Functional Features
|
Scaling
|
User can convert measured current, voltage and frequency values into pressure, temperature and flow values. Three conversion functions available: linearity, square and square root
|
Filtering
|
Average sliding filter (sampling number: 1 ~ 50) First-order linear filter (First-order coefficient: 0.01 ~ 1)
|
Switch
|
The measurement value will be automatically displayed at the moment the switch change of state. The latest 8 state changes will be stored in the memory.
|
Pressure Tare
|
Tare value is set through the user interface
|
Pressure stability indicator
|
Stability time and criteria is selectable
|
General Specifications
|
Input Channels
|
Top: 1 electrical signal measurement channel, φ4mm banana jacks
|
Right side: 2 channels for external digital pressure modules, 5-core dedicated aviation plug
|
Bottom: embedded digital pressure module (model ADT158Ex), field switchable.
|
Internal: 1 embedded atmospheric pressure sensor
|
Barometric Accuracy
|
±55 Pa
|
Measurement Rate
|
mV, V, mA & frequency: 3 times/sec
|
Pressure module: 1~10 times/sec selectable (3 as default)
|
Barometer: 1 time/sec
|
Power
|
4000mAh, 14.4Wh explosion-proof intelligent lithium battery, charging time = 6~8 hours, battery can be charged independently Typical working time 100 hours (measurement mode)
|
Environmental
|
Guaranteed temperature range of technical specifications: (-10 ~ 50)°C
|
Operating temperature: (-20 ~ 50)°C
|
Storage temperature: (-30 ~ 70)°C
|
Humidity: 0% to 95% RH, non-condensing.
|
Altitude: 3000 meters
|
Warm-up Time
|
10 min to fully meet technical specifications
|
Port Protection Voltage
|
30V max
|
Explosion-proof Grade
|
ATEX & IECEX: Ex ia IIC T4 Ga (Ta = -10°C ~ 50°C)
|
CSA: Class I, Division 1, Group A,B,C and D, T4 Ex ia IIC T4 Ga Class I, Zone 0, AEx ia IIC T4 Ga
|
UKCA-EX
|
CE Certification
|
TUV IEC61326, IEC61010
|
Rohs Compliance
|
Rohs II Directive 2011/65/EU, EN50581:2012
|
Protection Level
|
IP67, 1 meter drop test
|
Communication
|
Isolate USB-Type C (slave), Bluetooth
|
Display
|
4.4-inch color display capacitive screen, transflective, with LED backlight
|
Size
|
6.97" x 4.13" x 2.04" (177 mm x 105 mm x 52 mm) which doesn't include bottom ADT158Ex.
|
Weight
|
1.98 lb (0.9 kg)
|
Electrical Specifications
|
Specification
|
Range
|
Accuracy
|
Resolution
|
Note
|
Voltage Measurement
|
±300mV
|
0.015%RDG + 0.005%FS
|
1uV
|
Impedance: >100MΩ
|
±30V
|
0.015%RDG + 0.005%FS
|
0.1mV
|
Impedance: >1MΩ
|
Current Measurement
|
±30mA
|
0.015%RDG + 0.005%FS
|
0.1uA
|
Impedance: < 40Ω
|
Frequency Measurement
|
0.01~50000Hz (auto range)
|
0.005% RDG + 2 last digit
|
6-digit auto-resolution
|
Min threshold voltage: 2.5V
|
Units: Hz, kHz, MHz, CPM, CPH, s, ms, us
|
Switch On-Off Measurement
|
Inspection voltage: (3 ~ 30)V
|
Response speed: < 10ms, supports wet and dry switch
|
Pulse Count
|
0 ~ 9999999, optional rising edge and falling edge
|
Min threshold voltage: 2.5V
|
Loop Power
|
20V ± 10%, max output impedance: 320Ω, max load current: 25mA
|
Pressure Technichal Specifications
|
Pressure resolution
|
4, 5, or 6 digit resolution (user selectable)
|
Temperature Compensation
|
-10°C~50 °C
|
Pressure Module Type
|
Built-in digital pressure module: ADT158Ex, for more detailed information, please see ADT 158Ex datasheet. External digital pressure module: ADT161Ex,for more detailed information, please see ADT 161Ex datasheet.
|
Specifications
|
Refer to the technical specification of the ADT158Ex and ADT161Ex
|
High Static Pressure and Differential Pressure Synthesis Index
|
Two modules must be with the same range; Typical Differential pressure accuracy of 0.002%FS or 0.02%RD, whichever is greater when using two 0.02%FS external modules. Typical Differential pressure accuracy of 0.002%FS or 0.05%RD, whichever is greater when using two 0.05%FS external modules.
|
Gauge Pressure [1]
|
P/N
|
Pressure Range
|
Media [2]
|
Accuracy (%FS)
|
Burst Pressure
|
(psi)
|
(bar)
|
V15
|
-15
|
-1
|
G
|
0.02
|
3×
|
GP2
|
2
|
0.16
|
G
|
0.05
|
3×
|
GP5
|
5
|
0.35
|
G
|
0.05
|
3×
|
GP10
|
10
|
0.7
|
G
|
0.02
|
3×
|
GP15
|
15
|
1
|
G
|
0.02
|
3×
|
GP30
|
30
|
2
|
G
|
0.02
|
3×
|
GP50
|
50
|
3.5
|
G , L
|
0.02
|
3×
|
GP100
|
100
|
7
|
G , L
|
0.02
|
3×
|
GP150
|
150
|
10
|
G , L
|
0.02
|
3×
|
GP300
|
300
|
20
|
G , L
|
0.02
|
3×
|
GP500
|
500
|
35
|
G , L
|
0.02
|
3×
|
GP600
|
600
|
40
|
G , L
|
0.02
|
3×
|
GP1K
|
1000
|
70
|
G , L
|
0.02
|
3×
|
GP1.5K
|
1500
|
100
|
G , L
|
0.02
|
3×
|
GP2K
|
2000
|
140
|
G , L
|
0.02
|
3×
|
GP3K
|
3000
|
200
|
G , L
|
0.02
|
3×
|
GP5K
|
5000
|
350
|
G , L
|
0.02
|
3×
|
GP10K
|
10000
|
700
|
G , L
|
0.02
|
2×
|
GP15K
|
15000
|
1000
|
G , L
|
0.05
|
2x
|
GP20K
|
20000
|
1400
|
G , L
|
0.05
|
1.5x
|
GP25K
|
25000
|
1600
|
G , L
|
0.05
|
1.5x
|
GP30K
|
30000
|
2000
|
G , L
|
0.05
|
1.5x
|
GP36K
|
36000
|
2500
|
G , L
|
0.05
|
1.5x
|
GP40K
|
40000
|
2800
|
G , L
|
0.05
|
1.35x
|
GP50K
|
50000
|
3500
|
G , L
|
0.1
|
1.2x
|
[1] Sealed gauge pressure for above 1000 psi
|