The Mavolux 5032 B & C are high precision luxmeter's that are optimally suited for measuring very high light intensities (brightest daylight, head lights, etc.) without any additional accessories. Each has a USB interface for powering the meter, meter control, and data display/storage using GLUX 2 software (included). The Mavolux 5032 C is typically used in general applications and the Mavolux 5032 B has improved low light sensitivity for certification and inspection applications. The MAVOLUX 5032 B has been approved for certification and official inspection procedures.
Automatic / Manual Measuring Range Selection
Either Lux or footcandles can be selected as a measurement
unit. The measuring range is matched automatically to the
measurement value. The current measuring range can be locked
by pressing a key, or any one of four or five (MAVOLUX 5032B
USB) measuring ranges can be selected manually.
Function „Hold“
The current measurement value can be held at the display by
pressing the data hold key.
Function „Mem“
Storage in memory of upto 100 values, read-out in the display or
via USB port
Measuring rate
2 measurements per second
Light sensor
Silicon photo diode with V (λ) filter, complies with accuracy
class
MAVOLUX 5032 C USB: DIN 5032-7 Class C
MAVOLUX 5032 B USB: DIN 5032-7 Class B
Display
LCD display size 50 mm x 25 mm
Display/Digit Height 7 segments characters/13 mm
3 ½ digits
Overload Display „OL“ appears in display
Power Supply
Battery 1,5 V alkaline-manganese cell
(IEC LR 6), size AA
Battery Life approx. 45 hours continuous operation,
When the meter is connected to a PC,
power will be supplied by the PC via the
USB cable
Battery test symbol is automatically displayed
when battery voltage drops below
appr.1V
Battery Saver Circuit
The instrument is automatically switched off, if the
measurement value remains unchanged for approximately
4 minutes, and if none of the operating elements has been
activated during this time.
Continuous operation
For using the instrument in the continuous operation mode,
keep the key HOLD pushed down and then press the key
ON/OFF. Pushing the key ON/OFF again will end the
operation
Electromagnetic Compatibility EMC
The MAVOLUX meet the Specifications 89/336/EWG dt.
01.01.1996
Mechanical Design
Dimensions Basic instrument 65 x 120 x 19 mm
Sensor 31 x 105 x 30 mm
Cable between Instrument
and Sensor MAVOLUX 5032 C-USB: coiled cord,
permanently attached
MAVOLUX 5032 B-USB: plug-in coiled
cord.
Cable Length 1,5 m; available also in 3 m, 5 m or
10 m – on special order
Weight approx. 200 g without battery
- Light technicians
- Control of light sources
- Street lights
- Lighting of work places, public buildings, sports facilities
- Quality control and quality assurance in the manufacture of lamps and light sources
- Light boards / LED billboards
- Light designers and architects
- Measurements in agriculture and horticulture