BK Precision BA8100 80V EIS Battery Analyzer

SKU
BA8100
Manufacturer:
B&K Precision
B&K Precision

Key Features:

  • Fixed frequency measurements from the front panel
  • Swept frequency using included software
  • Maximum input voltage of 80 V
  • Programmable DC and AC current settings
  • EIS frequency range of 50 mHz to 10 kHz
  • Impedance Z, phase angle θ, voltage and current measurements
  • Simple 4-wire test connection
  • LAN, USB (COM), and RS232 interfaces standard
$3,590.00

Secured and trusted checkout with

The BK Precision BA8100 uses the Electrochemical Impedance Spectroscopy (EIS) technology out of the research lab and into the industrial market for diagnosing single battery cells or battery stacks up to 80 V.

Note: All specifications apply to the unit after a temperature stabilization time of 30 minutes over an ambient temperature range of 23 °C ± 5 °C.

Input
Voltage0.5 V to 80 V
Power200 W maximum (dissipated by internal load)
Settings
CurrentDC0.5 A to 3 A (dissipated by internal load)
Modulating AC Current50 mArms to 300 mArms (10% of DCA typical)
Modulating FrequencyRange0.05 Hz to 10 kHz
Resolution (Auto)0.001 Hz to 0.1 Hz
Accuracy0.01% of setting
Displayed Parameters
PrimaryX, Z, V, I, Freq.
SecondaryRs, Rp, Xs, Xp, Cs, Cp, Ls, Lp, Q, D, and θ
Measurements

Voltage 4-wire
Range0 V to 80 V
Resolution0.1 mV
Accuracy (ACI = 0) (1)0.05% of reading + 0.05% of full scale

DC Sink Current
Range0 A to 3 A
Resolution0.1 mA
Accuracy (ACI = 0) (1)0.5% of reading + 0.1% of full scale
Resistance R Auto sampling interval (8 to 10 cycles) and auto range enabledDisplayed Digits5 digits including sign
Resolution1 μΩ
Accuracy (10 Hz to 10 kHz)±(0.5% of reading + 5 μΩ)
Impedance Z Auto sampling interval (8 to 10 cycles) and auto range enabledDisplayed Digits5 digits including sign
Resolution1 μΩ
Accuracy (10 Hz to 10 kHz)±(0.5% of reading + 5 μΩ)

Phase Angle θ
Range-180° to 180°
Resolution0.01°
Accuracy±0.3°
General
AC Input100 VAC to 240 VAC, 50/60 Hz
UUT Connector4-terminal banana jack
I/O InterfacesLAN, USB (Virtual COM), RS232
Operating Temperature32° F to 104° F (0° C to 40° C)
Storage Temperature14° F to 122° F (-10° C to 50° C)
Dimension (W x H x D)8.5 “ x 3.5” x 14.6” (215 x 90 x 370 mm)
Weight9.9 lbs (4.5 kg)
SoftwareBundled software for fixed or frequency sweep, Nyquist and Bode plots
Warranty3 Years
Standard AccessoriesPower cord, kelvin clip test leads, calibration certificate, calibration test fixture (TLC81)

(1) The accuracy of DC voltage and current should be under no modulation condition (i.e. AC current = 0).

Write Your Own Review