We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The Fluke 107 is a compact, simple to operate digital multimeter constructed to fit the way you work. It is the perfect first-pass troubleshooting tool: small enough to fit in a shirt pocket or work bag, durable enough to withstand daily use, and extremely easy to use. All without compromising safety. For everyday applications, where True-RMS accuracy isn’t needed, the Fluke 107 delivers reliable measurements continuously.
Accuracy is available for one year after calibration, operating from 18 °C to 28 °C and relative humidity from 0% to 75%. The format of the accuracy specification is: ± ([% of reading] + [least significant digit])
Precision specifications | ||||
AC voltage (40 Hz to 500 Hz) 1) | Range | Resolution | Accuracy | |
6.000 V 60.00 V 600.0 V | 0.001 V 0.01 V 0.1 V | 1.0 % + 3 | ||
DC voltage | 6.000 V 60.00 V 600.0 V | 0.001 V 0.01 V 0.1 V | 0.5 % + 3 | |
Exchange millivolts | 600.0 mV | 0.1 mV | 3.0 % + 3 | |
Diode test 2) | 2.000 V | 0.001 V | 10 % | |
Resistance (ohm) | 400.0 Ω 4.000 kΩ 40.00 kΩ 400.0 kΩ 4.000 MΩ 40.00 MΩ | 0.1 Ω 0.001 kΩ 0.01 kΩ 0.1 kΩ 0.001 MΩ 0.01 MΩ | 0.5 % + 3 0.5 % + 2 0.5 % + 2 0.5 % + 2 0.5 % + 2 1.5% + 3 | |
Capacitor 3) | 50.00 nF 500.0 nF 5.000 μF 50.00 μF 500.0 μF 1000 μF | 0.01 nF 0.1 nF 0.001 μF 0.01 μF 0.1 μF 1 μF | 2 % + 5 2 % + 5 5 % + 5 5 % + 5 5 % + 5 5 % + 5 | |
Frequency 4) Hz (10 Hz – 100 kHz) | 50.00 Hz 500.0 Hz 5.000 kHz 50.00 kHz 100.0 kHz | 0.01 Hz 0.1 Hz 0.001 kHz 0.01 kHz 0.1 kHz | 0.1 % + 3 | |
Duty cycle 4) | 1 % to 99 % | 0.10 % | 1 % typical case 5) | |
AC current (40 Hz - 200 Hz) | 4.000 A 10.00 A | 0.001 A 0.01 A | 1.5 % + 3 | |
DC | 4.000 A 10.00 A | 0.001 A 0.01 A | 1.5 % + 3 | |
1) All AC current, frequency, and duty cycle are specified from 1% to 100% of the range. Input values below 1% of range are not specified. 2) Typically, the open circuit test voltage is 2.0 V and the short circuit current is < 0.6 mA. 3) The parameters do not include errors due to test lead capacitance and capacitor substrate (up to 1.5 nF in the 50 nF range). 4) All AC current, frequency, and duty cycle are specified from 1% to 100% of the range. Input values below 1% of range are not specified. 5) Typically, the frequency is 50 Hz or 60 Hz and the duty cycle is 10% to 90%. | ||||
Features | Overload protection | Input impedance (nominal value) | Common mode rejection ratio | Normal mode rejection ratio |
AC voltage | 600 V 1) | >10 MΩ <100 pF 2) | >60 dB at 50 Hz or 60 Hz at DC current | − |
Exchange millivolts | 600 mV | >1M, <100 pF | >80 dB at 50 Hz or 60 Hz | − |
DC voltage | 600 V 1) | >10 MΩ <100 pF | >100 dB at 50 Hz or 60 Hz at DC current | >60 dB at 50 Hz or 60 Hz |
1) Up to 6 x 105 V Hz 2) For medium voltage (alternating current), the input impedance is approximately 1 MΩ. | ||||
General technical indicators | ||||
The highest voltage applied to any terminal and ground | 600 V | |||
Display (LCD) | 6000 counts, updated three times per second | |||
Type of battery | Two AAA batteries (NEDA 24A, IEC LR03) | |||
Battery life | At least 200 hours | |||
Temperature | ||||
Operating temperature | 0 °C to 40 °C | |||
Storage temperature | -30 °C to 60 °C | |||
Relative humidity | ||||
Working humidity | No condensation below 10 °C ≤ 90 % at 10 °C to 30 °C; ≤ 75 % at 30 °C to 40 °C | |||
Working humidity, 40 MΩ range | ≤ 80 % at 10 °C to 30 °C; ≤ 70 % at 30 °C to 40 °C | |||
Altitude | ||||
Working elevation | 2000 m | |||
Storage altitude | 12,000 m | |||
Temperature coefficient | 0.1 X (specified accuracy) / °C (<18 °C or >28 °C) | |||
Current input fuse protection | 11A, 1000V fast-melting type, must use Fluke designated parts | |||
Volume (height x width x length) | 142 mm x 69 mm x 28 mm | |||
Weight | 200 g | |||
Protection level | IEC 60529: IP 40 | |||
Safety | IEC 61010-1: 600 V CAT III, pollution degree 2 | |||
Electromagnetic environment | IEC 61326-1: Portable | |||
Electromagnetic compatibility | Applicable only in Korea. Class A equipment (industrial broadcasting and communication equipment)1) | |||
1) This product meets the requirements of industrial (Class A) electromagnetic wave equipment, which should be noted by the seller or user. The device is suitable for use in a work environment, not a home environment. |
Write Your Own Review