Hioki IM3570 Impedance Analyzer

SKU
IM3570
Manufacturer:
Hioki
Hioki

Key Features:

  • 4 Hz to 5 MHz measurement frequencies
  • Rapid-speed testing, maximum speeds of 1.5ms (1 kHz) and 0.5ms (100 kHz) in LCR mode
  • Heightened accuracy measurements, basic accuracy of Z parameter: ±0.08%
  • LCR measurement, DCR measurement, sweep measurement, constant measurement and high-speed testing with just one instrument
  • Analyzer mode enables frequency sweeps, level sweeps and time interval measurements
Product Options:
$10,995.00

Secured and trusted checkout with

Essential for applications needing low-ESR measurement on the order of several milliohms, Hioki is offering their IM3570 Impedance Analyzer. The IM3570 is equipped for measurement frequencies of 4 Hz to 5 MHz and test signal levels of 5 mV to 5 V. Even more, it’s suited for advanced measurements like LCR measurement with AC signals, resistance measurement with direct current (Rdc) and sweep measurement.

*Please note: This product is not supplied with measurement probes or text fixtures; all are sold separately.

Measurement modes

LCR mode, Analyzer mode (Sweeps with measurement frequency and measurement level), Continuous measurement mode

Measurement parameters

Z, Y, θ, Rs (ESR), Rp, Rdc (DC resistance), X, G, B, Cs, Cp, Ls, Lp,
D (tanδ), Q

Measurement range

100 mΩ to 100 MΩ, 12 ranges (All parameters are determined according to Z)

Display range

Z, Y, Rs, Rp, Rdc, X, G, B, Ls, Lp, Cs, Cp :
±(0.000000 [unit] to 9.999999G [unit], Absolute value display for Z and Y only
θ : ±(0.000° to 180.000°), D : ±(0.000000 to 9.999999)
Q : ±(0.00 to 99999.99), Δ % : ±(0.0000% to 999.9999%)

Basic accuracy

Z ±0.08% rdg. θ: ±0.05°

Measurement frequency

4 Hz to 5 MHz (5 digits setting resolution, minimum resolution 10 mHz)

Measurement signal level

Normal mode:
V mode/CV mode: 5 mV to 5 Vrms (up to 1 MHz)
10 mV to 1 Vrms (1.0001 MHz to 5 MHz), 1 mVrms steps
CC mode: 10 μA to 50 mArms (up to 1 MHz)
10 μA to 10 mArms (1.0001 MHz to 5 MHz), 10 μArms steps
Low impedance high accuracy mode:
V mode/CV mode: 5 mV to 1 Vrms (up to 100 kHz), 1 mVrms steps
CC mode:10 μA to 100 mArms (100 mΩ and 1Ω ranges of up to 100 kHz), 10 μArms steps

Output impedance

Normal mode: 100 Ω, Low impedance high accuracy mode: 10 Ω

Display

5.7-inch color TFT, display can be set to ON/OFF

Measurement time

0.5 ms (100 kHz, FAST, display OFF, representative value)

Functions

DC bias measurement, Comparator, BIN measurement (classification), Panel loading/saving, Memory function

Interfaces

EXT I/O (handler), RS-232C, GP-IB, USB communication, USB memory, LAN

Power supply

90 to 264 V AC, 50/60 Hz, 150 VA max.

Dimensions and mass

330 mm (12.99 in) W × 119 mm (4.69 in) H × 307 mm (12.09 in) D, 5.8 kg (204.6 oz)

Accessories

Power cord × 1, Instruction manual × 1, PC communication instruction manual (CD-R) × 1

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account
  • Piezoeletric elements
  • C-D and low ESR measurement of functional polymer capacitors
  • DCR and L-Q measurement of inductors (coils and transformers)