Fluke Calibration 9190A Ultra-Cool Field Metrology Well (-95 to 140 C)

SKU
9190A
Manufacturer:
Fluke Calibration
Fluke Calibration

Key Features:

  • Wide temperature range: From –95 °C to 140 °C
  • Excellent accuracy: Accuracy using built-in reference thermometer readout: ± 0.05 °C full range; Display accuracy: ± 0.2 °C full range
  • Best-in-class stability: ± 0.015 °C full range
  • Fast cooling time: 23 °C to –90 °C: 80 minutes; 23 °C to –95 °C: 90 minutes; 140 °C to 23 °C: 60 minutes
  • Portability: Weighs only 16 kg (35 lb); Built-in front and back handles for easy two-handed carry
  • Best measurement practices: Conforms to EURAMET cg-13 guidance on measurement practices for temperature block calibrators
  • Ergonomics: Two-handed carry; built-in handles on front and back; All controls and inputs on the front panel
  • Stability indicator: Shows whether block temperature is stable and within limits defined by the user
  • PC interface: USB and RS-232 support remote operation from a computer
  • Dual heater zone: Active heater zone control compensates for differential temperatures between zones; Minimizes axial temperature gradients
  • Ramp-and-soak set-points: Define up to eight set-point temperatures; automatically ramp to and “soak” at each set-point; Specify soak time, number of cycles, and direction
  • Fast time to temperature and small footprint: Compact, efficient free piston Stirling cooler technology (FPSC)
  • Product Options:
    $22,375.00

    Secured and trusted checkout with

    Great for clean room environments

    Calibration baths are the most stable and uniform temperature sources available, but they aren’t a good fit for clean rooms. The size of a bath limits its portability, and bath fluids can easily spill and give off vapors. The 9190A Ultra-Cool Field Metrology Well is a great alternative. Its wide temperature range brackets the coldest and highest temperature ranges required for pharmaceutical, biomedical and food processing applications. The 9190A is small and lightweight, making it easy to transport. And since it does not use heat transfer fluids, clean rooms stay clean. The 9190A cooling and heating times are faster than a calibration bath—that means calibration work gets done more quickly.

    An accurate temperature source is critical for dependable process measurements

    Unreliable process measurements can have a damaging impact on business, leading to poor product quality, recalls, fines, waste, and lost profits. Ultimately, measurements are only as good as the temperature sources used to calibrate the measurement equipment. The 9190A Ultra-Cool Field Metrology Well incorporates the best technology and design expertise gained from decades of dry-block development experience. The 9190A conforms with EURAMET cg-13 guidelines for best measurement practices for temperature block calibrators. As a result, you can be assured that the 9190A specifications for accuracy, stability, axial (vertical) uniformity, radial (well-to-well) uniformity, loading, and hysteresis have been thoroughly and carefully defined and tested. With a 9190A Ultra-Cool Field Metrology Well, you can be confident you’re using the most accurate and stable ultra-cool dry-block calibrator available. And that will have a positive impact on your business.

     

     

                                         Fittings

     

     

