We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The Meriam M200LS Lab Standard Smart Manometer offers the best available pressure measurement accuracy in a portable, handheld format. The NIST traceable accuracy is typically ±0.01% of F.S. and is independent of temperature effect over an operating temperature range of 59º to 86º F. Pressure ranges from 28” H2O F.S. to 500 PSIG F.S. are available.
M2 pressure sensors are available to measure gauge, differential, absolute or vacuum pressure. The display can read out in any of eleven pressure units or can be linearly scaled for special user units. Differential models offer a square root flow function to display flow rate in user defined units.
An adjustable damping feature minimizes the effects of pulsating pressures. A Min/Max function captures the extremes of pulsating or varying pressure signals and a Hold function can freeze the display at any point of interest. Auto Record documents up to 240 readings for future recall.
Base Unit | 14 ounce (6.5" x 3.6" x 2.25") polycarbonate, permanently dissipative, ESD protection | ||||||
Keypad | Sealed membrane type with 5 dome contact style buttons. | ||||||
Display | 5 significant digit LCD (0.25" high) 2 line x 16 alphanumeric characters with contract adjustment. Two stage backlight with visual over range indication. | ||||||
Power | 4 each AA batteries providing over 100 hours of continuous use | ||||||
Wetted Parts |
| ||||||
Connections | 1/8" female NPT, 316L SS | ||||||
Traceable Accuracy | Pressure: ±0.01% of F.S. DN0028 ±0.02% of F.S. Accuracy statement includes all affects of temperature over specified operating range | ||||||
Temperature Spec |
|