We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The Fluke 710 is designed to enable users to perform quick, easy tests on HART smart control valves. With the built in HART communication function, users can source a 4-20 mA signal to cause the smart control valve to move, while simultaneously interpreting the valve’s HART feedback signal to determine whether the valve is moving to the expected position.
Functions | mA source, mA simulate, mA read, mA read/loop power, and volts read. |
Ranges | mA (0 to 24 mA) and Volts (0 to 30 V DC) |
Resolution | 1 µA on mA ranges and 1 mV on voltage range |
Accuracy | 0.01% ± 2 counts, all ranges (@ 23 °C ± 5 °C [73.4 °F ± 41 °F]) |
Operating temperature range | -10°C to 55°C (14°F to 131°F) |
Humidity range | 10 to 95% non-condensing |
Stability | 20 ppm of F.S. /°C from -10°C to 18°C and 28°C to 55°C |
Display | 128 x 64 pixels, LCD Graphic w/backlight, .34 high digits |
Power | 6 AAA alkaline |
Battery life | 40 hours continuous use (measure mode using) |
Loop compliance voltage | 24 V DC @ 20mA |
Over-voltage protection | 240 V AC |
Overload current protection | 28 mA DC |
EMC | EN61326 Annex A (Portable Instruments) |
Dimensions (L x W x D) | 15 x 9 x 3 cm (6 x 3.6 x 1.3 in) |
Weight | 0.3 kg (9.5 oz) |
Included accessories | Traceable calibration certificate with data, batteries, test leads, USB upload cable and safety |
Warranty | Three year |
Write Your Own Review