We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The MN375 is a compact AC current probe. It is designed to meet the most stringent demands in industry and electrical contracting. The Model MN375 meets EN 61010, 600V Cat. III safety standards and is CE marked and UL approved. The Model MN375 has a 0.78" (20mm) jaw opening and will accommodate conductors up to 250MCM. The unique hooked jaw design facilitates clamping and hooking into conductors. Polycarbonate materials and ultrasonic welding are used throughout to ensure ruggedness and overall lasting reliability. The Model MN375 is a high performance models and uses a nickel core to improve accuracy and phase shift on the low end of the measurement range. The Model MN375 is an excellent probe to be used in power and low level current measurement application.
ELECTRICAL | |
Nominal Range | 10A |
Measurement Range | 0.1 to 10A |
Transformation Ratio | Voltage output |
Output Signal | 100mV/A (1V @ 10A) |
Accuracy (10A Range) 0.1 to 1A |
1% of Reading ± 2mV |
Phase Shift (10A Range) 0.1 to 1A |
Not Specified |
Overload | – |
Frequency Range | 40 to 3kHz |
Limit Operating | 20A permanently |
Crest Factor | – |
Working/Common Mode Voltage | 600Vrms |
Output Termination | 5 ft Lead |
MECHANICAL | |
Operating Temperature | 14° to 131°F (-10° to 55°C) |
Storage Temperature | -40° to 158°F (-40° to 70°C) |
Operating Relative Humidity | 10 to 35°C 85% RH (without roll-off above 35°C) |
Jaw Opening | 0.83" (21mm) |
Maximum Conductor Size | 0.78" (20mm) |
Dimensions | 5.47 x 2.00 x 1.18" (139 x 51 x 30mm) |
Weight | 6.5 oz (180g) |
Polycarbonate Material | Polycarbonate with fiberglass charge, UL94 V0 |
SAFETY | |
Electrical | EN 61010-2-32 |
UL Approval | Yes – Canada and United States |
Double Insulation | Yes |
CE Mark | Yes |
Note: Reference conditions: 20 to 26°C, 20 to 75% RH, external magnetic field <40A/m, 48 to 65Hz sine wave, distortion factor less than 1%, no DC component, no external current carrying conductor, test sample centered. Load impedance >1MΩ
Write Your Own Review