We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The 5609 Secondary PRT Probe from Fluke Calibration combines both stability and precision into one elite secondary reference PRT. The 5609 covers an expansive temperature range of -200°C to 670°C with 100 mm minimum immersion.
The PRT is constructed with an Inconel™ sheath and the sensors with reference-grade platinum that contain four-wire connections. The 5609 comes with a wide selection of length options as well as a 90° bend (nine inches from the tip) option for dry well reference.
As standard, each probe includes its resistance value at the triple point of water. If calibration is desired please call to discuss.
Temperature range | –200 °C to 670 °C |
Nominal resistance at 0.01 °C | 100 Ω ± 0.5 Ω |
Temperature coefficient | 0.0039250 Ω/Ω/°C |
Relative Accuracy | ± 0.007 °C at –200 °C |
Short-term repeatability | ± 0.007 °C at 0.010 °C |
Drift[3] | ± 0.01 °C at 0.010 °C |
Hysteresis | ± 0.01 °C maximum |
Sensor length | 30 mm ± 5 mm (1.2 in ± 0.2 in) |
Sensor location | 3 mm ± 1 mm from tip (0.1 in ± 0.1 in) |
Sheath material | Inconel™ 600 |
Minimum insulation resistance | 500 MΩ at 23 °C, 10 MΩ at 670 °C |
Transition junction temperature range | –50 ºC to 200 °C |
Transition junction dimensions | 71 mm x 12.5 mm (2.8 in x 0.49 in) |
Minimum immersion length | 100 mm (3.9 in) |
Maximum immersion length | 305 mm (12 in) |
Response time | 12 seconds typical |
Self heating (in 0 °C bath) | 50 mW/°C |
Lead-wire cable type | PTFE 24 AWG |
Lead-wire length | 1.8 m (6 ft) |
Lead-wire temperature range | –50 °C to 250 °C |