Fluke 712B RTD Temperature Calibrator

SKU
FLUKE-712B
Manufacturer:
Fluke
Fluke

Key Features:

  • The 712B can measure and simulate (13) different RTD types and resistance
  • Measure 4 to 20 mA signals while simultaneously sourcing a temperature signal
  • Hanging tool designed in and included with every unit
  • Configurable 0 % and 100 % source settings for quick 25 % linearity checks
  • Linear ramp and 25 % step auto ramp based on 0 % and 100 % settings
  • Dual inputs and backlit display for easy interpretation of measurements
  • Power down settings remembered at power up for easy restart of tests
  • 1-year and 2-year specifications and traceable certification of calibration
  • Product Options:
    $1,970.99

    Secured and trusted checkout with

    For the temperature calibration professional that wants a highly accurate, easy-to-use, single function RTD temperature calibrator the Fluke 712B is the ideal test tool.
    General Specifications
    Maximum voltage applied between any terminal and earth ground or between any two terminals30 V
    Operating temperature-10 °C to 50 °C
    Storage temperature-30 °C to 60 °C
    Operating altitude2,000 meters
    Storage altitude12,000 meters
    Relative humidity (% RH operating without condensation)Non condensing
    90% (10 °C to 30 °C)
    75% (30 °C to 40 °C)
    45% (40 °C to 50 °C)
    (Without condensation)
    Vibration requirementsMIL-T-28800E, Class 2
    Drop text requirements1 meter
    IP RatingIEC 60529: IP52
    Electromagnetic environmentIEC 61326-1, Portable
    SafetyIEC 61010-1, Max 30 V to earth, Pollution Degree 2
    Power supply4 AA NEDA 1.5A IEC LR6 batteries
    Size (H x W x L)52.5 mm x 84 mm x 188.5 mm
    Weight515 g
    DC mA Measurement
    Resolution0-24 mA
    Range0.001 mA
    Accuracy (% of reading + counts)0.010% + 2 μA
    Temperature coefficient± (0.002% of reading + 0.002% of range) /°C ( <18 °C or >28 °C)
    Ohms Measurement
    Ohms rangeAccuracy (% of reading +counts)
    0.00 Ω to 400.00 Ω0.015% + 0.05 Ω
    400.0 Ω to 4000.0 Ω0.015% + 0.5 Ω
    Note: Read accuracy is based on 4-wire input. For 3-wire ohm measurements, assuming all three leads are matched, add 0.05 Ω (0.00 Ω~400.00 Ω), 0.2 Ω (400.0 Ω~4000.0 Ω) to the specifications.
    Temperature coefficient± (0.002% of reading + 0.002% of range) /°C ( <18 °C or >28 °C)
    Ohms Source
    Ohms range1.0 Ω to 400.0 Ω
    1.00 Ω to 400.00 Ω
    400.0 Ω to 1500.0 Ω
    1500.0 Ω to 4000.0 Ω
    Excitation current from measurement device0.1 mA to 0.5 mA
    0.5 mA to 3 mA
    0.05 mA to 0.8 mA
    0.05 mA to 0.4 mA
    Accuracy (% of reading + counts)0.015% + 0.1 Ω
    0.015% + 0.05 Ω
    0.015% + 0.5 Ω
    0.015% + 0.5 Ω
    Resolution0.00 Ω to 400.00 Ω0.01 Ω
    400.0 Ω to 4000.0 Ω0.1 Ω
    Temperature coefficient± (0.002% of reading + 0.002% of range) /°C ( <18 °C or >28 °C)
    Supports pulsed transmitters and PLCs with pulse times as short as 5 ms
    RTD Input and Output
    RTD Type (α)Range (°C)Measure (°C)Source (°C)
    1-year2-yearSource current1-year2-year
    10Ω Pt(385)-200 to 100 °C1.5 °C3 °C1 mA1.5 °C3 °C
    100 to 800 °C1.8 °C3.6 °C1 mA1.8 °C3.6 °C
    50Ω Pt(385)-200 to 100 °C0.4 °C0.7 °C1 mA0.4 °C0.7 °C
    100 to 800 °C0.5 °C0.8 °C1 mA0.5 °C0.8 °C
    100 Ω Pt(385)-200 to 100 °C0.2 °C0.4 °C1 mA0.2 °C0.4 °C
    100 to 800 °C0.015%
    +0.18 °C
    0.03%
    +0.36 °C
    1 mA0.015%
    +0.18 °C
    0.03%
    +0.36 °C
    200 Ω Pt(385)-200 to 100 °C0.2 °C0.4 °C500 μA0.2 °C0.4 °C
    100 to 630 °C0.015%
    +0.18 °C
    0.03%
    +0.36 °C
    500 μA0.015%
    +0.18 °C
    0.03%
    +0.36 °C
    500 Ω Pt(385)-200 to 100 °C0.3 °C0.6 °C250 μA0.3 °C0.6 °C
    100 to 630 °C0.015%
    +0.28 °C
    0.03%
    +0.56 °C
    250 μA0.015%
    +0.28 °C
    0.03%
    +0.56 °C
    1000 Ω Pt(385)-200 to 100 °C0.2 °C0.4 °C250 μA0.2 °C0.4 °C
    100 to 630 °C0.015%
    +0.18 °C
    0.03%
    +0.36 °C
    250 μA0.015%
    +0.18 °C
    0.03%
    +0.36 °C
    100 Ω Pt(3916)-200 to 100 °C0.2 °C0.4 °C1 mA0.2 °C0.4 °C
    100 to 630 °C0.015%
    +0.18 °C
    0.03%
    +0.36 °C
    1 mA0.015%
    +0.18 °C
    0.03%
    +0.36 °C
    100 Ω Pt(3926)-200 to 100 °C0.2 °C0.4 °C1 mA0.2 °C0.4 °C
    100 to 630 °C0.015%
    +0.18 °C
    0.03%
    +0.36 °C
    1 mA0.015%
    +0.18 °C
    0.03%
    +0.36 °C
    10 Ω Cu(427)-100 to 260 °C1.5 °C3 °C1 mA1.5 °C3 °C
    120 Ω Ni(672)-80 to 260 °C0.15 °C0.3 °C1 mA0.15 °C0.3 °C
    50 Ω Cu(427)-180 to 200 °C0.4 °C0.7 °C1 mA0.4 °C0.7 °C
    100 Ω Cu(427)-180 to 200 °C0.2 °C0.4 °C1 mA0.2 °C0.4 °C
    YSI40015 to 50 °C0.2 °C0.4 °C250 μA0.2 °C0.4 °C
    1. Sensor inaccuracies not included.
    2. Resolution: 0.1 °C.
    3. Read accuracy is based on 4-wire input. For 3-wire RTD measurements, assuming all three RTD leads are matched, add 1.0 °C (Pt10 and Cu10), 0.6 °C (Pt50 and Cu50), 0.4 °C (Other RTD types) to the specifications.
    4. Source Accuracy in source mode is based on 0.5 mA~3 mA (1.00 Ω~400.00 Ω), 0.05 mA~0.8 mA (400.0 Ω~1500.0 Ω), 0.05 mA~0.4 mA (1500.0 Ω ~4000.0 Ω), excitation current (0.25 mA for Pt1000 range).
    5. Temperature Coefficient: ±0.05 °C /°C for measure, ± 0.05 °C /°C (<18 °C or >28 °C) for source.