BK Precision 393B 100000 Count True RMS Handheld Digital Multimeter

SKU
393B
Manufacturer:
B&K Precision
B&K Precision

Key Features:

  • DCV, ACV, AC+DC, DCI, ACI, resistance, capacitance, frequency, diode, temperature, and continuity measurement functions
  • True RMS AC and AC+DC measurements
  • Up to 100,000-count, dual display
  • dB, dBm, limits, peak-hold, REL (delta), MIN, MAX, and average math functions
  • HFR (High Frequency Rejection) mode applies a low pass filter for AC measurements (800 Hz cut-off)
  • Isolated USB interface with operating software for remote data logging
  • Dirt and water-resistant housing with rubberized protective case
  • CAT III 1000 V / CAT IV 600 V protection
Product Options:
$580.00

Secured and trusted checkout with

The BK Precision 393B is a True RMS multimeter with an 100,000 display count. The 393B DCV accuracy is ± (0.015% + 20 digits) and it has 20,000 readings of data log capacity.

Specifications are based on the following conditions/assumptions:

  • • Accuracy specifications: ± (% of reading + counts of least significant digit) at 23 °C ± 5 °C, with relative humidity less than 80% RH One year calibration cycle
    • Temperature coefficient is 0.1 x (specified accuracy)/°C for T < 18 °C, T > 28°C
    • AC Voltage and AC Current specifications are AC coupled, true RMS
    • For non-sinusoidal waveforms:
         - Add 1.0% to accuracy specification for Crest Factor 1.0 to 2.0
         - Add 2.5% to accuracy specification for Crest Factor 2.0 to 2.5
         - Add 4.0% to accuracy specification for Crest Factor 2.5 to 3.0
    • For best accuracy use REL (delta) function to compensate the offsets

DC Voltage

ModelRangeResolutionAccuracy





390B
40.00 mV1 μV0.03 + 3
400.0 mV10 μV



0.03 + 1
4.000 V100 μV
40.00 V1 mV
400.0 V10 mV
1000 V100 mV




391B
40.000 mV1 μV0.040 + 40
400.00 mV10 μV0.035 + 20
4.000 V100 μV


0.030 + 20
40.000 V1 mV
400.00 V10 mV
1000.0 V100 mV




393B
100.000 mV1 μV0.025 + 40
1000.00 mV10 μV0.020 + 20
10.0000 V100 μV

0.015 + 20
100.000 V1 mV
1000.00 V10 mV

 

DC Current

ModelRangeResolutionAccuracy



390B
40.00 mA1 μA

0.2 + 1
400.0 mA10 μA
4.000 A100 μA
10.00 A1 mA0.2 + 2



391B
40.000 mA1 μA
0.2 + 40
400.00 mA10 μA
4.0000 A100 μA
0.2 + 80
10.000 A1 mA


393B
10.0000 mA0.1 μA
0.1 + 40
100.000 mA1 μA
10.0000 A100 μA0.1 + 80

- Burden Voltage: 2 mV/mA at mA inputs and 60 mV/A at A inputs
- Max. Continuous Measuring Time: 10 minutes at mA inputs, 1 minute at A inputs
- Min. Rest Time: 20 minutes after continuous measuring
- Overload protection: AC/DC 400 mA at mA inputs, AC/DC 10 A at A inputs

 

Continuity

ModelRangeResolutionAccuracy
390B400.0 Ω0.1 Ω0.2 + 2
391B & 393B400.00 Ω0.01 Ω0.2 + 30

- Open Circuit Voltage: -1.2 V
- Short Test Current: -0.3 mA
- Threshold: Adjustable 10 Ω to 50 Ω
- Buzzer Response Time: < 10 ms

 

Diode Test

ModelRangeResolutionAccuracy
390B2.000 V1 mV1.5 + 2
391B & 393B2.0000 V100 μV1.5 + 20

 

