We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
Part of the GW Instek PPX series, the GW Instek PPX-1005 is a high precision DC power supply with 10V output voltage, 5A output current, and 50W output power.
The specifications apply when the PPX is powered on for at least 30 minutes under +20°C~+30°C.
DC Output Mode | PPX-1005 | PPX-2002 | PPX-2005 | PPX-3601 | PPX-3603 | PPX-10H01 | |
Output voltage | 10.000V | 20.000V | 20.000V | 36.000V | 36.000V | 100.00V | |
Output current | 5.0000A | 2.0000A | 5.0000A | 1.0000A | 3.0000A | 1.0000A | |
Output power | 50W | 40W | 100W | 36W | 108W | 100W | |
CONSTANT VOLTAGE OPERATION | |||||||
Line regulation*7 | ±(0.01% of setting + 1mV) | ±(0.01% of setting + 1mV) | ±(0.01% of setting + 1mV) | ±(0.01% of setting + 3mV) | ±(0.01% of setting + 3mV) | ±(0.01% of setting+ 7mV) | |
Load regulation*8 | ±(0.01% of setting + 2mV) | ±(0.01% of setting + 2mV) | ±(0.01% of setting + 3mV) | ±(0.01% of setting + 3mV) | ±(0.01% of setting + 4mV) | ±(0.01% of setting + 7mV | |
Transient response *1 | <50us | <100us | |||||
Ripple& noise | (Vrms*2/ Vpp*3) | 0.35mVrms/ <6mV Vpp | 0.5mVrms/ <8mV Vpp | 0.5mVrms/ <8mV Vpp | 0.8mVrms/ <10mV Vpp | 0.8mVrms/ <10mV Vpp | 1.2mVrms/ <15mV Vpp |
Rise time*4 | Rated load | 20ms or less | 50ms or less | 100ms or less | |||
No load | 20ms or less | 50ms or less | 100ms or less | ||||
Fall time*5 | Rated load | 10ms or less | 20ms or less | 50ms or less | |||
No load | 100ms or less | 150ms or less | 250ms or less | ||||
Setting range (105%) | 0V to 10.5V | 0V to 21.0V | 0V to 21.0V | 0V to 37.8V | 0V to 37.8V | 0V to 105.0V | |
Setting resolution | 1mV | 1mV | 1mV | 1mV | 1mV | 10mV | |
Setting accuracy (23℃±5℃) | ±(0.03% of setting + 3mV) | ±(0.03% of setting + 5mV) | ±(0.03% of setting + 5mV) | ±(0.03% of setting + 8mV) | ±(0.03% of setting + 8mV) | ±(0.03% of setting+ 20mV) | |
Maximum remote sensing compensation voltage (single line) | 1V | 3V | |||||
Temperature Coefficient (TYP.) *6 | 100 ppm/℃ | ||||||
CONSTANT CURRENT OPERATION | |||||||
Line regulation*7 | ±(0.02% of setting + 250uA) | ±(0.02% of setting + 100uA) | ±(0.02% of setting + 250uA) | ±(0.02% of setting + 50uA) | ±(0.02% of setting + 150uA) | ±(0.02% of setting + 50uA) | |
Load regulation*9 | ±(0.02% of setting + 250uA) | ±(0.02% of setting + 100uA) | ±(0.02% of setting + 250uA) | ±(0.02% of setting + 50uA) | ±(0.02% of setting + 150uA) | ±(0.02% of setting+ 50uA) | |
Ripple & noise | (Arms*2) | 2mA | 1mA | 2mA | 400uA | 1mA | 1mA |
Setting range (105%) | 0A to 5.25A | 0A to 2.1A | 0A to 5.25A | 0A to 1.050A | 0A to 3.15A | 0A to 1.050A | |
Setting resolution | 0.1mA | 0.1mA | 0.1mA | 0.1mA | 0.1mA | 0.1mA | |
Setting accuracy (23℃±5℃) | ±(0.05% of setting + 3.0mA) | ±(0.05% of setting + 1.0mA) | ±(0.05% of setting + 3.0mA) | ±(0.05% of setting + 0.5mA) | ±(0.05% of setting + 1.5mA) | ±(0.05% of setting+ 1.0mA) | |
Temperature Coefficient (TYP.) *6 | 200 ppm/℃ | ||||||
*1. Time for output voltage to recover within ±(0.1% + 10mV) of its rated output for a load change from 50% to 100% of its rated output current. *2. Measurement frequency bandwidth is 5 Hz to 1 MHz. *3. Measurement frequency bandwidth is 10 Hz to 20 MHz. *4. From 10%~90% of rated output voltage, with rated resistive load. *5. From 90%~10% of rated output voltage, with rated resistive load. *6. Temperature coefficient: after a 30 minute warm-up. *7. At 90~110Vac or 108~132Vac or 198~242Vac or 216~264Vac, constant load. *8. From No-load to Full-load, constant AC input voltage. Make sure that test leads and output terminals are well connected. It is suggested that utilize 4-wire connection when European terminal output models are employed. *9. For load voltage change, equal to the unit voltage rating, constant AC input voltage. |
Measurement and Display | PPX-1005 | PPX-2002 | PPX-2005 | PPX-3601 | PPX-3603 | PPX-10H01 | |
Voltage Range | H | 10.000V | 20.000V | 20.000V | 36.000V | 36.000V | 100.00V |
L | 1.0000V | 2.0000V | 2.0000V | 3.6000V | 3.6000V | 10.000V | |
Current Range | H | 5.0000A | 2.0000A | 5.0000A | 1.0000A | 3.0000A | 1.0000A |
M | 500.00mA | 200.00mA | 500.00mA | 100.00mA | 300.00mA | 100.00mA | |
