BK Precision 8600 150W Programmable DC Electronic Load

SKU
8600
Manufacturer:
B&K Precision
B&K Precision

Key Features:

  • Wide voltage and current range, 0 to 500 V, 0 to 240 A
  • CC/CV/CR/CW operating modes
  • Transient mode up to 25 kHz
  • 16-bit resolution voltage and current measurement
  • List mode function
  • Adjustable slew rate in CC mode
  • Short circuit test
  • Built-in battery test function with voltage level, capacity level, and timer stop conditions
  • CR-LED mode to simulate the loading behavior of typical LEDs
  • Remote voltage sensing to compensate for the effect of voltage drop in wires
  • Analog current control and monitoring
  • Standard USB (USBTMC-compliant), RS-232, and GPIB interfaces with SCPI protocol support
  • OVP/OCP/OPP/OTP and local/remote reverse voltage (LRV/RRV) protection
Product Options:
$1,545.00

Secured and trusted checkout with

BK Precision’s 8600 series offers the performance of a modular system DC electronic load in a more effectual benchtop form. The 8600 is a 150W DC electronic load with the ability to function in constant voltage (CV), constant current (CC), constant resistance (CR) and constant power (CW) modes. Its transient operation speeds reach up to 25 KHz, while still maintaining a high 16 bit measurement resolution and accuracy. Flexible triggering capabilities, analog current control and monitoring, list mode function and remote voltage sensing are some of the many functions and features the 8600 boasts. The electronic load is also equipped with standard USB (USBTMC-complaint), RS-232 and GPIB serial interfaces for the power to communicate remotely.

The 8600 150W DC electronic load by BK Precision is ideal for testing and evaluation an expanse of DC sources like photovoltaic arrays and battery chargers.

Model

8600

8601

8602

8610

8612

8614

8616

Input ratings

Input voltage

0 – 120 V

0 – 120 V

0 – 500 V

0 – 120 V

0 – 500 V

0 – 120 V

0 – 500 V

Input

current

Low

0 – 3 A

0 – 6 A

0 – 3 A

0 – 12 A

0 – 3 A

0 – 24 A

0 –6 A

High

0 – 30 A

0 – 60 A

0 – 15 A

0 – 120 A

0 – 30 A

0 – 240 A

0 –60 A

Input power

150 W

250 W

200 W

750 W

1500 W

1200 W

Minimum

operating

voltage

Low

0.11 V at 3 A

0.18 V at 6 A

1 V at 3 A

0.12 V at 12 A

0.36 V at 3 A

0.15 V at 24 A

0.36 V at 6 A

High

1.1 V at 30 A

1.1 V at 60 A

4.5 V at 15 A

1.2 V at 120 A

3.6 V at 30 A

1.5 V at 240 A

3.6 V at 60 A

CV mode

 

Range

Low

0 – 18 V

0 – 50 V

0 – 18 V

0 – 50 V

0 – 18 V

0 – 50 V

High

0 – 120 V

0 – 500 V

0 – 120 V

0 – 500 V

0 – 120 V

0 – 500 V

 

Resolution

Low

0.1 mV

1 mV

0.1 mV

1 mV

0.1 mV

1 mV

High

1 mV

10 mV

1 mV

10 mV

1 mV

10 mV

 

 

Accuracy

Low

±(0.05%+

0.02%  FS)

±(0.025%+

0.05%  FS)

±(0.05%+

0.025%  FS)

±(0.025%+0.05% FS)

±(0.025%+

0.025%  FS)

±(0.025%+

0.05%  FS)

High

±(0.05%+

0.025%  FS)

±(0.025%+

0.05%  FS)

±(0.05%+

0.025%  FS)

±(0.025%+0.05% FS)

CC mode

 

Range

Low

0 – 3 A

0 – 6 A

0 – 3 A

0 – 12 A

0 – 3 A

0 – 24 A

0 – 6 A

High

0 – 30 A

0 – 60 A

0 – 15 A

0 – 120 A

0 – 30 A

0 – 240 A

0 – 60 A

 

Resolution

Low

0.01 mA

0.1 mA

1 mA

0.1 mA

1 mA

0.1 mA

High

0.1 mA

1 mA

10 mA

1 mA

10 mA

1 mA

 

Accuracy

Low

±(0.05%+0.05% FS)

±(0.05%+0.1% FS)

±(0.05%+0.05% FS)

±(0.05%+0.1% FS)

±(0.05%+0.05% FS)

High

±(0.05%+0.05% FS)

±(0.05%+0.1% FS)

±(0.05%+0.05% FS)

±(0.05%+0.1% FS)

±(0.05%+0.05% FS)

CR mode

 

Range

Low

0.05 Ω – 10 Ω

0.3 Ω – 10 Ω

0.02 Ω – 10 Ω

0.15 Ω – 10 Ω

0.01 Ω – 10 Ω

0.01 Ω – 10 Ω

High

10 Ω - 7.5 kΩ

Resolution

16 bit

 

Accuracy

Low

0.01%+0.08 S

High

0.01%+0.0008 S

CW mode

Range

150 W

250 W

200 W

750 W

1500 W

1200 W

Resolution

10 mW

100 mW

Accuracy

0.1% + 0.1% FS

0.2% + 0.2% FS

0.1% + 0.1% FS

0.2% + 0.2%  FS

Transient mode (CC mode)

