Accuracy
Accuracy: +/-0.025% FS or optional +/-0.05% F.S. No temperature effect from 23 to 122 F.
Engineering Units
Inches of H2O (select 4 C, 20 C, or 60 F), PSI, mm of Hg, inches of Hg, mBars, Bars, kPa, kg/cm2, cm of H2O and user selectable scaling.
Media Compatibility
DN: Non-isolated differential sensor for clean, dry, non-corrosive gases. GI & AI: Isolated Gauge or Absolute sensor for fluids compatible with 316L SS.
Pressure Limits
Twice range on Gauge and Absolute (GI and AI) units. Twice range on Differential (DN) units when pressurized on high side only and 150 PSI (10.5 Kg/cm2) static when applied to both sides of the sensor simultaneously.
Leak Test
Allows users to view Min/Max pressure values and calculates leak rate per minute in pressure units. The feature is very handy when troubleshooting or pressure testing pneumatic systems.
Damping Rates
Exponential damping with user selectable time constant from 0.1 to 25 seconds.
Min/Max Capture
Capture speed is equal to the selected damping rate.
Auto Record
Document up to 240 readings (automatically or manually) to reduce paperwork and field time. Saved readings can be recalled at any time.
Field Re-cal
The M2 series can be recalibrated in the field for zero, span, and linearity. The user has the option to select a 1 point or 5 point calibration procedure. This feature enables users with accurate primary standards to calibrate their devices locally if desired.
Auto Shut-Off
User selectable shut-off time period. Programmable at 10, 20, 30, 45, 60 minutes.
Base Unit
14 ounce (6.5" X 3.6" X 2.25") polycarbonate, permanently dissipative, ESD protection.
Keypad
Sealed membrane type with 5 dome contact style buttons.
Display
5 significant digit LCD (0.25" high) 2 line x 16 alphanumeric characters with contrast adjustment. Two stage backlight with visual overrange indication.
Power
4 each AA batteries providing over 100 hours of continuous use.
Wetted Parts
DN: 316L SS, brass, viton
GI: 316L SS
AI: 316L SS
Connections
1/8" female NPT, 316L SS
NIST Traceable Accuracy
Pressure: +/-0.025% of F.S. or +/-0.05% F.S. (all ranges except DN0028 +/-0.1% of F.S. See "Pressure Ranges" table for full scale value.)
Temperature Specifications
Storage: -40 F to 140 F (-40 C to 60 C)
Operating: 23 F to 122 F (-5 C to 50 C)
Certifications
IS and ATEX (pending)
EEx ia IIC T4 (pending)
(Tamb. -5 C to +50 C) (pending)
Note 1: Connector Kit (not for use with GI2000) includes: 1/8" tubing, "T" unions (2),1/8" MNPT adapters (6), 3/16" Barb to 1/8" FNPT (1) & 1/4" MNPT to 1/8" FNPT reducer (2)