Wika Mensor’s CPC4000 Industrial Pressure Controller delivers an expansive pressure range from -1 … 210 bar (-15 … 3,045 psi). It’s available as a desktop or as a 19" rack-mounting kit. The CPC4000 is equipped for up to two reference pressure sensors and an optional barometer for showing barometric pressure or can be used to imitate gauge or absolute pressure.
With an accuracy of up to 0.02 % IS-50, and able to control pressure with a high stability, it is particularly apt as a production tool for transmitter manufacturing, a calibration and maintenance tool for pressure measuring instruments or as a factory/working standard for the calibration of all types of pressure measuring instruments.
The CPC4000’s touchscreen and effortless menu navigation provide for a user friendly operation. More still, for the utmost user flexibility, depending on the application, the operator can choose between three set-point methods:
- Direct entry of the pressure value (set point) which will be controlled via touchscreen keypad.
- Define steps to reach the desired pressure value by either defining fixed pressure increments or a percentage of span value.
- User-defined programmable test sequences.
Reference pressure transducers | ||
Pressure range | Standard | Optional |
Accuracy 1) | 0.02 % FS 2) | 0.02 % IS-50 3) |
Gauge pressure | 0 ... 0.35 to 0 ... 210 bar (0 ... 5 to 0 ... 3,045 psi) | 0 ... 1 to 0 ... 210 bar (0 ... 15 to 0 ... 3,045 psi) |
Bi-directional | -0.17 ... 0.17 to -1 ... 210 bar (-2.5 ... 2.5 to -15 ... 3,045 psi) | -1 ... 10 to -1 ... 210 bar (-15 ... 145 to -15 ... 3,045 psi) |
Absolute pressure 4) | 0 ... 1 to 0 ... 211 bar abs. (0 ... 15 to 0 ... 3,060 psi abs.) | 0 ... 1 to 0 ... 211 bar abs. (0 ... 15 to 0 ... 3,060 psi abs.) |
Precision 5) | 0.008 % FS | 0.008 % FS |
Calibration interval | 365 days | 365 days |
Optional barometric reference | ||
Function | The barometric reference can be used to switch pressure types 6), absolute <=> gauge. With gauge pressure transducers, the measuring range of the transducers must begin with -1 bar (-15 psi) in order to carry out a complete absolute pressure emulation. | |
Measuring range | 552 ... 1,172 mbar abs. (8 ... 17 psi abs.) | |
Accuracy 1) | 0.02 % of reading | |
Pressure units | 39 and two freely programmable | |
1) It is defined by the total measurement uncertainty, with the coverage factor (k = 2) and includes the intrinsic performance of the instrument, the measurement uncertainty of the reference instrument, long-term stability, influence of ambient conditions, drift and temperature effects over the compensated range with recommended zero point adjustment every 30 days. 2) FS = full span 3) 0.02 % IS-50 accuracy: Between 0 ... 50 % of the full scale, the accuracy is 0.02% of half of the full scale value and between 50 ... 100 % of the full scale, the accuracy is 0.02 % of reading. 4) The minimum calibrated range of absolute transducer(s) is 600mTorr 5) It is defined as the combined effects of linearity, repeatability and hysteresis throughout the stated compensated temperature range 6) For a pressure type emulation, we recommend a native absolute pressure transducer, since the zero point drift can be eliminated through a zero point adjustment. |
Base instrument | |
Instrument | |
Instrument version | Standard: desktop case Option: 19" rack-mounting kit |
Dimensions | See Data Sheet |
Weight | approx. 12.7 kg (28 lbs) incl. all internal options |
Warm-up time | approx. 15 min |
Display | |
Screen | 7.0" color LCD with resistive touchscreen |
Resolution | 4 ... 6 digits depending on range and units |
Connections | |
Pressure connections | 4 ports with 7/16"- 20 F SAE, 1 port with 1/8" F NPT and 1 port with 10-32 UNF female |
Filter elements | The instrument has a 40-micron filters on all pressure ports. |
Pressure port adapters | Standard: without Option: 6 mm tube fitting, 1/4” tube fitting, 1/4” female NPT fittings, 1/8” female NPT fittings or 1/8” female BSP fittings |
Barometer port adapters | Standard: barb fitting Option: 6mm tube fitting, 1/4“ tube fitting |
Permissible pressure media | Dry, clean air or nitrogen (ISO 8573-1:2010 class 5.5.4 or better) |
Wetted parts | Aluminum, brass, 316 and 316L stainless steel, Buna N, FKM/FPM, PCTFE, PEEK, PTFE, PPS, glass-filled epoxy, RTV, ceramic, silicone, silicone grease, Urethane |
Overpressure protection | Safety relief valve fixed to reference pressure transducer and adjusted to customized measuring range |
Permissible pressure | |
Supply port | 110 % FS or 0.69 bar (10 psi), whichever is greater |
Measure/Control port | max. 105 % FS |
Voltage supply | |
Power supply | AC 100 ... 240 V, 50/60 Hz |
Power consumption | max. 150 VA |
Permissible ambient conditions | |
Storage temperature | -20 ... 70 °C (-4 ... 158 °F) |
Humidity | 5 … 95 % r. h. (relative humidity non-condensing) |
Compensated temperature range | 15 ... 45 °C (59 ... 113 °F) |
Mounting position | horizontal |
Control parameter | |
Control stability | < 0.005 % FS of the active range |
Control mode | precision, high speed and custom |
Control time | 10 s (regarding a 10 % FS pressure increase in a 50 ml test volume) |
Control range | 0 ... 100 % FS |
Minimum control pressure | 0.0017 bar (0.025 psi) over exhaust pressure or 0.05% FS whichever is greater |
Overshoots | < 0.3 % FS in high speed control mode (typical <0.1% FS in precision control mode) |
Test volume | 50 ... 1,000 ccm |
Communication | |
Interface | Standard: Ethernet, IEEE-488, USB, RS-232. |
Command sets | Mensor, WIKA SCPI, others optional |
Response time | approx. 100 ms |
Internal program | up to 24 sequences with up to 99 steps each |
Approvals and certificates | |
EC declaration of conformity | |
EMC directive 7) | EN 61326-1 emission (group 1, class A) and interference immunity (industrial application) |
Low voltage directive | EN 61010-1 |
Certificate | |
Calibration 8) | Standard: A2LA calibration certificate (standard on factory) Option: DKD/DAkkS calibration certificate |
7) Warning! This is class A equipment for emissions and is intended for use in industrial environments. In other environments, e.g. residential or commercial installations, it can intefere with other equipment under certain conditions. In such circumstances the operator is expected to take the appropriate measures. 8) Calibration in a horizontal position/operating position. |
- Oil and gas industry
- Industry (laboratory, workshop and production)
- Transmitter and pressure gauge manufacturers
- Calibration service companies and service industry