BK Precision 5493C 6 1/2 Digit Digital Benchtop Multimeter

SKU
5493C
Manufacturer:
B&K Precision
B&K Precision

Key Features:

  • 12 measurement functions: DCV, ACV, DCI, ACI, 2 and 4-wire resistance, capacitance, frequency, diode, temperature, continuity test, DCV ratio
  • High-contrast 4.3-inch full color LCD
  • Measurement speed up to 1000 readings/s
  • DCV basic accuracy up to 35 ppm (0.0035%)
  • Histogram, bar graph, trend chart, and numeric display options
  • Display Min, Max, Peak-to-Peak, Average, Standard Deviation statistics simultaneously
  • Null, dB, dBm, %, mx+b, and run/stop math functions
  • AC frequency measurements to 1MHz
  • Internal storage buffer for up to 10,000 readings
  • 2 and 4-wire resistance measurements down to 10 Ω
  • Dual measurement display
  • Rear input terminals for rack use
  • Standard USB (USBTMC-compliant), LAN, and RS232 interfaces, supporting SCPI commands. GPIB option
  • Front panel USB host port to export measurements, screen captures and save/recall settings
  • LabVIEW driver and operating software for remote instrument control, measurement monitoring, and data logging included
Product Options:
$1,115.00

Secured and trusted checkout with

The BK Precision 5493C is a 6 1/2 bench multimeter that is engineered for accuracy, repeatability and ease-of-use. Includes rear input terminals for rack use.

Specifications are based on the following conditions:

-  One year calibration cycle

-  Calibration temperature TCAL = 23 °C

-  Accuracy specifications: ± (% of reading + % of range), valid for TCAL ±5 ºC

-  All 24 hours accuracy specifications are typical

-  Add temperature  coefficient for every ºC outside TCAL ±5 ºC

-  For DC: The technical specifications are valid after a warm-up time of 60 minutes, with the integration time set to 10 or 100 NPLC, and auto zero enabled.

-  For AC: The technical specifications are valid after a warm-up time of 60 minutes, slow AC filtering enabled, using a sine wave input signal.

   

DC Voltage

 

 

Range(1)

 

 

Resolution

 

Full Scale

Reading

Accuracy

 

Temperature

Coefficient / °C

24 Hours(2)

TCAL ± 1 °C

90 Days

TCAL ± 5 °C

1 Year

TCAL ± 5 °C

100 mV

0.1 µV

119.9999 mV

0.0030 + 0.0030

0.0040 + 0.0035

0.0050 + 0.0035

0.0005 + 0.0005

1 V

1 µV

1.199999  V

0.0020 + 0.0006

0.0030 + 0.0007

0.0040 + 0.0007

0.0005 + 0.0001

10 V

10 µV

11.99999  V

0.0015 + 0.0004

0.0020 + 0.0005

0.0035 + 0.0005

0.0005 + 0.0001

100  V

100  µV

119.9999  V

0.0020 + 0.0006

0.0035 + 0.0006

0.0045 + 0.0006

0.0005 + 0.0001

1000  V

1 mV

1050.000  V

0.0020 + 0.0006

0.0035 + 0.0010

0.0045 + 0.0010

0.0005 + 0.0001

(1) 20% overrange on all ranges except 1000 V range which can only test 5% overrange.

(2) Relative to calibration standards.

   

DC Current

 

 

Range

  

Burden

Voltage

 

 

Resolution

 

Full Scale

Reading

Accuracy

 

Temperature

Coefficient / °C

24 Hours

TCAL ± 1 °C

90 Days

TCAL ± 5 °C

1 Year

TCAL ± 5 °C

100 µA

< 0.011 V

0.1 nA

119.9999 µA

0.010 + 0.020

0.040 + 0.025

0.050 + 0.025

0.0020 + 0.0030

1 mA

< 0.11 V

1 nA

1.199999 mA

0.010 + 0.006

0.030 + 0.006

0.050 + 0.006

0.0020 + 0.0005

10 mA

< 0.05 V

10 nA

11.99999 mA

0.010 + 0.020

0.030 + 0.020

0.050 + 0.020

0.0020 + 0.0020

100 mA

< 0.5 V

0.1 µA

119.9999 mA

0.010 + 0.004

0.030 + 0.005

0.050 + 0.005

0.0020 + 0.0005

1 A

< 0.7 V

1 µA

1.199999 A

0.050 + 0.006

0.080 + 0.010

0.100 + 0.010

0.0050 + 0.0010

3 A

< 2.0 V

1 µA

3.15000 A

0.180 + 0.020

0.200 + 0.020

0.200 + 0.020

0.0050 + 0.0020

10 A(3)

< 0.5 V

10 µA

11.99999 A

0.050 + 0.010

0.120 + 0.010

0.120 + 0.010

0.0050 + 0.0010

(3) 10 A range available on front panel only.