    Specifications
    Temperature range at 23°C–95°C to 140°C (–139°F to 284°F)
    Display accuracy±0.2°C full range
    Accuracy with external reference3±0.05°C full range
    Stability±0.015°C full range
    Axial uniformity at 40 mm (1.6 in)±0.05°C full range
    Radial gradient±0.01°C full range
    Loading effect(with a 6.35 mm reference probe and three 6.35 mm probes)
    ±0.006°C full range
    (versus display with one 6.35 mm probe)
    ±0.25°C at –95°C
    ±0.10°C at 140°C
    Operating conditions0°C to 35°C, 0% to 90%
    RH (non-condensing) < 2000 m altitude
    Environmental conditions for all specifications except temperature range13°C to 33°C
    Immersion (well) depth160 mm (6.3 in)
    Well diameter30 mm (1.18 in)
    Heating time1–95°C to 140°C: 40 min
    Cooling time123°C to –90°C: 80 min
    23°C to –95°C: 90 min
    140°C to 23°C: 60 min
    Stabilization time215 min
    Resolution0.01°
    DisplayLCD, °C or °F user selectable
    Size (H x W x D)480 x 205 x 380 mm (18.8 x 8.0 x 14.9 in)
    Weight16 kg (35 lb)
    Power requirements100 V to 115 V (±10%) 50/60 Hz, 575 W
    200 V to 230 V (±10%) 50/60 Hz, 575 W
    System fuse ratings115 V: 6.3 A T 250 V
    230 V: 3.15 A T 250 V
    4–20 mA fuse (-P model only)50 mA F 250 V
    Computer interfaceRS-232, USB Serial, and 9930 interface-it temperature calibration software included
    SafetyIEC 61010-1, Installation Category II, Pollution degree 2
    Electromagnetic environmentIEC 61326-1: Basic
    RefrigerantsR32 (Difluoromethane)
    < 20 g, ASHRAE safety group A2L
    R704 (Helium)
    < 20 g, ASHRAE safety group A1
    -P Specifications
    Built-in reference thermometer readout accuracy (4-wire reference probe)3±0.010°C at –95°C
    ±0.013°C at –25°C
    ±0.015°C at 0°C
    ±0.020°C at 50°C
    ±0.025°C at 140°C
    Reference resistance range0 Ω to 400 Ω
    Reference resistance accuracy40 Ω to 42 Ω: ±0.0025 Ω
    42 Ω to 400 Ω: ±60 ppm of reading
    Reference characterizationsITS-90, CVD, IEC-751, resistance
    Reference measurement capability4 wire
    Reference probe connection6-pin din with INFO-CON technology
    Built-in RTD thermometer readout accuracyNI-120: ±0.015°C at 0°C
    PT-100 (385): ±0.02°C at 0°C
    PT-100 (3926): ±0.02°C at 0°C
    PT-100 (JIS): ±0.02°C at 0°C
    RTD resistance range0 Ω to 400 Ω
    Resistance accuracy40 Ω to 25 Ω: ±0.002 Ω
    25 Ω to 400 Ω: ±80 ppm of reading
    RTD characterizationsPT-100 (385), (JIS), (3926), NI-120, resistance
    RTD measurement capability2-wire, 3-wire, and 4-wire RTD with jumpers only
    RTD connection4-terminal input
    Built-in TC thermometer readout accuracy5Type J: ±0.70°C at 140°C
    Type K: ±0.75°C at 140°C
    Type T: ±0.60°C at 140°C
    Type E: ±0.60°C at 140°C
    Type R: ±1.60°C at 140°C
    Type S: ±1.60°C at 140°C
    Type M: ±0.65°C at 140°C
    Type L: ±0.65°C at 140°C
    Type U: ±0.70°C at 140°C
    Type N: ±0.75°C at 140°C
    Type C: ±1.00°C at 140°C
    TC millivolt range–10 mV to 75 mV
    Voltage accuracy0.025% of reading +0.01 mV
    Internal cold junction compensation accuracy±0.35°C (ambient of 13°C to 33°C)
    TC connectionMiniature connectors (ASTM E1684)
    Built-in mA readout accuracy0.02% of reading + 0.002 mA
    mA rangeCal 4-22 mA, Spec 4-24 mA
    mA connection2 terminal input
    Loop power function24 V DC loop power
    Built-in electronics temperature coefficient
    (0°C to 13°C, 33°C to 50°C)
    ±0.005% of range per °C
    1. For ambient temperature of 23°C.
    2. Time from when the SETPOINT is reached to when the unit is with in stability specification.
    3. The temperature range may be limited by the reference probe connected to the readout.
    The built-in reference accuracy does not include the sensor probe accuracy. It does not include the probe uncertainty or probe characterization errors.
    4. Measurement accuracy specifications apply within the operating range and assume 4 wires for PRTs. With 3-wire RTDs add 0.05 Ω to the measurement accuracy plus the maximum possible difference between the resistances of the lead wires.
    5. The thermocouple input readout is sensitive to EM fields in the frequency range of 500 MHz to 700 MHz.
    Write Your Own Review