The Mavolux 5032 B & C are high precision luxmeter's that are optimally suited for measuring very high light intensities (brightest daylight, head lights, etc.) without any additional accessories. Each has a USB interface for powering the meter, meter control, and data display/storage using GLUX 2 software (included). The Mavolux 5032 C is typically used in general applications and the Mavolux 5032 B has improved low light sensitivity for certification and inspection applications. The MAVOLUX 5032 B has been approved for certification and official inspection procedures.
Specifications
Automatic / Manual Measuring Range Selection Either Lux or footcandles can be selected as a measurement unit. The measuring range is matched automatically to the measurement value. The current measuring range can be locked by pressing a key, or any one of four or five (MAVOLUX 5032B USB) measuring ranges can be selected manually.
Function „Hold“ The current measurement value can be held at the display by pressing the data hold key.
Function „Mem“ Storage in memory of upto 100 values, read-out in the display or via USB port
Measuring rate 2 measurements per second
Light sensor Silicon photo diode with V (λ) filter, complies with accuracy class MAVOLUX 5032 C USB: DIN 5032-7 Class C MAVOLUX 5032 B USB: DIN 5032-7 Class B
Display LCD display size 50 mm x 25 mm Display/Digit Height 7 segments characters/13 mm 3 ½ digits Overload Display „OL“ appears in display
Power Supply Battery 1,5 V alkaline-manganese cell (IEC LR 6), size AA Battery Life approx. 45 hours continuous operation, When the meter is connected to a PC, power will be supplied by the PC via the USB cable Battery test symbol is automatically displayed when battery voltage drops below appr.1V
Battery Saver Circuit The instrument is automatically switched off, if the measurement value remains unchanged for approximately 4 minutes, and if none of the operating elements has been activated during this time.
Continuous operation For using the instrument in the continuous operation mode, keep the key HOLD pushed down and then press the key ON/OFF. Pushing the key ON/OFF again will end the operation
Electromagnetic Compatibility EMC The MAVOLUX meet the Specifications 89/336/EWG dt. 01.01.1996
Mechanical Design Dimensions Basic instrument 65 x 120 x 19 mm Sensor 31 x 105 x 30 mm Cable between Instrument and Sensor MAVOLUX 5032 C-USB: coiled cord, permanently attached MAVOLUX 5032 B-USB: plug-in coiled cord. Cable Length 1,5 m; available also in 3 m, 5 m or 10 m – on special order Weight approx. 200 g without battery