We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The BK Precision 393B is a True RMS multimeter with an 100,000 display count. The 393B DCV accuracy is ± (0.015% + 20 digits) and it has 20,000 readings of data log capacity.
Specifications are based on the following conditions/assumptions:
- • Accuracy specifications: ± (% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than 80% RH One year calibration cycle
• Temperature coefficient is 0.1 x (specified accuracy)/°C for T < 18 °C, T > 28°C
• AC Voltage and AC Current specifications are AC coupled, true RMS
• For non-sinusoidal waveforms:
- Add 1.0% to accuracy specification for Crest Factor 1.0 to 2.0
- Add 2.5% to accuracy specification for Crest Factor 2.0 to 2.5
- Add 4.0% to accuracy specification for Crest Factor 2.5 to 3.0
• For best accuracy use REL (delta) function to compensate the offsets
DC Voltage
Model | Range | Resolution | Accuracy |
390B | 40.00 mV | 1 μV | 0.03 + 3 |
400.0 mV | 10 μV | 0.03 + 1 | |
4.000 V | 100 μV | ||
40.00 V | 1 mV | ||
400.0 V | 10 mV | ||
1000 V | 100 mV | ||
391B | 40.000 mV | 1 μV | 0.040 + 40 |
400.00 mV | 10 μV | 0.035 + 20 | |
4.000 V | 100 μV | 0.030 + 20 | |
40.000 V | 1 mV | ||
400.00 V | 10 mV | ||
1000.0 V | 100 mV | ||
393B | 100.000 mV | 1 μV | 0.025 + 40 |
1000.00 mV | 10 μV | 0.020 + 20 | |
10.0000 V | 100 μV | 0.015 + 20 | |
100.000 V | 1 mV | ||
1000.00 V | 10 mV |
DC Current
Model | Range | Resolution | Accuracy |
390B | 40.00 mA | 1 μA | 0.2 + 1 |
400.0 mA | 10 μA | ||
4.000 A | 100 μA | ||
10.00 A | 1 mA | 0.2 + 2 | |
391B | 40.000 mA | 1 μA | 0.2 + 40 |
400.00 mA | 10 μA | ||
4.0000 A | 100 μA | 0.2 + 80 | |
10.000 A | 1 mA | ||
393B | 10.0000 mA | 0.1 μA | 0.1 + 40 |
100.000 mA | 1 μA | ||
10.0000 A | 100 μA | 0.1 + 80 |
- Burden Voltage: 2 mV/mA at mA inputs and 60 mV/A at A inputs
- Max. Continuous Measuring Time: 10 minutes at mA inputs, 1 minute at A inputs
- Min. Rest Time: 20 minutes after continuous measuring
- Overload protection: AC/DC 400 mA at mA inputs, AC/DC 10 A at A inputs
Continuity
Model | Range | Resolution | Accuracy |
390B | 400.0 Ω | 0.1 Ω | 0.2 + 2 |
391B & 393B | 400.00 Ω | 0.01 Ω | 0.2 + 30 |
- Open Circuit Voltage: -1.2 V
- Short Test Current: -0.3 mA
- Threshold: Adjustable 10 Ω to 50 Ω
- Buzzer Response Time: < 10 ms
Diode Test
Model | Range | Resolution | Accuracy |
390B | 2.000 V | 1 mV | 1.5 + 2 |
391B & 393B | 2.0000 V | 100 μV | 1.5 + 20 |
Resistance
Model | Range | Resolution | Accuracy |
390B | 400.0 Ω | 0.1 Ω | 0.2 + 2 |
4.000 kΩ | 1 Ω | 0.2 + 1 | |
40.00 kΩ | 10 Ω | ||
400.0 kΩ | 100 Ω | ||
4.000 MΩ | 1 kΩ | 1.0 + 1 | |
40.00 MΩ | 10 kΩ | 2.0 + 20 | |
391B | 400.00 Ω | 0.01 Ω | 0.2 + 30 |
4.0000 kΩ | 0.1 Ω | ||
40.000 kΩ | 1 Ω | ||
