | ±(0.7% + 2) True RMS | |
AC bandwidth | 20 kHz with low pass filter; 3 db @ 1 kHz | |
Maximum resolution | 0.1 mV | |
Current DC | Maximum amps | 10 A (20 A for 30 seconds maximum) |
Amps accuracy | ±(0.2% + 2) | |
Maximum resolution | 0.01 µA | |
Current AC | Maximum amps | 10 A (20 A for 30 seconds maximum) |
Amps accuracy | ±(1.0% + 2) True RMS | |
Maximum resolution | 0.1 µA | |
Resistance | Maximum resistance | 50 MΩ |
Accuracy | ±(0.2% + 1) | |
Maximum resolution | 0.1 Ω | |
Capacitance | Maximum capacitance | 9,999 µF |
accuracy | ±(1% + 2) | |
Maximum resolution | 0.01 nF | |
Frequency | Maximum frequency | 200 kHz |
Accuracy | ±(0.005% + 1) | |
Maximum resolution | 0.01 Hz | |
Duty cycle | Maximum duty cycle | 99.9% |
Accuracy | ±(0.2% per khz + 0.1%) | |
Maximum resolution | 0.1% | |
Temperature measurement | –200.0°C – 1090°C –328.0°F – 1994.0°F excluding probe | |
80 BK temperature probe | –40.0°C – 260°C –40.0°F – 500°F, 2.2°C or 2% whichever is greater | |
Conductance | Maximum conductance | 60.00 nS |
Accuracy | ±(1.0% + 10) | |
Maximum resolution | 0.01 nS | |
Diode | Range | 3 V |
Resolution | 1 mV | |
Accuracy | ±(2% + 1) | |
Duty cycle range | Accuracy | Within ±(0.2% per kHz + 0.1%) |
Environmental Specifications | ||
Operating temperature | –20°C to + 55°C | |
Storage temperature | –40°C to + 60°C | |
Humidity (without condensation) | 0% – 90% (0°C – 35°C) 0% – 70% (35°C – 55°C) | |
Operating Altitude | 2000 m | |
Safety Specifications | ||
Overvoltage category | EN 61010–1 to 1000 V CAT III, 600V CAT IV | |
Agency approvals | UL, CSA, TÜV, VDE listed | |
Mechanical and General Specifications | ||
Size | 201 x 98 x 52 mm (with holster) | |
Weight | 355 g 624 g – with holster | |
Display | Digital | 6000 counts updates 4/sec. 19,999 counts in high–resolution mode |
Analog | 32 segments, updates 40/sec | |
Frequency | 19,999 counts, updates 3/sec at > 10 Hz | |
Warranty | Lifetime | |
Battery Life | Alkaline | ~400 hours typical, without backlight |
Shock | 1 Meter drop per IEC 61010–1:2001 | |
Vibration | Per MIL–PRF–28800 for a Class 2 instrument |
Each Fluke 87V Meter includes
- 9V battery
- Test Leads
- Rubber Holster with Stand
- Manual
What is "Limited Lifetime Warranty"?
Each Fluke 20, 70, 80, 170 and 180 series DMM will be free from defects
in material and workmanship for its lifetime. As used herein, "lifetime" is
defined as seven years after Fluke discontinues manufacturing the product,
but the warranty period shall be at least ten years from date of purchase.
In what way is the warranty limited?
- It does not cover battery or fuses.
- It does not cover abuse or usage outside of specifications.
- It does not cover normal wear and tear of mechanical parts.
- It covers the LCD for 10 years only (state-of-the-art for LCDs).
- It covers original user purchaser only