Specifications are based on the following conditions:
- One year calibration cycle
- Calibration temperature TCAL = 23 °C
- Accuracy specifications: ± (% of reading + % of range), valid for TCAL ±5 ºC
- All 24 hours accuracy specifications are typical
- Add temperature coefficient for every ºC outside TCAL ±5 ºC
- For DC: The technical specifications are valid after a warm-up time of 60 minutes, with the integration time set to 10 or 100 NPLC, and auto zero enabled.
- For AC: The technical specifications are valid after a warm-up time of 60 minutes, slow AC filtering enabled, using a sine wave input signal.
DC Voltage
Range(1) |
Resolution |
Full Scale Reading | Accuracy |
Temperature Coefficient / °C | ||
24 Hours(2) TCAL ± 1 °C | 90 Days TCAL ± 5 °C | 1 Year TCAL ± 5 °C | ||||
100 mV | 0.1 µV | 119.9999 mV | 0.0030 + 0.0030 | 0.0040 + 0.0035 | 0.0050 + 0.0035 | 0.0005 + 0.0005 |
1 V | 1 µV | 1.199999 V | 0.0020 + 0.0006 | 0.0030 + 0.0007 | 0.0040 + 0.0007 | 0.0005 + 0.0001 |
10 V | 10 µV | 11.99999 V | 0.0015 + 0.0004 | 0.0020 + 0.0005 | 0.0035 + 0.0005 | 0.0005 + 0.0001 |
100 V | 100 µV | 119.9999 V | 0.0020 + 0.0006 | 0.0035 + 0.0006 | 0.0045 + 0.0006 | 0.0005 + 0.0001 |
1000 V | 1 mV | 1050.000 V | 0.0020 + 0.0006 | 0.0035 + 0.0010 | 0.0045 + 0.0010 | 0.0005 + 0.0001 |
(1) 20% overrange on all ranges except 1000 V range which can only test 5% overrange.
(2) Relative to calibration standards.
DC Current
Range |
Burden Voltage |
Resolution |
Full Scale Reading | Accuracy |
Temperature Coefficient / °C | ||
24 Hours TCAL ± 1 °C | 90 Days TCAL ± 5 °C | 1 Year TCAL ± 5 °C | |||||
100 µA | < 0.011 V | 0.1 nA | 119.9999 µA | 0.010 + 0.020 | 0.040 + 0.025 | 0.050 + 0.025 | 0.0020 + 0.0030 |
1 mA | < 0.11 V | 1 nA | 1.199999 mA | 0.010 + 0.006 | 0.030 + 0.006 | 0.050 + 0.006 | 0.0020 + 0.0005 |
10 mA | < 0.05 V | 10 nA | 11.99999 mA | 0.010 + 0.020 | 0.030 + 0.020 | 0.050 + 0.020 | 0.0020 + 0.0020 |
100 mA | < 0.5 V | 0.1 µA | 119.9999 mA | 0.010 + 0.004 | 0.030 + 0.005 | 0.050 + 0.005 | 0.0020 + 0.0005 |
1 A | < 0.7 V | 1 µA | 1.199999 A | 0.050 + 0.006 | 0.080 + 0.010 | 0.100 + 0.010 | 0.0050 + 0.0010 |
3 A | < 2.0 V | 1 µA | 3.15000 A | 0.180 + 0.020 | 0.200 + 0.020 | 0.200 + 0.020 | 0.0050 + 0.0020 |
10 A(3) | < 0.5 V | 10 µA | 11.99999 A | 0.050 + 0.010 | 0.120 + 0.010 | 0.120 + 0.010 | 0.0050 + 0.0010 |
(3) 10 A range available on front panel only.
DC Resistance(4)
Range(5) |
Test Current |
Resolution |
Full Scale Reading | Accuracy |
Temperature Coefficient / °C | ||
24 Hours TCAL ± 1 °C | 90 Days TCAL ± 5 °C | 1 Year TCAL ± 5 °C | |||||
10 Ω | 10 mA | 10 µΩ | 11.99999 Ω | 0.0050 + 0.0040 | 0.0080 + 0.0060 | 0.0100 + 0.0080 | 0.0006 + 0.0008 |
100 Ω | 10 mA | 0.1 mΩ | 119.9999 Ω | 0.0030 + 0.0020 | 0.0080 + 0.0030 | 0.0100 + 0.0040 | 0.0006 + 0.0005 |
1 kΩ | 1 mA | 1 mΩ | 1.199999 kΩ | 0.0020 + 0.0005 | 0.0080 + 0.0010 | 0.0100 + 0.0010 | 0.0006 + 0.0001 |
10 kΩ | 100 µA | 10 mΩ | 11.99999 kΩ | 0.0020 + 0.0005 | 0.0080 + 0.0010 | 0.0100 + 0.0010 | 0.0006 + 0.0001 |
100 kΩ(6) | 50 µA | 100 mΩ | 119.9999 kΩ | 0.0020 + 0.0005 | 0.0080 + 0.0010 | 0.0100 + 0.0010 | 0.0006 + 0.0001 |
1 MΩ | 5 µA | 1 Ω | 1.199999 MΩ | 0.0020 + 0.0010 | 0.0080 + 0.0010 | 0.0100 + 0.0010 | 0.0010 + 0.0002 |
10 MΩ | 500 nA | 10 Ω | 11.99999 MΩ | 0.0150 + 0.0010 | 0.0200 + 0.0010 | 0.0400 + 0.0010 | 0.0030 + 0.0004 |
100 MΩ | 500 nA || 10 MΩ | 100 Ω | 119.9999 MΩ | 0.3000 + 0.0100 | 0.8000 + 0.0100 | 0.8000 + 0.0100 | 0.1500 + 0.0002 |
(4) Specifications apply to 2-wire and 4-wire resistance measurements using NULL for offset. Without NULL, add 0.2 Ω additional error for 2-wire measurements.
