We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
Fluke 87V True RMS Digital Multimeter
Specifications | ||
Voltage DC | Maximum voltage | 1000 V |
Accuracy | ±(0.05% + 1) | |
Maximum resolution | 10 µV | |
Voltage AC | Maximum voltage | 1000 V |
Accuracy | ±(0.7% + 2) True RMS | |
AC bandwidth | 20 kHz with low pass filter; 3 db @ 1 kHz | |
Maximum resolution | 0.1 mV | |
Current DC | Maximum amps | 10 A (20 A for 30 seconds maximum) |
Amps accuracy | ±(0.2% + 2) | |
Maximum resolution | 0.01 µA | |
Current AC | Maximum amps | 10 A (20 A for 30 seconds maximum) |
Amps accuracy | ±(1.0% + 2) True RMS | |
Maximum resolution | 0.1 µA | |
Resistance | Maximum resistance | 50 MΩ |
Accuracy | ±(0.2% + 1) | |
Maximum resolution | 0.1 Ω | |
Capacitance | Maximum capacitance | 9,999 µF |
accuracy | ±(1% + 2) | |
Maximum resolution | 0.01 nF | |
Frequency | Maximum frequency | 200 kHz |
Accuracy | ±(0.005% + 1) | |
Maximum resolution | 0.01 Hz | |
Duty cycle | Maximum duty cycle | 99.9% |
Accuracy | ±(0.2% per khz + 0.1%) | |
Maximum resolution | 0.1% | |
Temperature measurement | –200.0°C – 1090°C –328.0°F – 1994.0°F excluding probe | |
80 BK temperature probe | –40.0°C – 260°C –40.0°F – 500°F, 2.2°C or 2% whichever is greater | |
Conductance | Maximum conductance | 60.00 nS |
Accuracy | ±(1.0% + 10) | |
Maximum resolution | 0.01 nS | |
Diode | Range | 3 V |
Resolution | 1 mV | |
Accuracy | ±(2% + 1) | |
Duty cycle range | Accuracy | Within ±(0.2% per kHz + 0.1%) |
Environmental Specifications | ||
Operating temperature | –20°C to + 55°C | |
Storage temperature | –40°C to + 60°C | |
Humidity (without condensation) | 0% – 90% (0°C – 35°C) 0% – 70% (35°C – 55°C) | |
Operating Altitude | 2000 m | |
Safety Specifications | ||
Overvoltage category | EN 61010–1 to 1000 V CAT III, 600V CAT IV | |
Agency approvals | UL, CSA, TÜV, VDE listed | |
Mechanical and General Specifications | ||
Size | 201 x 98 x 52 mm (with holster) | |
Weight | 355 g 624 g – with holster | |
Display | Digital | 6000 counts updates 4/sec. 19,999 counts in high–resolution mode |
Analog | 32 segments, updates 40/sec | |
Frequency | 19,999 counts, updates 3/sec at > 10 Hz | |
Warranty | Lifetime | |
Battery Life | Alkaline | ~400 hours typical, without backlight |
Shock | 1 Meter drop per IEC 61010–1:2001 | |
Vibration | Per MIL–PRF–28800 for a Class 2 instrument |
Fluke PRV240 Proving Unit
Output voltage | 240 V ac rms or dc | ±10 % ≥1 MΩ |
LoZ >3 kΩ load | 60 V ac typical | |
LED power indicator | Turns on when output voltage is present | |
Battery | 4 x AA Alkaline batteries NEDA | 24 A IEC LR03 |
Battery life | 5000 (5-second duration) test cycles with >1 MΩ load, 300 tests minimum with >3 kΩ load | |
Operating temperature | -10 °C to +50 °C | |
Operating humidity | 0 % to 90 % (0 °C to 35 °C) | 0 % to 70 % (35 °C to 55 °C) |
Operating altitude | 2000 m | |
Dimensions | 11.7 cm x 7.4 cm x 2.8 cm (4.6 in x 2.9 in x 1.1 in), Pollution Degree 2 | |
Weight | 0.23 kg (8 oz) includes batteries | |
Safety | IEC61010-1, IEC61010-2-030 | |
Certification | CSA-CUS, CE | |
Warranty | One Year | |
Electromagnetic Compatibility (EMC) | IEC 61326-1 | Portable EM environment; CISPR 11, Group 1, Class A |
US (FCC) | 47 CFR 15 subpart B, this product is considered an exempt device per clause 15.103 | |
Korea (KCC) | Class A Equipment (Industrial Broadcasting & Communication Equipment) | |
This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it. This equipment is intended for use in business environments and not to be used in homes |
Fluke 87V Meter includes
- • 9V battery
- • Test Leads
- • Rubber Holster with Stand
Fluke PRV240 Proving Unit includes
- • TPAK hanging strap
- • Four AA batteries
- • Instruction sheet