The 30 Series is a wide range pressure calibrator that fits easily into your shirt pocket.
Its internal sensors give you 0.05% of reading accuracy and resist damage from overpressure.
ACCURACY-PRESSURE MEASUREMENT
0 to 110% of Range: | ± (0.05% of Reading + Floor Term) | Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. |
Floor Term: | 0.005% of FS or display resolution, whichever is greater | |
Vacuum: | ± (0.25% of Reading + Vacuum Floor Term) | |
Vacuum Floor Term: | 0.004 PSI for 16 and 36 PSI |
ACCURACY-MILLIAMPERES
Accuracy: | ± (0.025% of Reading + 0.001 mA) | Includes all effects of linearity, hysteresis, repeatability, temperature, and stability for one year. |
Range: | 0 to 55 mA | mA can be displayed as a percentage, where 0 to 100% corresponds to either 4 - 20 mA or 10 - 50 mA. |
Resolution: | 0.001 mA | Jacks are compatible with safety sheathed banana plugs. |
Connection: | 4 mm jacks, 19 mm (0.75") spacing |
|
OPERATING TEMPERATURE
Temperature Range: | 0 to 50°C (32 to 122°F) | Non-condensing. No change in accuracy over operating temperature range. Gauge must be zeroed to achieve rated specification. |
DISPLAY
Screen: | 2 lines of up to 6 digits, | LCD readable in sunlight. |
Display Rate: | 2 rdgs/sec (adjustable to 1, 3, or 3.7 rdgs/sec) | Changes to reading rate requires free ConFigM30 software. |
POWER
1 x 9V: | 90 hours, typical | Uses 1 alkaline 9V battery. Use USB interface reduces operating time. |
|
| Do not use AC power supply in hazardous locations. |
DATA/COMMUNICATION
Digital Interface: | micro-USB | Free Labview drivers and programming instructions available online. |
|
| Do not use USB in hazardous locations. |
ENCLOSURE
Model 31 ≤ 300 PSI: | 342 g (12.0 oz) | Weight includes batteries. |
*Model 31 ≤ 300 PSI: | 428 g (15.0 oz) | *SPSS optional manifold. |
Model 31 ≥ 1500 PSI: | 360 g (13.0 oz) | Nylon carry case included. |
Model 33: | 400 g (14.0 oz) |
|
*Model 33: | 485 g (17.0 oz) |
|
Housing: | Extruded aluminum |
|
Keypad and Labels: | UV resistant polyester |
|
SENSOR ≥ 1500 PSI
Wetted Materials: | 316 stainless steel | All welded, with a permanently filled diaphragm seal. |
Diaphragm Seal Fluid: | Dow Corning 200 |
|
Connection: | Bottom, 1/8" female NPT |
|
SENSOR ≤ 300 PSI
Wetted Materials: | 316 stainless steel, Viton* | *Viton o-ring is internal. |
| Anodized aluminum penetrated with PTFE (polytetrafluoroethylene) | All welded with a permanently filled diaphragm seal. |
Diaphragm Seal Fluid: | Dow Corning 200 |
|
Connection: | Bottom, 1/8" female NPT |
|
STORAGE TEMPERATURE
Temperature Range: | -20 to 70°C (-4 to 158°F) | Batteries should be removed if stored for more than one month. |
SPECIAL FEATURES
The following requires the use of our free ConfigM30 software.
User Defined Unit: | Allows the user to define and display any pressure units not included, or to use the gauge to display force, level, or other pressure related parameters. Reduce or increase displayed resolution. |
Remove: | Unwanted pressure units from Units button. |
Password Protect: | Changes to configuration or calibration factor(s). |
Save: | Configuration to a file. |
Copy: | Configuration to other calibrators. |
Dimensions
Optional Pressure Pump Systems
A includes System A (pneumatic) 0 to 30 psi | |
AH includes System A (pneumatic) 0 to 580 psi | |
B includes System B (pneumatic) -25 inHg to 30 psi | |
BHincludes System B (pneumatic) -25 inHg to 580 psi | |
C includes System C (oil) 0 to 3,000 psi | |
CH includes System C (oil) 0 to 5,000 psi | |
DO includes System D (oil) 0 to 5,000 psi | |
DW includes System D (water) 0 to 5,000 psi | |
EOincludes System E (oil) 0 to 10,000 psi | |
FOVincludes System F (oil) 0 to 15,000 psi | |
FWVincludes System F (water) 0 to 15,000 psi | |
GOincludes System G (oil) 0 to 15,000 psi | |
GWincludes System G (water) 0 to 15,000 psi | |
HOincludes System H (oil/pneumatic) -25 inHg to 580 psi and 0 to 5,000 psi |
*Not recommended for oxygen service.
- • Deadweight tester replacement
- • Flow computer calibration
- • Gauge calibration
- • Pressure regulator adjustment
- • Pressure transmitter calibration
Every 30 Series Digital Pressure Calibrator comes with a factory test report with test data at 5 temperatures, proving our gauges meet specifications throughout their full temperature range.