(5) 20% overrange on all ranges
(6) To eliminate noise interference, it is recommended to use shielded test cables for resistance measurements above 100 kΩ.
Capacitance
Range(1) |
Full Scale Reading | Accuracy |
Temperature Coefficient / °C | ||
24 Hours TCAL ± 1 °C | 90 Days TCAL ± 5 °C | 1 Year TCAL ± 5 °C | |||
1 nF | 1.199999 nF | 0.5 + 0.10 | 0.5 + 0.40 | 1.0 + 0.50 | 0.02 + 0.001 |
10 nF | 11.99999 nF | 0.2 + 0.05 | 0.5 + 0.10 | 1.0 + 0.10 | 0.02 + 0.001 |
100 nF | 119.9999 nF | 0.2 + 0.05 | 0.5 + 0.10 | 1.0 + 0.10 | 0.02 + 0.001 |
1 µF | 1.199999 µF | 0.2 + 0.05 | 0.5 + 0.05 | 1.0 + 0.10 | 0.02 + 0.001 |
10 µF | 11.99999 µF | 0.2 + 0.05 | 0.5 + 0.05 | 1.0 + 0.10 | 0.02 + 0.001 |
100 µF | 119.9999 µF | 0.2 + 0.05 | 0.5 + 0.05 | 1.0 + 0.10 | 0.02 + 0.001 |
1 mF | 1.199999 mF | 0.2 + 0.05 | 0.5 + 0.05 | 1.0 + 0.20 | 0.02 + 0.001 |
10 mF | 11.99999 mF | 0.5 + 0.20 | 0.5 + 0.30 | 1.0 + 0.50 | 0.02 + 0.001 |
(1) 20% overrange on all ranges.
True RMS AC Voltage(2)(3)
100 mV, 1 V, 10 V, 100 V, and 750 V ranges
Frequency | 24 Hours TCAL ± 1 °C | 90 Days TCAL ± 5 °C | 1 Year TCAL ± 5 °C | Temperature Coefficient / °C |
3 Hz to 5 Hz | 1.00 + 0.02 | 1.00 + 0.02 | 1.00 + 0.03 | 1.00 + 0.03 |
5 Hz to 10 Hz | 0.35 + 0.02 | 0.35 + 0.03 | 0.35 + 0.03 | 0.35 + 0.03 |
10 Hz to 20 kHz | 0.04 + 0.02 | 0.05 + 0.03 | 0.06 + 0.03 | 0.07 + 0.03 |
20 kHz to 50 kHz | 0.10 + 0.04 | 0.11 + 0.05 | 0.12 + 0.05 | 0.13 + 0.05 |
50 kHz to 100 kHz | 0.55 + 0.08 | 0.60 + 0.08 | 0.60 + 0.08 | 0.60 + 0.08 |
100 kHz to 300 kHz | 4.00 + 0.50 | 4.00 + 0.50 | 4.00 + 0.50 | 4.00 + 0.50 |
Range(4) |
Resolution | Full Scale Reading |
100 mV | 1 µV | 119.9999 mV |
1 V | 10 µV | 1.199999 V |
10 V | 100 µV | 11.99999 V |
100 V | 1 mV | 119.9999 V |
750 V(5) | 10 mV | 787.500 V |
(2) Valid for sine wave input > 0.3% of range and greater than 1 mVrms.
(3) Three filters are available for low-frequency performance: 3 Hz, 20 Hz, and 200 Hz. No additional errors will occur for frequencies above these filter settings.
(4) 20% overrange on all ranges except 750 V range which can only test 5% overrange.
(5) 750 V range is limited to 8 x 107 Volt-Hz.