400.00 kΩ | 10 Ω | 0.3 + 30 | |
4.0000 MΩ | 100 Ω | 1.0 + 30 | |
40.000 MΩ | 1 kΩ | 1.5 + 30 | |
393B | 1000.00 Ω | 0.01 Ω | 0.050 + 30 |
10.0000 kΩ | 0.1 Ω | 0.025 + 30 | |
100.000 kΩ | 1 Ω | ||
1000.00 kΩ | 10 Ω | 0.3 + 30 | |
10.0000 MΩ | 100 Ω | 1.0 + 30 | |
40.000 MΩ | 1 kΩ | 1.5 + 30 |
True RMS AC Voltage
Model | Range | Frequency | Accuracy |
390B | 40.00 mV(1), 400.0 mV(1), 4.000 V, 40.00 V, 400.0 V(1), 1000 V(2) | 40 Hz to 70 Hz(3) | 0.5 + 2 |
70 Hz to 1 kHz(3) | 1.5 + 4 | ||
1 kHz to 5 kHz(3) | 3.0 + 4 | ||
5 kHz to 100 kHz(4) (5) | 5.0 + 20 | ||
391B | 40.000 mV, 400.00 mV | 40 Hz to 65 Hz | 1.00 + 50 |
65 Hz to 1 kHz | 3.00 + 50 | ||
1 kHz to 3 kHz | 5.00 + 50 | ||
4.0000 V, 40.000 V | 40 Hz to 45 Hz | 1.50 + 50 | |
45 Hz to 65 Hz | 0.70 + 50 | ||
65 Hz to 1 kHz | 1.50 + 50 | ||
1 kHz to 10 kHz | 3.00 + 50 | ||
10 kHz to 50 kHz | 5.00 + 50 | ||
50 kHz to 100 kHz | 10.0 + 50 | ||
400.00 V, 1000.0 V | 40 Hz to 45 Hz | 1.50 + 50 | |
45 Hz to 65 Hz | 0.70 + 50 | ||
65 Hz to 1 kHz | 1.50 + 50 | ||
393B | 100.000 mV, 1000.00 mV | 40 Hz to 65 Hz | 0.70 + 50 |
66 Hz to 1 kHz | 1.50 + 50 | ||
1 kHz to 3 kHz | 3.00 + 50 | ||
10.0000 V, 100.000 V | 40 Hz to 45 Hz | 1.00 + 50 | |
45 Hz to 65 Hz | 0.40 + 50 | ||
65 Hz to 1 kHz | 1.00 + 50 | ||
1 kHz to 10 kHz | 2.00 + 50 | ||
10 kHz to 20 kHz | 3.00 + 50 | ||
20 kHz to 50 kHz | 5.00 + 50 | ||
50 kHz to 100 kHz | 10.0 + 50 | ||
1000.00 V | 40 Hz to 45 Hz | 1.00 + 50 | |
45 Hz to 65 Hz | 0.40 + 50 | ||
65 Hz to 1 kHz | 1.00 + 50 |
(1) The bandwidth is 40 Hz to 5 kHz
(2) The bandwidth is 40 Hz to 1 kHz
(3) Below 10% of range, add 2 digits to accuracy
(4) Below 10% of range, add 10 digits to accuracy, < 50 kHz
(5) Below 10% of range, add 20 digits to accuracy, > 50 kHz
True RMS AC Current
Model | Range | Frequency | Accuracy |
390B | 40.00 mA, 400.0 mA, 4.000 A, 10.00 A | 40 Hz to 70 Hz | 0.8 + 2 |
70 Hz to 1 kHz | 2.0 + 4 | ||
1 kHz to 10 kHz | |||
391B | 40.000 mA, 400.00 mA, 4.0000 A, 10.000 A | 40 Hz to 65 Hz | 0.8 + 80 |
65 Hz to 1 kHz | 3.0 + 80 | ||
393B | 10.0000 mA, 100.000 mA, 10.0000 A | 40 Hz to 65 Hz | 0.7 + 80 |
65 Hz to 1 kHz | 2.0 + 80 |
- Below 5% of range, add 70 digits to accuracy
- Max. Continuous Measuring Time: 10 minutes at mA inputs, 1 minute at A inputs
- Min. Rest Time: 20 minutes after continuous measuring
- Overload Protection: AC/DC 400 mA at mA inputs, AC/DC 10 A at A inputs
Frequency
Model | Range | Resolution | Accuracy |
390B | 400.0 Hz | 0.1 Hz | 1 digit (3 3/4-digit mode) 5 digit (4 3/4-digit mode) |
4.000 kHz | 1 Hz | ||
40.00 kHz | 10 Hz | ||
100.0 kHz | 100 Hz | ||
391B & 393B | 40.000 Hz | 0.001 Hz | 0.1 + 10 |
400.00 Hz | 0.01 Hz | ||
4.0000 kHz | 0.1 Hz | ||
40.000 kHz | 1 Hz | ||