T1 & T2 (1)

20 µs – 3600  s / Resolution: 10 µs

Accuracy

5 µs + 100 ppm

 

Slew Rate (2)

Low

0.001-2.5 A/ms

0.001-1 A/ms

0.001-0.25 A/µs

0.0001-0.1 A/µs

0.001-0.25 A/µs

0.0001-0.1 A/µs

High

0.001-2.5 A/µs

0.001-1 A/µs

0.01-2.5 A/µs

0.001-1 A/µs

0.01-2.5 A/µs

0.001-1 A/µs

(1) Fast pulse trains with large transitions may not be achievable.
(2) The slew rate specifications are not warranted, but are descriptions of typical performance. The actual transition time is defined as the time for the input to change from 10% to 90%, or vice versa, of the
programmed current values. In case of very large load changes, e.g. from no load to full load, the actual transition time will be larger than the expected time. The load will automatically adjust the slew rate to fit within the range
(high or low) that is closest to the programmed value.

 

 

Model

8600

8601

8602

8610

8612

8614

8616

Readback voltage

 

Range

Low

0 – 18 V

0 – 18 V

0 – 50 V

0 – 18 V

0 – 50 V

0 – 18 V

0 – 50 V

High

0 – 120 V

0 – 120 V

0 – 500 V

0 – 120 V

0 – 500 V

0 – 120 V

0 – 500 V

 

Resolution

Low

0.1 mV

1 mV

0.1 mV

1 mV

0.1 mV

1 mV

High

1 mV

10 mV

1 mV

10 mV

1 mV

10 mV

Accuracy

±(0.05%+0.05% FS)

Readback current

 

Range

Low

0 – 3 A

0 – 6 A

0 – 3 A

0 – 12 A

0 – 3 A

0 – 24 A

0 – 6 A

High

0 – 30 A

0 – 60 A

0 – 15 A

0 – 120 A

0 – 30 A

0 – 240 A

0 – 60 A

 

Resolution

Low

0.01 mA

0.1 mA

0.01 mA

1 mA

0.1 mA

1 mA

0.1 mA

High

0.1 mA

1 mA

0.1 mA

10 mA

1 mA

10 mA

1 mA

 

Accuracy

±(0.05%+

0.05%  FS)

±(0.05%+

0.1% FS)

±(0.05%+

0.05%  FS)

±(0.05%+

0.1% FS)

±(0.05%+

0.05%  FS)

±(0.05%+

0.1% FS)

±(0.05%+

0.05%  FS)

Readback power

Range

150 W

250 W

200 W

750 W

1500 W

1200 W

Resolution

10 mW

100 mW

Accuracy

±(1%+0.1% FS)

±(0.2%+0.2% FS)

±(0.1%+0.1% FS)

±(0.2%+0.2% FS)

Protection range (typical)

OPP

 150 W

 250 W

 200 W

760 W

1550 W

1250 W

 

OCP

Low

 3.3 A

 6.6 A

 3.3 A

13.2 A

3.3 A

26.4 A

6.6 A

High

 33 A

 66 A

 16.5 A

132 A

33 A

264 A

66 A

OVP

 120 V

 120 V

 500 V

130 V

530 V

130 V

530 V

OTP

185 ºF (85 °C)

General (typical)

Short circuit

 

Current (CC)

Low

3 A

6 A

3 A

12 A

3 A

24 A

6 A

High

30 A

60 A

15 A

120 A

30 A

240 A

60 A

Voltage (CV)

0 V

Resistance (CR)

 35 mΩ

 30 mΩ

 300 mΩ

10 mΩ

120 mΩ

6 mΩ

60 mΩ

Input terminal impedance

150 kΩ

300 kΩ

1 MΩ

300 kΩ

1 MΩ

300 kΩ

1 MΩ

AC input

110 V/220 V ±10%, 50/60 Hz

Operating temperature

32 ºF to 104 ºF (0 °C to 40 °C)

Storage temperature

14 ºF to 140 ºF (-10 °C to 60 °C)

Humidity

Indoor use, ≤ 95%

Safety

EN61010-1:2001, EU Low Voltage Directive 2006/95/EC

Electromagnetic

compatibility

Meets EMC Directive 2004/108/EC,  EN 61000-3-2:2006, EN 61000-3-3:1995+A1:2001+A2:2005

EN 61000-4-2/-3/-4/-5/-6/-11, EN 61326-1:2006

Dimensions (W x H x D)

8.5” x 3.5” x 15.2” (218  x 90 x 387 mm)

19” x 5.8” x 24.5” (485  x 147 x 621 mm)

Weight

9.9 lbs (4.5  kg)

54 lbs (24.6  kg)

Three-Year Warranty

Standard accessories

User manual, power cord

Optional accessories

TLPWR1 high current test leads, IT-E151 rackmount kit (models 8600, 8601, and 8602 only)

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account

Products In This Series

As low as

Now $1,850.00

As low as

Now $1,915.00

As low as

Now $3,230.00

As low as

Now $4,005.00