AC/DC Current Probe Model SL206 measures low DC or AC from 10mA to 80A. This battery-powered clamp-on probe may be used with digital multimeters, voltmeters, or other voltage-measuring instruments. The probe jaws clamp around the conductor under test, allowing current measurements without breaking the circuit. Hall sensor technology senses the magnetic field produced by the current measured, and generates a millivolt DC or AC output signal. The narrow jaw is designed for optimum use in crowded wiring in industrial and automotive environments.
ELECTRICAL | |
Nominal Range | 2A & 80A |
Measurement Range | 10mA to 80ADC; 50mA to 60AAC |
Transformation Ratio | Voltage output |
Output Signal | 2A: 1mV/mA 80A: 10mV/A |
Accuracy (1mV/mA Range) 10mA to 2AAC/DC | 2% of Reading ± 5mA |
Phase Shift (1mV/mA Range) | DC to 65Hz 1° |
Accuracy 50mA to 50ADC | 4% of Reading ± 20mA |
50mA to 40AAC | ±4% of Reading ± 20mA |
50 to 80ADC | ±12% of Reading |
40 to 60AAC | ±12% of Reading |
Phase Shift (1mV/mA Range) | DC to 65Hz 1° |
Overload | 120A continuous |
Frequency Range | DC to 8kHz |
Load Impedance | 100kW (1mV/mA); 10kW (1V/A) |
Working/Common Mode Voltage | 600Vrms |
Output Termination | 5 ft Lead with 4 mm safety banana plug |
MECHANICAL | |
Operating Temperature | 32° to 122°F (0° to 50°C) |
Storage Temperature | -22° to 176°F (-30° to 80°C) |
Operating Relative Humidity | 10° to 30°C: 85 ± 5% RH (without condensation) 40° to 50°C: 45 ± 5% RH (without condensation) |
Maximum Conductor Size | 0.46" (11.8mm) |
Dimensions | 9.09 x 1.42 x 2.64" (231 x 36 x 67mm) |
Weight | 11.6 oz (330g) with battery |
Polycarbonate Material | Handle: Lexan® 920A, UL94 V2 |
SAFETY | |
Electrical | EN 61010-2-32 |
Double Insulation | Yes |
CE Mark | Yes |
Note: Reference conditions: 23°C ± 5°K, 20 to 75% RH, 48 to 65Hz, external magnetic field <40A/m, no external current carrying conductor, test sample centered, load impedance 1Mohm.