Vaisala Indigo520 Universal Premium Transmitter

SKU
INDIGO520
Manufacturer:
Vaisala
Vaisala

Key Features:

  • Absolute pressure measurement range of 500 … 1100 hPa
  • Total accuracy ±0.15 hPa
  • Pressure units: hPa, mbar, kPa, Pa, inHg, mmH20,mmHg, torr, psia
  • Supports 2 detachable probes simultaneously
  • Large touchscreen display
  • HUMICAP & BAROCAP sensor technologies
  • IP66 and NEMA 4 rated metalenclosure
  • 4 configurable galvanicallyisolated analog output
  • 2 relays
  • Ethernet connection with web interface for remote access
  • Modbus® TCP/IP protocol
Product Options:
$1,085.00

Secured and trusted checkout with

The Vaisala Indigo520 supports two detachable probes simultaneously. The Indigo520 is compatible with Indigo probes that measures humidity, temperature, dew point, carbon dioxide, hydrogen peroxide, and moisture in oil measurements. The Indigo520 can also measure barometric pressure with an additional module.

 

Choose the humidity measurement probe that best suits your needs:

  • • HMP1 for laboratories
  • • HMP3 with a field replaceable sensor for general use
  • • HMP7 for demanding, high-humidity industrial and meteorology applications
  • • HMP9 for confined spaces
  • • TMP1 for temperature measurement only

 

Indigo520 Examples

Indigo Compatible Probes

Measurement typeProbe models
Humidity and temperatureHMP1, HMP3, HMP4, HMP5, HMP7, HMP8, HMP9
TemperatureTMP1
Dew pointDMP5, DMP6, DMP7, DMP8
CO2GMP251, GMP252
Vaporized hydrogen peroxideHPP271, HPP272
Moisture in oilMMP8

 

Other Compatible Devices

Device or seriesModels
MHT410 Moisture, Hydrogen and
Temperature Transmitter
MHT410

 

Measurement Performance

Barometric pressure (optional module)
Pressure range500 … 1100 hPa
Class A:
Linearity±0.05 hPa
Hysteresis±0.03 hPa
Repeatability±0.03 hPa
Calibration uncertainty±0.07 hPa
Accuracy at +20 °C / +68 °F±0.10 hPa
Temperature dependence±0.1 hPa
Total accuracy (-40 … +60 °C / -40 … +140 °F)±0.15 hPa
Long-term stability/year±0.1 hPa
Response time (100 % response):
One sensor2 s
Pressure unitshPa, mbar, kPa, Pa, inHg, mmH20, mmHg, torr, psia

 

Mechanical Specifications

NEMA ratingNEMA 4
Housing classificationIK08, DIN EN ISO 11997-1: Cycle B
(VDA 621-415)
Housing materialAlSi10Mg (DIN 1725)
Display window materialStrengthened glass (IK08)
Display size5 in
Weight1.5 kg (3.3 lb)
Dimensions (H × W × D)142 × 182 × 67 mm
(5.63 × 7.17 × 2.64 in)
Cable diameters for cable glands
M20×1.5 glands5.0 … 8.0 mm (0.20 … 0.31 in)
M20×1.5 glands with split bushing7 mm (0.28 in)
M16×1.5 glands2.0 … 6.0 mm (0.08 … 0.24 in)

 

Inputs and Outputs

Operating power 1)
Protective extra-low voltage (PELV)
version
15 … 35 V DC, 24 V AC ±20 %
50/60 Hz, max. current 2 A (power
supply input is galvanically isolated)
Fuse size for power supply: 3 A
Isolation voltage: 500 V AC,
1000 V DC
PELV power cable temp. rating≥ +80 °C (+176 °F)
AC (mains) power version100 … 240 V AC 50/60 Hz,
max. current 1 A (power supply input
is galvanically isolated)
Fuse size for power supply: 10 A
Isolation voltage: 1500 V AC
Power over Ethernet versionPower over Ethernet (PoE)
IEEE 802.3at Type 2 Class 4
Max. current 600 mA, max. power
consumption 25.5 W
Isolation voltage: 500 V AC,
1000 V DC
Analog outputs
Number of analog outputs4
Selectable voltage output types0 … 1 V, 0 … 5 V, 0 … 10 V, scalable
Selectable current output types4 … 20 mA, 0 … 20 mA, scalable
Max. wire size2.5 mm2 (14 AWG)
Accuracy of analog outputs at
+20 °C (+68 °F)
±0.05 % full scale
Temperature dependence±0.005 % / °C full scale
External loads:
Current outputsRL < 500 Ω
0 … 1 V outputRL > 2 kΩ
0 … 5 V and 0 … 10 V outputsRL > 10 kΩ
Relay outputs
Number and type of relays2 pcs, SPDT
Max. switching power, current, voltage30 W, 1 A, 40 V DC / 28 V AC
Max. wire size in PELV version2.5 mm2 (14 AWG)
Max. wire size in AC (mains) version1.5 mm2 (16 AWG)
Ethernet interface
Supported standards10BASE-T, 100BASE-TX
Connector8P8C (RJ45)
Supported protocolsModbus TCP/IP (port 502), HTTPS
(port 8443)

1) The power supply option is selected when ordering the transmitter.

 

Operating Environment

For use in wet locationsYes
Operating humidity0 … 100 %RH
Maximum operating altitude3000 m (9843 ft)
IP ratingIP66 1)
Operating temperature
With display-20 … +55 °C (-4 … +131 °F)
Without display-40 … +60 °C (-40 … +140 °F)
Without display with barometer module-40 … +55 °C (-40 … +131 °F)
Storage temperature
With display-30 … +60 °C (-22 … +158 °F)
Without display-40 … +60 °C (-40 … +140 °F)

1) Evaluated by Eurofins, not by UL

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account