We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
MN353 is a high accuracy and low phase shift AC current probe with voltage output. It is designed for use with DMMs or current measuring instruments with voltage inputs. The voltage output is proportional to the current measured (through the jaws). The MN353 has an integral 5 ft (1.5m) lead terminated by shrouded banana plugs.
ELECTRICAL | |
Nominal Range | 150A |
Measurement Range | 0.1 to 150A |
Transformation Ratio | Voltage output |
Output Signal | 10mV/A (1V @ 100Aac) |
Accuracy (150A Range) 0.1 to 1A |
2% of Reading ± 0.2mV |
1 to 80A | 1% of Reading ± 0.2mV |
80 to 150A | 1.5% of Reading ± 0.2mV |
Phase Shift (150A Range) 1 to 20A |
≤3.0° |
20 to 100A | ≤2.0° |
100 to 150A | ≤2.5° |
Overload | 240A for 10 min ON, 30 min OFF |
Frequency Range | 40 to 10kHz |
Limit Operating | 200A permanently to 1kHz; Derating above 3kHz: 200A x (1/0.333F), F in kHz to 1kHz |
Crest Factor | 3 @ 150A Peak with an error (due to CF) of 3% |
Working/Common Mode Voltage | 600Vrms |
Output Termination | 5 ft Lead |
MECHANICAL | |
Operating Temperature | 14° to 131°F (-10° to 55°C) |
Storage Temperature | -40° to 158°F (-40° to 70°C) |
Operating Relative Humidity | 10 to 35°C 85% RH (without roll-off above 35°C) |
Jaw Opening | 0.83" (21mm) |
Maximum Conductor Size | 0.78" (20mm) |
Dimensions | 5.47 x 2.00 x 1.18" (139 x 51 x 30mm) |
Weight | 6.5 oz (180g) |
Polycarbonate Material | Polycarbonate with fiberglass charge, UL94 V0 |
SAFETY | |
Electrical | EN 61010-2-32 |
UL Approval | Yes – Canada and United States |
Double Insulation | Yes |
CE Mark | Yes |
Note: Reference conditions: 20 to 26°C, 20 to 75% RH, external magnetic field <40A/m, 48 to 65Hz sine wave, distortion factor less than 1%, no DC component, no external current carrying conductor, test sample centered. Load impedance >1mΩ
Write Your Own Review