 

DC Resistance(4)

 

 

Range(5)

 

Test

Current

 

 

Resolution

 

Full Scale

Reading

Accuracy

 

Temperature

Coefficient / °C

24 Hours

TCAL ± 1 °C

90 Days

TCAL ± 5 °C

1 Year

TCAL ± 5 °C

10 Ω

10 mA

10 µΩ

11.99999 Ω

0.0050 + 0.0040

0.0080 + 0.0060

0.0100 + 0.0080

0.0006 + 0.0008

100 Ω

10 mA

0.1 mΩ

119.9999 Ω

0.0030 + 0.0020

0.0080 + 0.0030

0.0100 + 0.0040

0.0006 + 0.0005

1 kΩ

1 mA

1 mΩ

1.199999 kΩ

0.0020 + 0.0005

0.0080 + 0.0010

0.0100 + 0.0010

0.0006 + 0.0001

10 kΩ

100 µA

10 mΩ

11.99999 kΩ

0.0020 + 0.0005

0.0080 + 0.0010

0.0100 + 0.0010

0.0006 + 0.0001

100 kΩ(6)

50 µA

100 mΩ

119.9999 kΩ

0.0020 + 0.0005

0.0080 + 0.0010

0.0100 + 0.0010

0.0006 + 0.0001

1 MΩ

5 µA

1 Ω

1.199999 MΩ

0.0020 + 0.0010

0.0080 + 0.0010

0.0100 + 0.0010

0.0010 + 0.0002

10 MΩ

500 nA

10 Ω

11.99999 MΩ

0.0150 + 0.0010

0.0200 + 0.0010

0.0400 + 0.0010

0.0030 + 0.0004

100 MΩ

500 nA || 10 MΩ

100 Ω

119.9999 MΩ

0.3000 + 0.0100

0.8000 + 0.0100

0.8000 + 0.0100

0.1500 + 0.0002

(4) Specifications apply to 2-wire and 4-wire resistance measurements using NULL for offset.  Without  NULL, add 0.2 Ω additional error for 2-wire measurements.

(5) 20% overrange on all ranges

(6) To eliminate noise interference, it is recommended to use shielded test cables for resistance measurements above 100 kΩ.

 

Capacitance

 

 

Range(1)

 

Full Scale

Reading

Accuracy

 

Temperature

Coefficient / °C

24 Hours

TCAL ± 1 °C

90 Days

TCAL ± 5 °C

1 Year

TCAL ± 5 °C

1 nF

1.199999 nF

0.5 + 0.10

0.5 + 0.40

1.0 + 0.50

0.02 + 0.001

10 nF

11.99999 nF

0.2 + 0.05

0.5 + 0.10

1.0 + 0.10

0.02 + 0.001

100 nF

119.9999 nF

0.2 + 0.05

0.5 + 0.10

1.0 + 0.10

0.02 + 0.001

1 µF

1.199999 µF

0.2 + 0.05

0.5 + 0.05

1.0 + 0.10

0.02 + 0.001

10 µF

11.99999 µF

0.2 + 0.05

0.5 + 0.05

1.0 + 0.10

0.02 + 0.001

100 µF

119.9999 µF

0.2 + 0.05

0.5 + 0.05

1.0 + 0.10

0.02 + 0.001

1 mF

1.199999 mF

0.2 + 0.05

0.5 + 0.05

1.0 + 0.20

0.02 + 0.001

10 mF

11.99999 mF

0.5 + 0.20

0.5 + 0.30

1.0 + 0.50

0.02 + 0.001

(1) 20% overrange on all ranges.

 

True RMS AC Voltage(2)(3)

100 mV, 1 V, 10 V, 100 V, and 750 V ranges

 

Frequency

24 Hours

TCAL ± 1 °C

90 Days

TCAL ± 5 °C

1 Year

TCAL ± 5 °C

Temperature

Coefficient / °C

3 Hz to 5 Hz

1.00 + 0.02

1.00 + 0.02

1.00 + 0.03

1.00 + 0.03

5 Hz to 10 Hz

0.35 + 0.02

0.35 + 0.03

0.35 + 0.03

0.35 + 0.03

10 Hz to 20 kHz

0.04 + 0.02

0.05 + 0.03

0.06 + 0.03

0.07 + 0.03

20 kHz to 50 kHz

0.10 + 0.04

0.11 + 0.05

0.12 + 0.05

0.13 + 0.05

50 kHz to 100 kHz

0.55 + 0.08

0.60 + 0.08

0.60 + 0.08

0.60 + 0.08

100 kHz to 300 kHz

4.00 + 0.50

4.00 + 0.50

4.00 + 0.50

4.00 + 0.50

    

 

Range(4)

 

Resolution

Full Scale

Reading

100 mV

1 µV

119.9999 mV

1 V

10 µV

1.199999  V

10 V

100 µV

11.99999  V

100 V

1 mV

119.9999  V

750 V(5)

10 mV

787.500 V

(2) Valid for sine wave input > 0.3% of range and greater than 1 mVrms.

(3) Three filters are available for low-frequency performance: 3 Hz, 20 Hz, and 200 Hz. No additional errors will occur for frequencies above these filter settings.

