BK Precision 394B 50000 Count Process Multimeter

SKU
394B
Manufacturer:
B&K Precision
B&K Precision

Key Features:

  • Source/Measure/Simulate 0-20 mA and 4-20 mA DC current
  • Simultaneously monitor mA and % of scale
  • Adjustable manual and automatic output current ramp and step modes
  • Built-in 24 V loop power supply for testing process transmitters eliminates the need for an
    external supply
  • HART® mode inserts 250 Ω resistor in series with loop power output when evaluating devices using the HART communication protocol
  • Measurement functions: DCV, ACV, AC+DC, DCI, ACI, resistance, frequency, continuity, diode test
  • True RMS AC and AC+DC measurements
  • 50,000-count, dual display
  • dB, dBm, limits, peak-hold, REL (Δ), MIN, MAX, average math functions
  • HFR (High Frequency Rejection) mode applies a low pass filter for AC measurements (800 Hz cut-off)
  • Frequency measurement to 100 kHz
  • Dirt and water-resistant housing with rubberized protective case
  • Isolated USB interface with operating software for remote data logging
  • CAT III 1000 V / CAT IV 600 V protection
Product Options:
$790.00

Secured and trusted checkout with

The BK Precision 394B is a process multimeter with an 50,000 display count. The 394B current output ranges is 0-20 mA or 4-20 mA and the loop power supply is >24 V.

Specifications are based on the following conditions/assumptions:

  • • Accuracy specifications: ± (% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than 80% RH
    • One year calibration cycle
    • Temperature coefficient is 0.1 x (specified accuracy)/°C for T < 18 °C, T > 28°C
    • AC voltage and AC current specifications are AC coupled, true RMS
    • For non-sinusoidal waveforms:
         - Add 1.0% to AC accuracy specification for Crest Factor 1.4 to 2.0
         - Add 2.5% to AC accuracy specification for Crest Factor 2.0 to 2.5
         - Add 4.0% to AC accuracy specification for Crest Factor 2.5 to 3.0
    • For best accuracy use REL (delta) function to compensate the offsets
    • AC + DC accuracy: AC accuracy + DC accuracy + 1.0%
    • HFR accuracy: AC accuracy + 1.0% for 40 Hz to 400 Hz.
    • Overload protection: AC/DC 1000 V

Voltage

FunctionRangeAccuracy



AC (1)
50.000 mV
500.00 mV
Sine wave:
(0.7 + 20) for 40 Hz to 70 Hz
(1.5 + 40) for 71 Hz to 10 kHz
5.0000 V
50.000 V
500.00 V
1000.0 V(2)
Sine wave:
(0.5 + 20) for 40 Hz to 70 Hz
(1.5 + 40) for 71 Hz to 10 kHz
(3.0 + 80) for 1001 Hz to 10 kHz

DC
50.000 mV0.05 + 30
500.00 mV
5.0000 V
50.000 V
500.00 V
1000.0 V

0.05 + 5

(1) Below 5% of AC range, add 20 digits to accuracy.
(2) The bandwidth of range is 40 Hz to 1 kHz
Notes:
- Input impedance: 10 MΩ, < 100 pF
- Min. resolution: 1 μV in 50 mV range

 

Current

FunctionRangeAccuracy
AC (3)50.000 mA
1.000 A
Sine wave:
(1.0 + 20) for 40 Hz to 70 Hz
(2.0 + 40) for 71 Hz to 10 kHz
DC50.000 mA
1.000 A
0.05 + 5

(3) Below 5% of AC range, add 20 digits to accuracy.
Notes:
- Max. continuous measuring time: 10 minutes at mA input, 1 minute at A input
- Min. rest time: 20 minutes after continuous measuring
- Input impedance: 13 Ω at mA input and 0.1 Ω at A input
- Min. resolution: 1 μA in 50 mA range

 

Resistance

RangeResolutionTest CurrentAccuracy
500.00 Ω 0.01 Ω1 mA0.2 + 30
5.0000 kΩ0.1 Ω100 μA
0.2 + 10
50.000 kΩ1 Ω10 μA
500.00 kΩ10 Ω1 μA0.5 +10
5.0000 MΩ100 Ω100 nA1.0 + 10
50.00 MΩ10 kΩ10 nA2.0 + 10

Notes:
- Max. open circuit voltage: 3.5 V

 

Continuity

RangeResolutionTest CurrentAccuracy
500.00 Ω0.01 Ω1 mA0.1 + 30

Notes:
- Max. open circuit voltage: 3.5 V
- Continuity threshold: < 30 Ω

 

Diode Test

RangeResolutionTest CurrentAccuracy
2.000 V1 mV± 1 mA1.0 + 10

Notes:
- Max. open circuit voltage: ± 3.5 V

 

Frequency

RangeResolutionAccuracy
500.00 Hz0.01 Hz




± 3 digits
5.0000 kHz0.1 Hz
50.000 kHz1 Hz
100.00 kHz10 Hz

Notes:
- Min. frequency: 5 Hz

 

Process Multimeter Functions / Current Output


Range

Accuracy

Resolution
Output Adjustment Modes
RampStep
0 to 20 mA or 4 mA to 20 mA
(overrange up to 24 mA) 
± (0.05 + 5)1 μALinear (slow), 0% to 100% and back to 0% in 40 s
Linear (fast), 0% to 100% and back to 0% in 20 s
25% steps (coarse), 0% to 100%, 15 s for each step
25% steps (fine), 0% to 100%, 5 s for each step

 

Process Multimeter Functions / Loop Power


Range

Accuracy
Drive Capability
Normal250 Ω HART
50 mA± (0.05 + 5)30 V / 1.25 kΩ24 V / 1 kΩ

 

General

Display5 digit / 50,000 count
Measurement Speed10 samples per second
ConnectivityIR-USB
Power4 x 1.5 V AA size batteries
Battery Life (typical)100 hours
Auto Power OffAdjustable up to 20 minutes or never
Low Battery Indicator
OverrangeOL is displayed

Temperature
Operating14 ºF to 122 ºF (-10 °C to 50 °C)
at ≤ 80% relative humidity
Storage-4 ºF to 140 ºF (-20 °C to 60 °C)
SafetyLow Voltage Directive (LVD) 2014/35/EU, EN61010-1,
EN61010-2-30, 600 V CAT IV / 1000 V CAT III
Electromagnetic CompatibilityEMC Directive 2014/30/EU, EN61326-1:2013
Dimensions (W x H x D),
without holster
3.8" x 8.2" x 2" (95 mm x 207 mm x 52 mm)
Weight1.4 lbs (630 g)
Warranty3 Years
Standard AccessoriesTest leads, protective case, optical-isolated USB
cable, magnetic hanging kit, alkaline batteries
Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account

Products In This Series

As low as