We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
General | |
Operating Temperature Range | -20 to 60 °C (-5 to 140 °F) |
Storage Temperature Range | -30 to 60 °C (-22 to 140 °F) |
Relative Humidity Range | 10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing |
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing | |
Ranges and Resolution | 0.00 to 24.00 mA or -25.0 to 125.0% of 4-20 mA |
Accuracy | ≤ ± (0.05 % of Reading +0.01 mA) |
Noise | ≤ ± ½ Least Significant Digit |
Temperature effect | ≤ ± 0.005 %/°C of FS |
Size | 4.75 x 2.6 x 1.50 in, 120 x 66 x 38 mm (LxWxH) |
Weight | 8 ounces, 0.23 kg (including batteries) |
Batteries | Two “AA” Alkaline 1.5V (LR6) |
Optional NiMh Rechargeable battery kit | 120 VAC for North America Only; charger, four NiMh batteries, AC & DC cords [Part # 020-0103] |
Low Battery | Low battery indication with nominal 1 hour of operation left |
Protection against misconnection | Over-voltage protection to 135 vrms (rated for 30 seconds) |
Display | High contrast graphic liquid crystal display with 0.45” (11.4 mm) high digits. Bar graph with 1% resolution of 4-20 mA signal scale. |
Source/Power & Measure Two Wire Transmitters | |
Loop compliance voltage | High Power ≥ 24 DCV @ 20.00mA Low Power ≥ 15 DCV @ 20.00mA |
Loop drive capability & Battery Life
High Power
Low Power |
1200 Ω at 20 mA for 8 hours nominal; 950 Ω with Hart Resistor enabled
600 Ω at 20 mA for 11 hours nominal; 350 Ω with Hart Resistor enabled |
2-Wire Transmitter Simulation | |
Voltage burden | ≤ 2V at 20 mA |
Overload/Current limit protection | 24 mA nominal |
Loop voltage limits | 2 to 42 VDC (fuse-less protected from reverse polarity connections) |
Battery life | ≥ 220 hours nominal |
Read mA | |
Voltage burden | ≤ 1V at 20 mA |
Overload/Current limit protection | 25 mA nominal |
Battery life | ≥ 220 Hours nominal |