Resistance

ModelRangeResolutionAccuracy




390B
400.0 Ω0.1 Ω0.2 + 2
4.000 kΩ1 Ω

0.2 + 1
40.00 kΩ10 Ω
400.0 kΩ100 Ω
4.000 MΩ1 kΩ1.0 + 1
40.00 MΩ10 kΩ2.0 + 20




391B
400.00 Ω0.01 Ω

0.2 + 30
4.0000 kΩ0.1 Ω
40.000 kΩ1 Ω
400.00 kΩ10 Ω0.3 + 30
4.0000 MΩ100 Ω1.0 + 30
40.000 MΩ1 kΩ1.5 + 30




393B
1000.00 Ω0.01 Ω0.050 + 30
10.0000 kΩ0.1 Ω
0.025 + 30
100.000 kΩ1 Ω
1000.00 kΩ10 Ω0.3 + 30
10.0000 MΩ100 Ω1.0 + 30
40.000 MΩ1 kΩ1.5 + 30

 

True RMS AC Voltage

ModelRangeFrequencyAccuracy


390B


40.00 mV(1), 400.0 mV(1), 4.000 V, 40.00 V, 400.0 V(1), 1000 V(2)
40 Hz to 70 Hz(3)0.5 + 2
70 Hz to 1 kHz(3)1.5 + 4
1 kHz to 5 kHz(3)3.0 + 4
5 kHz to 100 kHz(4) (5)5.0 + 20










391B


40.000 mV, 400.00 mV
40 Hz to 65 Hz1.00 + 50
65 Hz to 1 kHz3.00 + 50
1 kHz to 3 kHz5.00 + 50




4.0000 V, 40.000 V
40 Hz to 45 Hz1.50 + 50
45 Hz to 65 Hz0.70 + 50
65 Hz to 1 kHz1.50 + 50
1 kHz to 10 kHz3.00 + 50
10 kHz to 50 kHz5.00 + 50
50 kHz to 100 kHz10.0 + 50

400.00 V, 1000.0 V
40 Hz to 45 Hz1.50 + 50
45 Hz to 65 Hz0.70 + 50
65 Hz to 1 kHz1.50 + 50











393B

100.000 mV, 1000.00 mV
40 Hz to 65 Hz0.70 + 50
66 Hz to 1 kHz1.50 + 50
1 kHz to 3 kHz3.00 + 50




10.0000 V, 100.000 V
40 Hz to 45 Hz1.00 + 50
45 Hz to 65 Hz0.40 + 50
65 Hz to 1 kHz1.00 + 50
1 kHz to 10 kHz2.00 + 50
10 kHz to 20 kHz3.00 + 50
20 kHz to 50 kHz5.00 + 50
50 kHz to 100 kHz10.0 + 50

1000.00 V
40 Hz to 45 Hz1.00 + 50
45 Hz to 65 Hz0.40 + 50
65 Hz to 1 kHz1.00 + 50

(1) The bandwidth is 40 Hz to 5 kHz
(2) The bandwidth is 40 Hz to 1 kHz
(3) Below 10% of range, add 2 digits to accuracy
(4) Below 10% of range, add 10 digits to accuracy, < 50 kHz
(5) Below 10% of range, add 20 digits to accuracy, > 50 kHz

 

True RMS AC Current

ModelRangeFrequencyAccuracy


390B


40.00 mA, 400.0 mA, 4.000 A, 10.00 A
40 Hz to 70 Hz0.8 + 2
70 Hz to 1 kHz
2.0 + 4
1 kHz to 10 kHz

391B

40.000 mA, 400.00 mA, 4.0000 A, 10.000 A
40 Hz to 65 Hz0.8 + 80
65 Hz to 1 kHz3.0 + 80

393B

10.0000 mA, 100.000 mA, 10.0000 A
40 Hz to 65 Hz0.7 + 80
65 Hz to 1 kHz2.0 + 80

- Below 5% of range, add 70 digits to accuracy
- Max. Continuous Measuring Time: 10 minutes at mA inputs, 1 minute at A inputs
- Min. Rest Time: 20 minutes after continuous measuring
- Overload Protection: AC/DC 400 mA at mA inputs, AC/DC 10 A at A inputs

