BK Precision 9832B 2000 VA Programmable AC Power Source

SKU
9832B
Manufacturer:
B&K Precision
B&K Precision

Key Features:

  • AC, DC and AC+DC power source
  • 3-Phase capability using multiple AC sources and 3-Phase kit (TL983P-KIT)
  • Low total harmonic distortion meets the IEC 61000-3-2 standard
  • Comprehensive measurement capabilities Vrms, Arms, Vdc, +Apk, -Apk, inrush current, frequency, power factor, apparent power, reactive power, true power, and crest factor
  • 0.98 power factor at AC input stage
  • Built-in standard waveforms sine, square and clipped sine
  • 30 built-in THD waveforms
  • Amplifier mode with 1.2 kHz bandwidth for generating user-defined arbitrary waveforms
  • Step, List and Pulse modes for generating power line disturbance (PLD) simulations
  • List mode supports 10 user-defined programs with up to 100 programmable steps
  • Generate custom harmonic waveforms on a PC and download them to the instrument’s 5 non-volatile memory locations
  • Digital I/O port supporting external trigger, transient indication, failure status indication, remote inhibit, RS232, and external analog output level programming interface
  • Comprehensive protection modes OVP, OCP,OPP, OTP, fan failure, output timer and key lock
  • Downloadable LabVIEWTM driver and soft panel software for remote control
  • Standard USB (USBTMC-compliant), RS232, LAN, and GPIB remote interfaces
Product Options:
$6,850.00

Secured and trusted checkout with

The 9830B Series programmable AC power sources provide high performance and low total harmonic distortion in a 3U form factor. The addition of positive and negative DC offset voltages expands the AC capabilities to operate in DC and AC+DC output coupling modes. The user can select built-in and user-defined harmonic waveforms or select from standard sine, square or clipped sine outputs.

3-Phase power can be achieved by connecting 3 units of the same model in a master and slave configuration using the optional 3-Phase sync adapters. The frequency range is adjustable from 45 to 600 Hz in 3-phase mode.

The 9832B max voltage is 300 V, max current is 20 A, and max power is 2000 VA.

Model

9832B

9833B

AC Output

Output Phase

Single

Maximum Power

2000 VA

3000 VA

Voltage

Range (1) (rms)

Low

0 to 150 V

High

0 to 300 V

Current (rms)

Low

20 A

30 A

High

10 A

15 A

 

Current (peak)

Low

65 A (< 100 Hz)

50 A (> 100 Hz)

97.5 A (< 100 Hz)

75 A (> 100 Hz)

High

32.5 A (< 100 Hz)

25 A (> 100 Hz)

48.75 A (< 100 Hz)

37.5 A (> 100 Hz)

Frequency

Range

Single phase

45 Hz to 1.2 kHz

3-Phase

45 Hz to 600 Hz

Phase Range

0 to 359.7 º

 

Total Harmonic Distortion (2)

45 Hz to 400 Hz

0.5 %

> 400 Hz to 1 kHz

1 %

> 1 k to 1.2 kHz

2 %

Line Regulation (3)

0.1 %

Load Regulation (3)

0.1 %

Temp. Coefficient

0.2 % per ºC

Crest Factor

45 Hz to 100 Hz

3.25

100 Hz to 1.2 kHz

2.5

Efficiency (4)

80 % (typical)

DC Output

Maximum Power

1000 W

1500 W

Voltage Range (1)

Low

0 to ± 212 V

High

0 to ± 424 V

Current

Low

10 A

15 A

High

5 A

7.5 A

Ripple and Noise (20 Hz to 20 MHz)

≤ 300 mVrms  /  ≤ 3 Vpp

Output Characteristics

Transient Response Time

1.5 ms (typical)

Output Impedance

≤ 1 Ω

Programming

 

Resolution

Voltage

0.1 V

Phase

0.1 º

Frequency

0.01 Hz (< 100 Hz), 0.1 Hz (> 100 Hz)

 

 

Accuracy

 

Voltage

AC

0.2 % + 0.2 % F.S.

DC

0.2 % + 0.4 % F.S.

Phase

0.15 %

Frequency

± 1 % (45 Hz to 100 Hz)

 

Measurement

 

Resolution

Voltage

0.1 V

Current

0.01 A

Power

0.01 W

Frequency

0.01 Hz (< 100 Hz)

0.1 Hz (> 100 Hz)

 

 

 

Accuracy

Voltage

AC

0.25 % + 0.25 % F.S.

DC

0.25 % + 0.5 % F.S.


Current

AC

0.25 % + 0.375 % F.S. (rms)

0.4 % + 0.75 % F.S. (Peak)

0.25 % + 0.25 % F.S. (rms)

0.25 % + 0.5 % F.S. (Peak)

DC

0.25 %+3 % F.S

0.25 %+2% F.S

Power

1 %  of F.S. for frequency ≤ 500 Hz

2 % of F.S. for frequency > 500 Hz

Frequency

0.5 %

AC Input

Voltage

190 V to 250 V

Frequency

47 Hz to 63 Hz

Maximum Power

2500 VA

3800 VA

Maximum Current

13.2 A

20 A

Power Factor

0.98 (typical)

General

 

Analog

BNC Input

Input

Voltage Range

0 to ± 12.5 V

Input Impedance

200 kΩ

Bandwidth

1.2 kHz

 

Storage Memory

10 programs,  up to 100 steps total (List mode)

5 memory locations for user-defined  waveforms

9 instrument settings

Remote Interface

Analog programming (5),  USB (USBTMC or virtual COM), RS232 (5), GPIB, and Ethernet

Command Response time

50 ms

Protection

OVP, OCP, OPP, OTP

Operating Temperature

32 °F to 104 °F (0 °C to 40 °C)

Storage Temperature

-40 °F to 185 °F (-40 °C to 85 °C)

Environmental Conditions

≤ 80% Relative Humidity up to 35 ºC, non-condensing

Dimensions  (W x H x D)

16.5” x 5.2” x 22” (420 x 132 x 560 mm)

Weight

52.9 lbs (24 kg)

Warranty

3 Years

Included Accessories

AC power cord with input connector, test report &

certificate of calibration

Optional Accessories

Rackmount ears with handles (RK3U)  &

3-Phase sync adapter  (TL983P), 3-Phase kit (TL983P-KIT)

Note: All specifications apply to the unit after a temperature stabilization time of 15 minutes over an
ambient temperature range of 23 °C ± 5 °C.
(1) - The maximum voltage is limited to 310 Vrms and ± 438 Vdc, 3 Phase mode not supported
(2) - > 66% to full range
(3) - AC mode with sine wave and remote sense enabled
(4) - 150 VAC (150 V range) and 300 VAC (300 V range) with nominal input AC voltage.
(5) - Analog programming pin available on digital I/O connector, not available in 3-Phase

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account
  • Pre-compliance testing
  • Simulate common grid faults, voltage sags, frequency, and phase disturbances, according to IEC61000-4-11/14/28/34
  • Electromagnetic compatibility (EMC), according to IEC61000-3-2
  • Evaluate transformers, TRIACs, SCRs, and passive components

Products In This Series

As low as

Now $8,620.00