L | 50.000mA | 20.000mA | 50.000mA | 10.000mA | 30.000mA | 10.000mA | |
LL | 5.0000mA | 2.0000mA | 5.0000mA | 1.0000mA | 3.0000mA | 1.0000mA | |
Measurement Resolution | Voltage (H) | 1mV | 10mV | ||||
Voltage (L) | 0.1mV | 1mV | |||||
Current (H) | 0.1mA | 0.1mA | |||||
Current (M) | 0.01mA | 0.01mA | |||||
Current (L) | 0.001mA | 0.001mA | |||||
Current (LL) | 0.0001mA | 0.0001mA | |||||
Measurement Accuracy | Voltage (H/L) | ±(0.03% of rdg + 2mV) | ±(0.03% of rdg+ 4mV) | ±(0.03% of rdg + 5mV) | ±(0.03% of rdg + 6mV) | ±(0.03% of rdg + 8mV) | ±(0.03% of rdg + 15mV) |
Temperature Coefficient *1 (TYP.) | 100 ppm/℃ | ||||||
Current (H/M) | ±(0.05% of rdg + 2.5mA) | ±(0.05% of rdg + 1.0mA) | ±(0.05% of rdg + 2.5mA) | ±(0.05% of rdg + 0.4mA) | ±(0.05% of rdg + 1.2mA) | ±(0.05% of rdg + 1.0mA) | |
Current (L/LL) | ±(0.1% of rdg + 40µA) | ±(0.1% of rdg + 24µA) | ±(0.1% of rdg + 40µA) | ±(0.1% of rdg + 16µA) | ±(0.1% of rdg + 28µA) | ±(0.1% of rdg + 24µA) | |
Temperature Coefficient *1 (TYP.) | 200 ppm/℃ | ||||||
*1. Temperature coefficient: after a 30 minute warm-up. |
Temperature measurement | PPX-1005 | PPX-2002 | PPX-2005 | PPX-3601 | PPX-3603 | PPX-10H01 | |
Temperature (K-Type Thermocouple) | Range | -200℃~+1372℃ | |||||
Resolution | 0.25℃ | ||||||
Accuracy | ±(0.5% + 2℃) | ||||||
Protection | PPX-1005 | PPX-2002 | PPX-2005 | PPX-3601 | PPX-3603 | PPX-10H01 | |
Over Voltage Protection (OVP) | Operation | Turns the output off, displays OVP and lights ALARM | |||||
Setting range | 0.5V to 11.0V | 1.0V to 22.0V | 1.0V to 22.0V | 1.8V to 39.6V | 1.8V to 39.6V | 5.0V to 110.0V | |
(5% to 110% of the rated output voltage) | |||||||
Setting Accuracy | ±(1% of rating) | ||||||
Over Current Protection (OCP) | Operation | Turns the output off, displays OCP and lights ALARM | |||||
Setting range | 0.25A to 5.5A | 0.1A to 2.2A | 0.25A to 5.5A | 0.05A to 1.1A | 0.15A to 3.3A | 0.05A to 1.1A | |
(5% to 110% of the rated output current) | |||||||
Setting Accuracy | ±(1% of rating) | ||||||
Over Temperature Protection (OTP) | Operation | Turns the output off, displays OTP and lights ALARM | |||||
Communications | PPX-1005 | PPX-2002 | PPX-2005 | PPX-3601 | PPX-3603 | PPX-10H01 | |
LAN | MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Subnet Mask | ||||||
USB | Type A: Host, Type B: Slave, Speed: 1.1/2.0, USB-CDC | ||||||
RS-232/RS-485 | Complies with the EIA-RS-232/RS-485 specifications (excluding the connector) | ||||||
Input ratings (AC rms) | PPX-1005 | PPX-2002 | PPX-2005 | PPX-3601 | PPX-3603 | PPX-10H01 | |
Nominal input voltage *1 | 100Vac / 120Vac / 220Vac / 240Vac(±10%), 50Hz / 60Hz, single phase | ||||||
Input frequency range | 47Hz to 63Hz | ||||||
Max. Inrush current | 25Amax or less | 20Amax or less | 30Amax or less | 35Amax or less | 40Amax or less | 30Amax or less | |
Max. power consumption | 200VA | 150VA | 300VA | 150VA | 300VA | 300VA | |
*1. Before connecting the power plug to an AC line outlet, make sure the voltage selector switches of the bottom panel in the correct position. It might be damaged the instrument by connecting to the wrong AC line voltage |
General | PPX-1005 | PPX-2002 | PPX-2005 | PPX-3601 | PPX-3603 | PPX-10H01 | |
Dimensions (mm) & Weight | 107(W)×124(H)×313(D) (not including protrusions),Approx. 5.5kg | ||||||
Operating temperature | 0˚C to 40˚C | ||||||
Storage temperature | -20˚C to 70˚C | ||||||
Operating humidity | 20% to 80% RH; No condensation | ||||||
Storage humidity | 20% to 85% RH; No condensation | ||||||
Altitude | Maximum 2000m | ||||||
EMC | Complies with the European EMC directive 2014/30/EU for Class Atest and measurement products. | ||||||
Safety | Complies with the European Low Voltage Directive 2014/35/EU and carries the CE-marking. | ||||||
Withstand voltage | Between input and chassis | No abnormalities at 1500 Vac for 1 minute. | |||||
Between input and output | No abnormalities at 3000 Vac for 1 minute. | ||||||
Between output and chassis | No abnormalities at 500 Vdc for 1 minute. | ||||||
Insulation resistance | Between input and chassis | 500 Vdc, 100MΩ or more | |||||
Between input and output | 500 Vdc, 100MΩ or more | ||||||
Between output and chassis | 500 Vdc, 100MΩ or more |