BK Precision 5335C Single-Phase AC / DC Power Meter

SKU
5335C
Manufacturer:
B&K Precision
B&K Precision

Key Features:

  • 600 Vrms ( Cat II ) and 20 Arms direct input ranges
  • Frequency ranges DC, 0.5 Hz to 100 kHz
  • 0.1% basic accuracy for voltage and current measurements
  • 4.3-inch color LCD (TFT)
  • Simultaneously measure and display up to 12 different measurement parameters
  • Capture inrush current, and voltage surge with the peak function
  • Harmonic measurements to the 50th order
  • Integration function with automatic range switching
  • Ability to measure electric energy which is produced or consumed
  • Pre-compliance testing according to IEC/EN62000-3-2 / 4-7
  • Standard USB (USBTMC-compliant), RS232, and LAN interfaces
  • Line and frequency filter capability for reducing unwanted signal noise
  • Optional universal breakout box to simplify connection between power meter and DUT
Product Options:
$2,845.00

Secured and trusted checkout with

The BK Precision 5335C is a single-phase AC / DC power meter that is for measuring and analyzing power consumption and power quality parameters quickly and accurately. The 5335C measures all AC and DC parameters, including power, current, voltage, power factor, frequency, and phase. This meter supports power measurements up to 600 Vrms and 20 Arms, with a bandwidth up to of 100 kHz.

Specifications are subject to the following conditions
Temperature: 23±5° C, humidity: 30 to 75% RH.
Warm-up time: 30 minutes

General Measurement Specifications




Basic measurements

Voltage, Current
Peak to peak, Maximum, Minimum, Average_rms, Average_rectified, DC,
Crest factor (current), Inrush (current)
PowerReal, Apparent, Reactive, DC, Power factor
TimeFrequency, Phase
IntegrationTotal power, Total current, Maximum power, Minimum power



Harmonic measurements

Type
Current, Voltage, Real power, Apparent power, Reactive power, Power factor, Phase, Percentage of total (Current, Voltage, Power)
RangeDC up to 50th order
Max. Frequency100 kHz
Input bandwidthDC, 0.5 Hz to 100 kHz
Measurement methodDigital sampling
A/D ConverterSimultaneous conversion of voltage and current inputs, Resolution: 18-bit, Maximum conversion rate: 10 μs
Line filterSelect OFF or ON (cutoff frequency at 500 Hz)
Peak (max,min)Voltage, current, or power
Input voltage continuous max.1.5 kV-peak or 1 kV-RMS, whichever is less
Input voltage transient (<1s) max.2 kV-peak or 1.5 kV-RMS, whichever is less
Input voltage common-mode max.600 Vrms
Voltage input impedance2 MΩ + 13 pF in parallel (typical)

Current input
impedance (typical)
5 mA to 200 mA range505 mΩ + 0.1 μH
0.5 A to 20 A range5 mΩ + 0.1 μH
Sensor input20 kΩ (50 mV to 2 V)
100 kΩ (2.5 V to 10 V)


Input current continuous max.
5 mA to 200 mA range30 A-peak or 20 A-RMS, whichever is less
0.5 A to 20 A range100 A-peak or 30 A-RMS, whichever is less
Sensor inputPeak value less than or equal to 5 times the rated range


Input current transient (<1s) max.
5 mA to 200 mA range30 A-peak or 20 A-RMS, whichever is less
0.5 A to 20 A range150 A-peak or 40 A-RMS, whichever is less
Sensor inputPeak value less than or equal to 10 times the rated range
Voltage Measurement Accuracy and Ranges
RangesCF=3: 15 V, 30 V, 60 V, 150 V, 300 V, 600 V
CF=6: 7.5 V, 15 V, 30 V, 75 V, 150 V, 300 V

Accuracy² (line, frequency, &
digital filter set to off)
DC to 1 kHz±(0.1% + 0.2% F.S.)
1 kHz < f ≤ 10 kHz±((0.07 f¹ )% + 0.3% F.S.)
10 kHz < f ≤ 100 kHz±(0.5% + 0.5% F.S.) ± [{0.04 × (f¹ - 10)}% ]