100.00 kHz | 10 Hz |
Temperature
Model | Range | Resolution | Accuracy |
390B | -200 °C to 1200 °C | 0.1 °C | 1.0 + 10 °C |
-328 °F to 2192 °F | 0.1 °F | 1.0 + 18 °F | |
391B & 393B | -200 °C to 10.0 °C | 0.1 °C | 1.0 + 2 °C |
-10.1 °C to 1200 °C | 0.1 °C | 1.0 + 1 °C | |
-328 °F to 50.0 °F | 0.1 °F | 1.0 + 4 °F | |
-50.1 °F to 2192 °F | 0.1 °F | 1.0 + 2 °F |
- Specified after 60 minutes of warm-up time
- Specification does not include error of the thermocouple probe
Supplementary Functions
Model | Function | Range | Accuracy |
390B | AC + DC | Same as voltage and current | AC accuracy ± 1.0% |
HFR | AC accuracy ± 1.0% for 40 Hz to 400 Hz | ||
Peak Hold | AC accuracy ± (3.0% + 200 digits) for 40 Hz to 1 kHz | ||
391B & 393B | AC + DC | Same as AC function | AC accuracy ± 1.0% |
HFR | Same as AC function | AC accuracy ± 1.0% for 40 Hz to 400 Hz | |
Peak Hold | 391B: Same as AC function x 1.25 393B: Same as AC function x 0.125 | AC accuracy ± (3.0% + 100 digits) for 40 Hz to 500 Hz | |
dB | 120.00 dB | Not specified | |
dBm (600 Ω) | 120.00 dBm | Not specified |
Capacitance
Model | Range | Resolution | Accuracy |
390B | 40.00 nF | 10 pF | 0.9 + 20 |
400.0 nF | 100 pF | 0.9 + 10 | |
4.000 μF | 1 nF | 0.9 + 2 | |
40.00 μF | 10 nF | ||
400.0 μF | 100 nF | ||
4.000 mF | 1 μF | 0.9 + 10 | |
40.00 mF | 10 μF | 0.9 + 20 | |
391B & 393B | 4.0000 nF | 0.1 pF | Not specified |
40.000 nF | 1 pF | 1.2 + 200 | |
400.00 nF | 10 pF | 0.8 + 20 | |
4.0000 μF | 100 pF | ||
40.000 μF | 1 nF | ||
400.00 μF | 10 nF | ||
4.0000 mF(1) | 100 nF | 1.2 + 200 | |
40.000 mF(1) | 1 μF | 1.2 + 400 |
(1) Available in manual range selection only
General
Model | 390B | 391B | 393B | |
Display | 4,000 / 40,000 count | 10,000 / 100,000 count | ||
Measurement Speed | 10 times per second | 3 times per second | ||
Data Log Capacity | 40,000 measurements | 20,000 measurements | ||
Connectivity | IR-USB and Bluetooth (class 2) | IR-USB | ||
Power | 4 x 1.5 V AA size batteries | |||
Battery Life (typical) | 50 hours | 100 hours | ||
Auto Power Off | Adjustable up to 30 minutes or never | |||
Low Battery Indicator | √ | |||
Overrange | OL or -OL is displayed | |||
Temperature | Operating | 14 ºF to 122 ºF (-10 °C to 50 °C) at ≤ 80% relative humidity | ||
Storage | -4 ºF to 140 ºF (-20 °C to 60 °C) | |||
Safety | EN61010-1 to 600 V CAT IV / 1000 V CAT III | |||
Electromagnetic Compatibility | EMC Directive 2014/30/EU, EN61326-1:2013 | |||
Dimensions (W x H x D), without case | 3.8" x 8.2" x 2" (95 mm x 207 mm x 52 mm) | |||
Weight | 1.4 lbs (630 g) | |||
Warranty | 3 Years | |||
Standard Accessories | Test leads, K-type thermocouple adapter, protective case, optical-isolated USB cable, magnetic hanging kit, alkaline batteries |