We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The Fluke 721Ex Precision Pressure Calibrator comes equipped with dual isolates pressure sensors, allowing for simultaneous static and differential pressure measurements to be done with one isntrument. It is ideal for gas custody applications, even in potentially hazardous atmospheres. The Fluke 721Ex can be configured by selecting either a 16 PSIG (1.1 bar) or 36 PSIG (2.48 bar) low pressure sensor then add any of 7 high pressure ranges including 100, 300, 500, 1000, 1500, 3000 or 5000 psig (6.9, 20, 24.5, 69, 103.4, 200, 345 bar).
The 721-Ex reads pressure in the following pressure psig, bar, mbar, kPa, kgcm2, cmH2O@4°C, cmH2O@20°C, mH2O@4°C, mH2O@20°C, inH2O@4°C, inH2O@20°C, inH2O@60°F, mmHg@0°C, inHg@0°C, ftH2O@4°C, ftH2O@20°C, ftH2O@60°F
Models 721-Ex available models based on pressure range:
FLUKE-721Ex-1603: 16 PSIG, 300 PSIG
FLUKE-721Ex-1615: 16 PSIG, 1500 PSIG
FLUKE-721Ex-1630: 16 PSIG, 3000 PSIG
FLUKE-721Ex-3615: 36 PSIG, 1500 PSIG
FLUKE-721Ex-3630: 36 PSIG, 3000 PSIG
FLUKE-721Ex-3650: 36 PSIG, 5000 PSIG
Specifications 1-Year | |||
Fluke-721Ex-1601 | Low pressure sensor | ||
Range | Resolution | Accuracy | |
-14 psi to +16 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
-12 psi to +100 psi, | 0.01 psi, | 0.025% of full scale | |
Fluke-721Ex-1603 | Low pressure sensor | ||
-14 psi to +16 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
-12 psi to +300 psi, | 0.01 psi, | 0.025% of full scale | |
Fluke-721Ex-1605 | Low pressure sensor | ||
-14 psi to +16 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
-12 psi to +500 psi, | 0.01 psi, | 0.025% of full scale | |
Fluke-721Ex-1610 | Low pressure sensor | ||
0 psi to +1000 psi, | 0.1 psi, | 0.025% of full scale | |
High pressure sensor | |||
-12 psi to +500 psi, | 0.01 psi, | 0.025% of full scale | |
Fluke-721Ex-1615 | Low pressure sensor | ||
Range | Resolution | Accuracy | |
-14 psi to +16 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
0 psi to +1500 psi, | 0.1 psi, | 0.025% of full scale | |
Fluke-721Ex-1630 | Low pressure sensor | ||
-14 psi to +16 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
0 psi to +3000 psi, | 0.1 psi, | 0.025% of full scale | |
Fluke-721Ex-1650 | Low pressure sensor | ||
-14 psi to +16 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
0 psi to +5000 psi, | 0.1 psi, | 0.035% of full scale | |
Fluke-721Ex-3601 | Low pressure sensor | ||
Range | Resolution | Accuracy | |
-14 psi to +36 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
-12 to +100 psi, | .01 psi, | 0.025% of full scale | |
Fluke-721Ex-3603 | Low pressure sensor | ||
-14 psi to +36 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
-12 psi to +300 psi, | 0.01 psi, | 0.025% of full scale | |
Fluke-721Ex-3605 | Low pressure sensor | ||
-14 psi to +36 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
-12 psi to +500 psi, | 0.01 psi, | 0.025% of full scale | |
Fluke-721Ex-3610 | Low pressure sensor | ||
Range | Resolution | Accuracy | |
-14 psi to +36 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
0 psi to +1000 psi, | 0.1 psi, | 0.025% of full scale | |
Fluke-721Ex-3615 | Low pressure sensor | ||
-14 psi to +36 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
0 psi to +1500 psi, | 0.1 psi, | 0.025% of full scale | |
Fluke-721Ex-3630 | Low pressure sensor | ||
-14 psi to +36 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
0 psi to +3000 psi, | 0.1 psi, | 0.025% of full scale | |
Fluke-721Ex-3650 | Low pressure sensor | ||
-14 psi to +36 psi, | 0.001 psi, | 0.025% of full scale | |
High pressure sensor | |||
0 psi to +5000 psi, | 0.1 psi, | 0.035% of full scale | |
General Specifications | |||
Instrument setup recall | Last used on power-up | ||
Environmental | Operating temperature -10°C to +45°C | ||
Altitude | 2,000 m | ||
Ingress protection | IP-40 | ||
Power requirements | 6.0 V DC | ||
Battery life | >35 hours, typical usage | ||
Batteries required | (4) Standard AA cells | ||
Dimensions (H x W x D) | 7.9 x 4.3 x 2.3 in (20 x 11 x 5.8 cm) | ||
Weight | 03539 kg (1 lb 3 oz) | ||
Connectors/ports pressure | Two, 1/8 in NPT, 700P Series Pressure Module; RTD Probe | ||
Electrical and Temperature Measurement 1-Year | |||
Function | Range | Resolution | Accuracy |
mA measure | 0 to 24 mA | 0.001 mA | ±0.015% of rdg |
Temperature Effect | |||
No effect on accuracy on all functions from 15°C to 35°C | |||
Add ±0.002% F.S./°C for temps outside of 15°C to 35°C | |||
Available Pressure Engineering Units | |||
psi, bar, mbar, kPa, kgcm2, cmH2O @ 4°C, cmH2O @ 20°C, mH2O @ 4°C, mH2O @ 20°C, inH2O @ 4°C, inH2O @ 20°C, inH2O @ 60°F, mmHg @ 0°C, inHg @ 0°C, ftH2O @ 4°C, ftH2O @ 20°C, ftH2O @ 60°F | |||
Available Pressure Engineering Units | |||
ATEX | Ex ia IIB T3 Gb (Ta= -10… +45 °C) | ||
KEMA 10 ATEX 0168X | |||
IECEx | Ex ia IIB T3 Gb (Ta= -10…+45 °C) | ||
II 2 G IECEx CSA 10.0013X |