Temperature
coefficient
For temperature changes after zero-level compensation or range change
+ 0.02% F.S. /°C to the DC voltage accuracy
Influence of self-generated heat caused by voltage input (U is the voltage reading (V))+ 0.0000001 × U²% to the AC voltage accuracy
+ 0.0000001 × U² % + 0.0000001 × U² % F.S. to DC current accuracy

¹ Input signal frequency in kHz
² Input waveform: Sine wave crest factor: 3, common-mode voltage: 0 V, power factor: 1
Frequency filter: Turn on when measuring ≤ 200 Hz

 

Current Measurement Accuracy and Ranges


Direct input range
CF= 3:5 mA, 10 mA, 20 mA, 50 mA, 100 mA, 200 mA, 0.5 A, 1 A, 2 A, 5 A, 10 A, 20 A
CF= 6:2.5 mA, 5 mA, 10 mA, 25 mA, 50 mA, 100 mA, 250 mA, 0.5 A, 1 A, 2.5 A, 5 A, 10 A


Sensor input range
External 1CF = 3: 2.5 V, 5 V, 10 V
CF = 6: 1.25 V, 2.5 V, 5 V
External 2CF= 3: 50 mV, 100 mV, 200 mV, 500 mV, 1 V, 2 V
CF= 6: 25 mV, 50 mV, 100 mV, 250 mV, 500 mV, 1 V


Accuracy ² (line, frequency, & digital filter set to off)
DC to 1 kHz±(0.1% + 0.2% F.S.)
1 kHz < f ≤ 10 kHz±{(0.07 f ¹)% + 0.3% F.S.}
10 kHz < f ≤ 100 kHz±(0.5% + 0.5% F.S.) ± [{0.04×(f ¹ -10)}%]



Temperature
coefficient
2.5 to 200 mA5 μA/ °C (after zero-level compensation, or range change)
500 mA to 20 A500 μA/ °C (after zero-level compensation, or range change)
Influence of internal sensor self-heating+ 0.00013 × I² % of reading to the AC current accuracies + 0.00013 × I² % of reading + 0.004 × I² mA (0.5 to 20 A) or
0.00013 × I² % of reading + 0.00004 × I² mA (2.5 to 200 mA), add to the DC current accuracy specifications
Power Measurement Accuracy





Real power
accuracy ² ³ (CF= 3)4
DC±(0.1% + 0.2% F.S.)
0.5 Hz ≤ f < 45 Hz±(0.3% + 0.2% F.S.)
45 Hz ≤ f ≤ 66 Hz±(0.1% + 0.1% F.S.)
66 Hz < f ≤ 1 kHz±(0.2% + 0.2% F.S.)
1 kHz < f ≤ 10 kHz±(0.1% + 0.3% F.S.) ± [{0.067×(f-1)}%]
10 kHz < f ≤ 100 kHz±(0.5% + 0.5% F.S.) ± [{0.09×(f-10)}%]
Apparent power (S)Voltage accuracy + current accuracy
Reactive power (Q)Apparent power accuracy + ( √1.0004-PF²) - ( √1-PF²) × 100%
Power factor (PF)±[(PF–PF/1.0002) + abs(cosØ - cos{Ø+sin-1(influence from the power factor when PF=0%/100)})] ± 1 digit when voltage and current are at the measurement range rated input
Phase angle (Ф)±[abs(Ø - cos-1(PF/1.0002)) + sin-1{(influence from the power factor when PF=0%)/100}] deg ± 1 digit when voltage and current are at the measurement range rated input
Temperature coefficientSame as the temperature coefficient for voltage and current
Frequency Measurement Accuracy
Frequency
measurement
range
Data update interval0.1 s0.25 s0.5 s1 s2 s5 s
Measurement range25 Hz ≤ f ≤ 100 kHz10 Hz ≤ f ≤ 100 kHz5 Hz ≤ f ≤ 100 kHz2.5 Hz ≤ f ≤ 100 kHz1.5 Hz ≤ f ≤ 50 kHz0.5 Hz ≤ f ≤ 20 kHz

Accuracy

±0.06%
(CF 3 and signal <30% F.S.) or,
(CF 6 and signal <60% F.S.), and
≤ 200 Hz with frequency filter on
Frequency filter500 Hz low-pass

