We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
Calibration, test, and diagnosing process loops. The PIECAL 532 is the best tool to do that!
Practical Instrument Electronics has introduced their PIE 532 loop calibrator. Offering all the functions of a loop calibrator, the 532 also delivers essential troubleshooting tools.
The 532 features the innovative Loop Diagnostic Mode, offered exclusively by PIE. The Loop Diagnostic Mode removes the need for a DMM by displaying the voltage, loop current, impedance, AC voltage and mode of operation simultaneously for the most in time efficiency. It’s capable of measuring ground fault error as a result of faulty wiring, flooded conduit and corrosion bridges. The unique diagnostic tool, available in every mode of operation, constantly updates parameters in real time and can even identify the most likely cause of loop failure.
With its Loop Diagnostic Mode, EZ-Dial knob for making adjustments, and the Dual Action EZ-Check switch for testing limits, the 532 from Practical Instrument Electronics is one of the leading Loop Calibrators available.
General Specifications: (Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration) | |
Operating Temperature Range | -20 to 60 °C (-5 to 140 °F) |
Storage Temperature Range | -30 to 60 °C (-22 to 140 °F) |
Relative Humidity Range | 10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing 10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing |
Size | 7.00 X 3.30 X 2.21 inches (177.8 x 83.8 x 56.1mm) |
Weight | 12.0 oz (340 grams) |
Battery | 9V Alkaline Optional 120 VAC 50/60 Hz AC adaptor available |
Miscellaneous | Low battery indication with nominal 1 hour of operation left Over-voltage protection to 120 Vrms (rated for 30 seconds) or 240 Vrms (rated for 15 seconds) Bar graph display with 1% resolution of 4-20 mA signal scale High contrast graphic liquid crystal display with 0.45” (11.4 mm) high digits |
Common Specifications for all Current Modes: | |
Ranges Accuracy Temperature Effect Resolution(s) | 0.000 to 24.000 mA, -25.00 to 125.00% of 4-20 mA ≤ ± (0.012 % of reading + 0.002 mA) ≤ ± 50 ppm/°C of range 0.001 mA and 0.01 % |
Source/Power and Measure 2-Wire Transmitter Specifications: | |
Loop Compliance Voltage | ≥ 24 Volts |
Loop Drive Capability | 1200 Ω at 20 mA for entire battery life, 950Ω w/HART® mode enabled |
Miscellaneous | Open loop or out of compliance conditions are indicated by appropriate error display Battery life in: Source mode ≥ 18 hrs at 12mA typical (HART® disabled) Power measure ≥ 10 hrs at 12mA typical HART® protocol mode is a selectable option at turn on. HART® protocol mode places a 250Ω resistor in series with the output Selectable EZ-Step(s) for Source Mode/2-Wire Transmitter Simulation: 2 to 16 selectable step settings Step size is determined by the selected high & low ranges Selectable time settings for stepping and soak: STEP: 5 to 900 seconds SOAK: 0 to 900 seconds |
Read mA Specifications: | |
Voltage Burden Overload/Current Limit Protection Battery Life | ≤ 2V at 20 mA nominal ≤24 mA typical ≥ 40 Hours |
2-Wire Transmitter Simulation Specifications: | |
Voltage Burden | ≤ 2 V at 20 mA |
Overload/Current Limit Protection | nominal ≤ 24 mA |
Loop Voltage Limits | 2-60 VDC |
Miscellaneous | Open loop or out of compliance conditions are indicated by appropriate error display Battery life ≥ 40 hour typical Selectable EZ-Step(s) for Source Mode/2-Wire Transmitter Simulation: 2 to 16 selectable step settings Step size is determined by the selected high & low ranges Selectable time settings for stepping and soak: STEP: 5 to 900 seconds SOAK: 0 to 900 seconds |
Voltage Specifications (Read Range): | |
Read Range: | 0.000 to 24.000 VDC then 24.01 to 30.00 (with 4X over range) Auto Ranging |
Resolutions | 0.001 V up to 24 V, 0.01 V up to 30.00 V |
Temperature Effect | ≤ ± 100 ppm/°C of range |
Input Resistance | ≥ 1 MΩ |
Accuracy | 0.000 to 24.000 VDC ≤ ± (0.012%RDG + 0.004 V) 24.01 to 30.00 VDC ≤ ± (0.025 % of reading ±0.04 V) |
Voltage Specifications (Source Range): | |
Source Range: Output Resistance Source Current Accuracy Battery life | 0.000 to 24.000 VDC ≤0.3 Ω ≥20.000 mA with Flashing indicator for over range ≤±(0.012%RDG + 0.004 V) ±50ppm/°C of range > 40 hour typical |