Continuity
Range |
Test Current | Accuracy |
Temperature Coefficient / °C | ||
24 Hours TCAL ± 1°C | 90 Days TCAL ± 5°C | 1 year TCAL ± 5°C | |||
1 kΩ | 1 mA | 0.002 + 0.030 | 0.008 + 0.030 | 0.010 + 0.030 | 0.0010 + 0.0020 |
Diode(6)
Range |
Test Current | Accuracy |
Temperature Coefficient / °C | ||
24 Hours TCAL ± 1°C | 90 Days TCAL ± 5°C | 1 year TCAL ± 5°C | |||
5 V | 1 mA | 0.002 + 0.030 | 0.008 + 0.030 | 0.010 + 0.030 | 0.0010 + 0.0020 |
(6) Applicable to the voltage measured at the input terminal. A 1 mA test current is typical. Changes in the current source will cause variations in voltage drop across the diode junction.
True RMS AC Current(1)(2)
Range(3), Burden Voltage |
Frequency | Accuracy |
Temperature Coefficient / °C | ||
24 Hours TCAL ± 1 °C | 90 Days TCAL ± 5 °C | 1 Year TCAL ± 5 °C | |||
100 µA, < 0.011 V 1 mA, < 0.11 V 10 mA, < 0.05 V 100 mA, < 0.5 V | 3 Hz to 5 Hz | 1.00 + 0.02 | 1.00 + 0.02 | 1.00 + 0.04 | 0.100 + 0.006 |
5 Hz to 10 Hz | 0.30 + 0.02 | 0.30 + 0.03 | 0.30 + 0.04 | 0.035 + 0.006 | |
10 Hz to 5 kHz | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.015 + 0.006 | |
5 kHz to 10 kHz | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.030 + 0.006 | |
1 A, < 0.7 V | 3 Hz to 5 Hz | 1.00 + 0.02 | 1.00 + 0.02 | 1.00 + 0.04 | 0.100 + 0.006 |
5 Hz to 10 Hz | 0.30 + 0.02 | 0.30 + 0.03 | 0.30 + 0.04 | 0.035 + 0.006 | |
10 Hz to 5 kHz | 0.10 + 0.04 | 0.10 + 0.04 | 0.10 + 0.04 | 0.015 + 0.006 | |
5 kHz to 10 kHz | 0.30 + 0.04 | 0.30 + 0.04 | 0.30 + 0.04 | 0.030 + 0.006 | |
3 A, < 2.0 V | 3 Hz to 5 Hz | 1.00 + 0.02 | 1.00 + 0.02 | 1.00 + 0.04 | 0.100 + 0.006 |
5 Hz to 10 Hz | 0.35 + 0.02 | 0.35 + 0.03 | 0.35 + 0.04 | 0.035 + 0.006 | |
10 Hz to 5 kHz | 0.23 + 0.04 | 0.23 + 0.04 | 0.23 + 0.04 | 0.015 + 0.006 | |
5 kHz to 10 kHz | 0.30 + 0.04 | 0.30 + 0.04 | 0.30 + 0.04 | 0.030 + 0.006 | |
10 A(4), < 0.5 V | 3 Hz to 5 Hz | 1.00 + 0.02 | 1.00 + 0.02 | 1.00 + 0.04 | 0.100 + 0.006 |
5 Hz to 10 Hz | 0.30 + 0.02 | 0.35 + 0.03 | 0.30 + 0.04 | 0.035 + 0.006 | |
10 Hz to 5 kHz | 0.15 + 0.04 | 0.15 + 0.04 | 0.15 + 0.04 | 0.015 + 0.006 | |
5 kHz to 10 kHz | 0.15 + 0.04 | 0.15 + 0.04 | 0.15 + 0.04 | 0.030 + 0.006 |
(1) Three filters are available for low-frequency performance: 3 Hz, 20 Hz, and 200 Hz. No additional errors will occur for frequencies above these filter settings.
(2) Valid for sinewave input > 1% of range and greater than 10 µA AC.
(3) 20% overrange on all ranges except 3 A range which can only test 5% overrange.
(4) 10 A range available on front panel only.
Specifications
General | ||
AC Input | 110 V/220 V ±10%, 50/60 Hz | |
Power Consumption | ≤ 30 VA | |
I/O Interface | RS232, USB host, USB (USBTMC-compliant) device, LAN, GPIB (5492CGPIB and 5493CGPIB only) | |
Temperature | Operating | 32 °F to 104 °F (0 °C to 40 °C) |
Storage | -40 ºF to 158 ºF (-40 ºC to 70 ºC) | |
Humidity | Indoor use, ≤ 95 % | |
Safety | EN 6110-1:2010, Low Voltage Directive (LVD) 2014/35/EU | |
Electromagnetic Compatibility | EN 61326-1:2013, EMC directive 2014/30/EU | |
Dimensions (W x H x D) | 8.85” x 3.93” x 13.97” (225 mm x 100 mm x 355 mm) | |
See Data Sheet