We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
Amprobe 33XR-A is a manual ranging digital multimeter with temperature and capacitance for field service technicians.
FEATURE | 33XR-A |
DC Voltage | 400.0 mV, 4.000 V, 40.00 V, 400.0 V, 1000 V |
AC Voltage | 400.0 mV, 4.000 V, 40.00 V, 400.0 V, 1000 V |
DC Current | 400.0 µA, 4.000 mA, 40.00 mA, 300.0 mA, 10 A |
AC Current | 400.0 µA, 4.000 mA, 40.00 mA, 300.0 mA, 10 A |
Resistance | 400.0 Ω, 40.00 kΩ, 4.000 MΩ |
Continuity | • |
Diode | • |
VolTect™ Non-Contact Voltage Detection | - |
Capacitance | 4.000 µF, 40.00 µF, 400.0 µF, 4000 µF |
Frequency | 4.000 kHz, 40.00 kHz, 400.0 kHz, 1 MHz |
Temperature | 1832 °F (1000 °C) |
Duty cycle | – |
Logic test | – |
Inductance | – |
dBm | – |
Specifications | |
DC Voltage ranges: | 400 mV, 4 V, 40 V, 400 V, 1000 V |
DC Voltage accuracy: | all ranges: ± (0.7 % rdg + 1 dgt) |
DC Voltage resolution: | 100 µV |
DC Voltage input impedance: | 10 MΩ |
AC Voltage (45 Hz to 500 Hz) ranges: | 400 mV, 4 V, 40 V, 400 V, 750 V |
AC Voltage (45 Hz to 500 Hz) accuracy: | 4 V Range, 200 Hz to 500 Hz: ± (2.0 % rdg + 4 dgts) |
AC Voltage (45 Hz to 500 Hz) accuracy: | All other ranges: ± (1.5 % rdg + 4 dgts) |
AC Voltage (45 Hz to 500 Hz) resolution: | 100 µV |
AC Voltage (45 Hz to 500 Hz) input impedance: | 10 MΩ |
DC Current ranges: | 400 µA, 4 mA, 40 mA, 300 mA, 10 A |
DC Current accuracy: | 400 µA to 300 mA ranges: ± (1.0 % rdg + 1 dgt) |
DC Current accuracy: | 10 A range: ± (2.0 % rdg + 3 dgts) |
DC Current resolution: | 0.1 µA |
DC Current voltage burden: | 400 µA range: 1 mV/µA, 4 mA range: 100 mV/1 mA, 40 mA range: 12 mV/1 mA, |
DC Current voltage burden: | 300 mA range: 4 mV/1 mA, 10 A range: 100 mV/1 A |
AC Current (45 Hz to 500 Hz) ranges: | 400 µA, 4 mA, 40 mA, 300 mA, 10 A |
AC Current (45 Hz to 500 Hz) accuracy: | 400 µA to 300 mA ranges: ± (1.5 % rdg + 4 dgts) |
AC Current (45 Hz to 500 Hz) accuracy: | 10 A range: ± (2.5 % rdg + 4 dgts) |
AC Current resolution: | 0.1 µA |
AC Current voltage burden: | See DC Current |
Resistance ranges: | 400 Ω 40 kΩ 4 MΩ |
Resistance accuracy: | 400 Ω and 40 kΩ ranges: ± (1.0 % rdg + 4 dgts) |
Resistance accuracy: | 4 MΩ range: ± (1.2 % rdg + 4 dgts) |
Resistance resolution: | 100 mΩ |
Write Your Own Review