 

Frequency

ModelRangeResolutionAccuracy


390B
400.0 Hz0.1 Hz

1 digit (3 3/4-digit mode)
5 digit (4 3/4-digit mode)
4.000 kHz1 Hz
40.00 kHz10 Hz
100.0 kHz100 Hz




391B & 393B
40.000 Hz0.001 Hz



0.1 + 10
400.00 Hz0.01 Hz
4.0000 kHz0.1 Hz
40.000 kHz1 Hz
100.00 kHz10 Hz

 

Temperature

ModelRangeResolutionAccuracy

390B
-200 °C to 1200 °C0.1 °C1.0 + 10 °C
-328 °F to 2192 °F0.1 °F1.0 + 18 °F



391B & 393B
-200 °C to 10.0 °C0.1 °C1.0 + 2 °C
-10.1 °C to 1200 °C0.1 °C1.0 + 1 °C
-328 °F to 50.0 °F0.1 °F1.0 + 4 °F
-50.1 °F to 2192 °F0.1 °F1.0 + 2 °F

- Specified after 60 minutes of warm-up time
- Specification does not include error of the thermocouple probe

 

Supplementary Functions

ModelFunctionRangeAccuracy



390B
AC + DC


Same as voltage and current
AC accuracy ± 1.0%
HFRAC accuracy ± 1.0% for 40 Hz to 400 Hz
Peak HoldAC accuracy ± (3.0% + 200 digits) for 40 Hz to 1 kHz





391B & 393B
AC + DCSame as AC functionAC accuracy ± 1.0%
HFRSame as AC functionAC accuracy ± 1.0% for 40 Hz to 400 Hz
Peak Hold391B: Same as AC function x 1.25
393B: Same as AC function x 0.125
AC accuracy ± (3.0% + 100 digits) for 40 Hz to 500 Hz
dB120.00 dBNot specified
dBm (600 Ω)120.00 dBmNot specified

 

Capacitance

ModelRangeResolutionAccuracy






390B
40.00 nF10 pF0.9 + 20
400.0 nF100 pF0.9 + 10
4.000 μF1 nF

0.9 + 2
40.00 μF10 nF
400.0 μF100 nF
4.000 mF1 μF0.9 + 10
40.00 mF10 μF0.9 + 20






391B & 393B
4.0000 nF0.1 pFNot specified
40.000 nF1 pF1.2 + 200
400.00 nF10 pF


0.8 + 20
4.0000 μF100 pF
40.000 μF1 nF
400.00 μF10 nF
4.0000 mF(1)100 nF1.2 + 200
40.000 mF(1)1 μF1.2 + 400

(1) Available in manual range selection only

 

General

Model390B391B393B
Display4,000 / 40,000 count10,000 / 100,000 count
Measurement Speed10 times per second3 times per second
Data Log Capacity40,000 measurements20,000 measurements
ConnectivityIR-USB and Bluetooth (class 2)IR-USB
Power4 x 1.5 V AA size batteries
Battery Life (typical)50 hours100 hours
Auto Power OffAdjustable up to 30 minutes or never
Low Battery Indicator
OverrangeOL or -OL is displayed

Temperature
Operating14 ºF to 122 ºF (-10 °C to 50 °C)
at ≤ 80% relative humidity
Storage-4 ºF to 140 ºF (-20 °C to 60 °C)
SafetyEN61010-1 to 600 V CAT IV / 1000 V CAT III
Electromagnetic CompatibilityEMC Directive 2014/30/EU, EN61326-1:2013
Dimensions (W x H x D), without case3.8" x 8.2" x 2" (95 mm x 207 mm x 52 mm)
Weight1.4 lbs (630 g)
Warranty3 Years
Standard AccessoriesTest leads, K-type thermocouple adapter, protective case, optical-isolated USB cable, magnetic hanging kit, alkaline batteries
Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account

Products In This Series

As low as

Now $460.00

As low as

Now $790.00