(4) 20% overrange on all ranges except 750 V range which can only test 5% overrange.

(5) 750 V range is limited to 8 x 107 Volt-Hz.

   

Continuity

 

 

Range

 

Test

Current

Accuracy

 

Temperature

Coefficient / °C

24 Hours

TCAL ± 1°C

90 Days

TCAL ± 5°C

1 year

TCAL ± 5°C

1 kΩ

1 mA

0.002 + 0.030

0.008 + 0.030

0.010 + 0.030

0.0010 + 0.0020

  

Diode(6)

 

 

Range

 

Test

Current

Accuracy

 

Temperature

Coefficient / °C

24 Hours

TCAL ± 1°C

90 Days

TCAL ± 5°C

1 year

TCAL ± 5°C

5 V

1 mA

0.002 + 0.030

0.008 + 0.030

0.010 + 0.030

0.0010 + 0.0020

(6) Applicable to the voltage measured at the input terminal. A 1 mA  test current is typical. Changes in the current source will cause variations in voltage drop across the diode junction.

  

True RMS AC Current(1)(2) 

 

 

Range(3), Burden Voltage

 

 

Frequency

Accuracy

 

Temperature

Coefficient / °C

24 Hours

TCAL ± 1 °C

90 Days

TCAL ± 5 °C

1 Year

TCAL ± 5 °C

 

100 µA, < 0.011  V

1 mA, < 0.11 V

10 mA, < 0.05 V

100 mA, < 0.5 V

3 Hz to 5 Hz

1.00 + 0.02

1.00 + 0.02

1.00 + 0.04

0.100 + 0.006

5 Hz to 10 Hz

0.30 + 0.02

0.30 + 0.03

0.30 + 0.04

0.035 + 0.006

10 Hz to 5 kHz

0.10 + 0.04

0.10 + 0.04

0.10 + 0.04

0.015 + 0.006

5 kHz to 10 kHz

0.10 + 0.04

0.10 + 0.04

0.10 + 0.04

0.030 + 0.006

 

 

1 A, < 0.7 V

3 Hz to 5 Hz

1.00 + 0.02

1.00 + 0.02

1.00 + 0.04

0.100 + 0.006

5 Hz to 10 Hz

0.30 + 0.02

0.30 + 0.03

0.30 + 0.04

0.035 + 0.006

10 Hz to 5 kHz

0.10 + 0.04

0.10 + 0.04

0.10 + 0.04

0.015 + 0.006

5 kHz to 10 kHz

0.30 + 0.04

0.30 + 0.04

0.30 + 0.04

0.030 + 0.006

 

 

3 A, < 2.0 V

3 Hz to 5 Hz

1.00 + 0.02

1.00 + 0.02

1.00 + 0.04

0.100 + 0.006

5 Hz to 10 Hz

0.35 + 0.02

0.35 + 0.03

0.35 + 0.04

0.035 + 0.006

10 Hz to 5 kHz

0.23 + 0.04

0.23 + 0.04

0.23 + 0.04

0.015 + 0.006

5 kHz to 10 kHz

0.30 + 0.04

0.30 + 0.04

0.30 + 0.04

0.030 + 0.006

 

 

10 A(4), < 0.5 V

3 Hz to 5 Hz

1.00 + 0.02

1.00 + 0.02

1.00 + 0.04

0.100 + 0.006

5 Hz to 10 Hz

0.30 + 0.02

0.35 + 0.03

0.30 + 0.04

0.035 + 0.006

10 Hz to 5 kHz

0.15 + 0.04

0.15 + 0.04

0.15 + 0.04

0.015 + 0.006

5 kHz to 10 kHz

0.15 + 0.04

0.15 + 0.04

0.15 + 0.04

0.030 + 0.006


(1) Three filters are available for low-frequency performance: 3 Hz, 20 Hz, and 200 Hz. No additional errors will occur for frequencies above these filter settings.
(2) Valid for sinewave input > 1% of range and greater  than 10 µA AC.
(3) 20% overrange on all ranges except 3 A range which can only test 5% overrange.
(4) 10 A range available on front panel only.

 

Specifications

General

AC Input

110 V/220  V ±10%,  50/60 Hz

Power Consumption

≤ 30 VA

I/O Interface

RS232, USB host, USB (USBTMC-compliant)  device, LAN, GPIB (5492CGPIB and 5493CGPIB only)

 

Temperature

Operating

32 °F to 104 °F (0 °C to 40 °C)

Storage

-40 ºF to 158 ºF (-40 ºC to 70 ºC)

Humidity

Indoor use, ≤ 95 %

Safety

EN 6110-1:2010, Low Voltage Directive  (LVD) 2014/35/EU

Electromagnetic Compatibility

EN 61326-1:2013, EMC directive  2014/30/EU

Dimensions  (W x H x D)

8.85”  x 3.93” x 13.97” (225 mm x 100 mm x 355 mm)

See Data Sheet

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account

Products In This Series

As low as

Now $890.00

As low as

Now $1,110.00

As low as

Now $1,335.00