¹ Input signal frequency in kHz
² Input waveform: Sine wave crest factor: 3, common-mode voltage: 0 V, power factor: 1
Frequency filter: Turn on when measuring ≤ 200 Hz
³ When power factor (PF)=0 (apparent power (S)):
           ±0.2% of S when 45 Hz ≤ f ≤ 66 Hz
           ±{(0.2+0.2×f)% of S} when 0.066 ≤ f ≤ 100 kHz
When 0<PF<1(phase angle (Ф)):
(power reading ) × [(power reading error %) + (power range %) × (power range/indicated apparent power value) + {tanФ× (influence when PF=o)%}]
When the line filter is turned ON:
           45 to 66 Hz: Add 0.3% of reading
           <45 Hz: Add 1% of reading
4 Accuracy when the crest factor is set to 6, the accuracy is obtained by doubling specified accuracies

 

Harmonic Measurement Parameters
Measurement methodPLL synchronization
Frequency rangePLL frequency source range 10 Hz to 1.2 kHz (typical)
FFT data length1024
Window functionRectangle
Fundamental frequency (Fund. freq.)10 Hz to 75 Hz75 Hz to 150 Hz150 Hz to 300 Hz300 Hz to 600 Hz600 Hz to 1200 Hz
Sample rate(Fund. freq.) x 1024(Fund. freq.) x 512(Fund. freq.) x 256(Fund. freq.) x 128(Fund. freq.) x 64
Window width124816
Upper limit of analysis orders50321684
Harmonic Measurement Accuracy (when line filter is off)
Frequency10 Hz ≤ f < 45 Hz45 Hz ≤ f ≤ 440 Hz440 Hz < f ≤ 1 kHz1 kHz < f ≤ 2.5 kHz2.5 kHz < f ≤ 5 kHz
Voltage and current±0.15% ± 0.35% F.S.±0.15% ± 0.35% F.S.±0.20% ± 0.35% F.S.±0.80% ± 0.45% F.S.3.05% ± 0.45% F.S.
Power±0.15% ± 0.50% F.S.±0.20% ± 0.50% F.S.±0.40% ± 0.50% F.S.1.56% ± 0.60% F.S.5.77% ± 0.60% F.S.
Oscilloscope Function
Channels2
MeasurementVoltage and current
Bandwidth (-3 dB)10 kHz
Sample rate100 kS/s
Record length300 points/channel
Horizontal scale (Accuracy ±4.0%)500 us, 1 ms, 2 ms, 5 ms, 10 ms, 20 ms, 50 ms, 100 ms, 200 ms, 500 ms
Vertical scale ranges
(Accuracy ±4.0%)
CF 3I: 2.5, 5, 10, 25, 50, 100, 250, 500 mA/div, 1 A, 2.5 A, 5 A, 10 A/div,
U: 7.5, 15, 30, 75, 150, 300 V/div
CF 6I: 5, 10, 20, 50, 100, 200, 500 mA/div, 1 A, 2 A, 5 A, 10 A, 20 A/div,
U: 15, 30, 60, 150, 300, 600 V/div
Maximum input voltage (DC+AC peak)1800 V
Maximum input current (DC+AC peak)60 A
Environmental and Safety
TemperatureOperating: 41 ºF to 104 ºF (5 °C to 40 °C)
Storage: -4 ºF to 122 ºF (-20 °C to 50 °C)
Humidity20% RH to 80% RH (non-condensing)
Electromagnetic compatibilityIEC 61326
SafetyIEC 61010-1, EN 61010-1, Measurement 600 V CAT II
General
Display4.3” TFT-LCD display, 480 x 272
Remote InterfacesUSB (USBTMC-Compliant), RS232, LAN
Power100 to 240 VAC, 50 / 60 Hz
Power Consumption50 VA max.
Dimensions (W x H x D)8.4” x 3.5” x 14” (214.5 mm × 88.2 mm × 354.6 mm)
Weight6.2 lbs (2.8 kg)
Warranty3 Years
Standard AccessoriesGetting started manual, instruction manual (downloadable), AC power cord, USB type A-to-type B cable, certificate of calibration

 