Ashcroft IXLdp Industrial Ultra-Low Variable Capacitance Pressure Transducer/Transmitter

SKU
IXLDP
Manufacturer:
Ashcroft
Ashcroft

Key Features:

  • Certified 0.25% and 0.5% accuracy
  • 0.1"- 200" - H2O pressure ranges
  • High overpressure protection
  • FM approved for hazardous locations
  • NEMA 4X metal construction
  • Six types of output signals available
  • Easy installation
  • On-board voltage regulation allows use of lower cost, non-precise, unregulated power supply
Product Options:
$806.27

Secured and trusted checkout with

The Ashcroft® Industrial IXLdp was designed for the measurement and control of very low pressure and flow in industrial and process plant environments. The Industrial IXLdp transmitter features a rugged NEMA 4X enclosure, built-in electrical terminal box isolated from the electronics and threaded process connections.

The Ashcroft IXLdp transmitter utilizes a state-of-the-art variable capacitance sensor with a glass-clad silicon chip. The Si-Glas technology combines the inherent high sensitivity of a variable capacitance transducer with the repeatability of a micro-machined, single-crystal silicon diaphragm.

The Si-Glas sensor is composed of sputtered metals and glass molecularly bonded to silicon. There are no epoxies or other organics in the sensor to contribute to drift or mechanical degradation over time.

PERFORMANCE SPECIFICATIONS
Ref. Temperature: 70°F ±2°F (21°C ±1°C)
Accuracy Class (of Span)
Two Options: ±0.25%, ±0.5%
Includes non-linearity (Terminal Point Method), hysteresis,
non-repeatability, zero offset and span setting errors.
Stability – Max. Change (Span/year): ±0.25 %
Standard Ranges (Inches W.C.)
Unidirectional Ranges:
Differential or Gauge
0/0.10      0/2.00      0/10.00      0/ 50.00
0/0.25      0/2.50      0/15.00      0/100.00
0/0.50      0/3.00      0/20.00      0/150.00
0/1.00      0/5.00      0/25.00      0/200.00
Bidirectional Ranges:
Compound
±0.05      ±0.50      ± 3.00      ± 20.00
±0.10      ±1.00      ± 5.00      ± 25.00
±0.20      ±2.00      ±10.00     ± 50.00
±0.25      ±2.50      ±15.00     ±100.00
Custom Ranges:
Special range calibrations – see options
Response Time: Standard: 250ms
Optional variable damping (0-30 sec) – see options
ENVIRONMENTAL SPECIFICATIONS
Temperature Limits:
Storage: –40 to 210°F
Operating: –20 to 185°F
XFM Operating: –4 to 104°F
Compensated: 0 to 160°F
Relative Humidity: 0 to 95%
Thermal Coefficients:
             0.25% Acc.          0.5% Acc.
ZERO ±0.01% Span/°F   ±0.02% Span/°F
SPAN ±0.01% Span/°F   ±0.02% Span/°F
Vibration Sweep:
Less than 0.2% Span/g temporary effect 10-130 Hz
FUNCTIONAL SPECIFICATIONS
Overpressure Limits:
Proof differential pressure:         20 psid
Burst differential pressure:         50 psid
Maximum static (line) pressure: 100 psi
Static pressure effect: less than 0.5% Span
Mounting Position Effect:
1˝ W.C. and higher     0.1% Span/g
0.25˝ up to 0.5” W.C.  0.5% Span/g
0.1˝ W.C.                   0.8% Span/g
Note: Calibrated horizontally standard unless
otherwise specified. Mounting Position Effect easily
corrected with zero potentiometer.
ELECTRICAL SPECIFICATIONS
Output Signal:                    Power:
4-20 mA (2 wire)                  12-36Vdc
0-5 Vdc (3 wire)                   12-36Vdc
1-5 Vdc (3 wire)                   12-36Vdc
1-6 Vdc (3 wire)                   12-36Vdc
±2.5 Vdc (3 wire)                 12-36Vdc
±5 Vdc (3 wire)                    12-36Vdc
Output Signal is Independent of Power Supply
Changes: 12-36 Vdc range without effect on
output signal
Reverse Wiring Protected
Internal Zero and Span: ±10% F.S. Adjustment
Supply Current: 2.6mA typical for voltage output
Warm-up Time:
Full specification: Less than one second
Fast Response, Turndown & Variable Dampening
Optional
PHYSICAL SPECIFICATIONS
Enclosure:
NEMA 4X, 300 series stainless steel
Process Connections: Two 1⁄4 NPT female
Electrical Connections: Two 1⁄2˝ female electrical
conduit connections isolated from the electronics.
Separate access cover for terminal connections
Media: Clean, dry and noncorrosive gas (consult
factory for use on other media)
NOT FOR USE ON LIQUIDS

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account
  • HVAC
  • Fume Hood Control
  • Lab/clean Room Pressurization
  • Laminar Flow
  • Furnace/Stack Draft
  • Leak Detection
  • Medical
  • Fan Tracking
  • Filter Monitoring
  • Glovebox
  • Velocity Measurements