We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The BK Precision BA8100 uses the Electrochemical Impedance Spectroscopy (EIS) technology out of the research lab and into the industrial market for diagnosing single battery cells or battery stacks up to 80 V.
Note: All specifications apply to the unit after a temperature stabilization time of 30 minutes over an ambient temperature range of 23 °C ± 5 °C.
Input | ||
Voltage | 0.5 V to 80 V | |
Power | 200 W maximum (dissipated by internal load) | |
Settings | ||
Current | DC | 0.5 A to 3 A (dissipated by internal load) |
Modulating AC Current | 50 mArms to 300 mArms (10% of DCA typical) | |
Modulating Frequency | Range | 0.05 Hz to 10 kHz |
Resolution (Auto) | 0.001 Hz to 0.1 Hz | |
Accuracy | 0.01% of setting | |
Displayed Parameters | ||
Primary | X, Z, V, I, Freq. | |
Secondary | Rs, Rp, Xs, Xp, Cs, Cp, Ls, Lp, Q, D, and θ | |
Measurements | ||
Voltage 4-wire | Range | 0 V to 80 V |
Resolution | 0.1 mV | |
Accuracy (ACI = 0) (1) | 0.05% of reading + 0.05% of full scale | |
DC Sink Current | Range | 0 A to 3 A |
Resolution | 0.1 mA | |
Accuracy (ACI = 0) (1) | 0.5% of reading + 0.1% of full scale | |
Resistance R Auto sampling interval (8 to 10 cycles) and auto range enabled | Displayed Digits | 5 digits including sign |
Resolution | 1 μΩ | |
Accuracy (10 Hz to 10 kHz) | ±(0.5% of reading + 5 μΩ) | |
Impedance Z Auto sampling interval (8 to 10 cycles) and auto range enabled | Displayed Digits | 5 digits including sign |
Resolution | 1 μΩ | |
Accuracy (10 Hz to 10 kHz) | ±(0.5% of reading + 5 μΩ) | |
Phase Angle θ | Range | -180° to 180° |
Resolution | 0.01° | |
Accuracy | ±0.3° | |
General | ||
AC Input | 100 VAC to 240 VAC, 50/60 Hz | |
UUT Connector | 4-terminal banana jack | |
I/O Interfaces | LAN, USB (Virtual COM), RS232 | |
Operating Temperature | 32° F to 104° F (0° C to 40° C) | |
Storage Temperature | 14° F to 122° F (-10° C to 50° C) | |
Dimension (W x H x D) | 8.5 “ x 3.5” x 14.6” (215 x 90 x 370 mm) | |
Weight | 9.9 lbs (4.5 kg) | |
Software | Bundled software for fixed or frequency sweep, Nyquist and Bode plots | |
Warranty | 3 Years | |
Standard Accessories | Power cord, kelvin clip test leads, calibration certificate, calibration test fixture (TLC81) |
(1) The accuracy of DC voltage and current should be under no modulation condition (i.e. AC current = 0).
Write Your Own Review