Confirm the functionality of refrigerant dryers with Vaisala’s DMT132 Dewpoint Transmitter.
The DMT132 maintains a Td accuracy of +/-1°C (+/-1.8°F) in the typical operating measurement range of -3°C to 20°C (26.6°F to 68°F) while the Vasiala HUMICAP sensor, resistant to compressor oil and many other chemicals, ensures superior long-term stability.
Performance | |
DEW POINT | |
Measurement range | -20 ... +50 °C (-4 ... +122 °F) Td |
Accuracy (at +20 °C (+68 °F) | ±1 °C for -3...20 °C (+26.6 ... +68 °F) Td ±2 °C for -15...-3 °C (+5 ... +26.6 °F) Td |
* when dew point is below 0 °C (+32 °F), the transmitter outputs frostpoint | |
Response time at 20 °C (+68 °F) gas temperature and 1 bar pressure | |
-14 -> +3 °C (+7 -> +37 °F) Td | 17 s (63%) 40 s (90%) |
+3 -> -14 °C (+37 -> +7 °F) Td | 33 s (63%) 85 s (90%) |
Operating Environment | |
Operating temperature Operating pressure Relative humidity Sample flow rate Measured gases | -10 ... + 60 °C (+14 ... +140 °F) 0 ... 20 bar 0 ... 100 %RH no effect on measurement accuracy gases non-corrosive gases |
Outputs | |
Analog output (scalable) | 4...20 mA, 2-wire |
Resolution for current output | 0.002 mA |
Accuracy of analog outputs at +20 °C | ±0.05% full scale |
Typical temperature dependence | ±0.005% of full scale/ °C |
Connector | 4-pin M8 (IEC 60947-5-2) |
LED indication available for defined dew point limit/ error state indication RS485 serial line for service use | |
General | |
Sensor | Vaisala HUMICAP®180R |
Recommended calibration interval (in refrigerant dryer application) | 2 years |
Mechanical connection | G 1/2" ISO |
Operating voltage | 10 .... 28 VDC |
External load | max 100 ohm for supply voltages <20 VDC max 500 ohm for supply voltages 20...28 VDC |
Weight | 65 g (2.3 oz) |
Housing material | PPS + 40% GF |
Housing classification | IP65 (NEMA 4) |
Storage temperature range | -40 ... + 80 °C (-40 ... +176 °F) |
Start-up time | 3 s |
Complies with EMC standard EN61326-1, Electrical equipment for measurement control and laboratory use - EMC requirements; Industrial environment |