MadgeTech’s RHTemp1000IS compact humidity and temperature logger is FM approved as intrinsically safe for hazardous environments.
This logger has the ability to measure and record temperatures from -40 °C to +80 °C (-40 °F to +176 °F) with a superior accuracy of ±0.5 °C. The RHTemp1000IS has the memory capacity for 16,384 time and date stamped readings, and is able to be programmed to start up to 24 months in advance.
The RHTemp1000IS is compatible with the latest MadgeTech 4 Software making starting, stopping, and downloading fast and simple. Graphical, tabular, and summary data is provided for analysis and can be displayed as customizable engineering units.
The RHTemp1000IS has been Factory Mutual certified as intrinsically safe for Class I, Division 1, groups A, B, C and D and non-incendive for Class I, Division 2, groups A, B, C and D.
General
Memory | 16,384 readings (all channels enabled) |
Reading Rate | 1 reading every second up to 1 reading every 24 hours |
Battery Life | 2 years typical at 25 °C, 15 minute reading intervals |
Material | 316 Stainless Steel/Radel |
Dimensions | 1.7 in x 0.97 in x 0.97 in (42 mm x 24.6 mm x 24.6 mm) |
Operating Environment | -40 °C to +80 °C (-40 °F to +176 °F), 0 %RH to 100 %RH non-condensing |
Required Interface Package | IFC400 or IFC406 |
Approvals | CE, IS Rated |
Temperature
Temperature Sensor | Resistance Temperature Detector (RTD) |
Temperature Range | -40 °C to +80 °C (-40 °F to +176 °F) |
Temperature Resolution | 0.01 °C (0.018 °F) |
Temperature Response Time in Air | t63 - 8 minutes to 63% of change t90 - 68 minutes to 90% of change |
Temperature Calibrated Accuracy | ±0.5 °C (0 °C to 55 °C) |
Humidity
Humidity Sensor | Capacitive Polymer |
Humidity Range | 0 %RH to 100 %RH (non-condensing) |
Humidity Resolution | 0.1 %RH |
Humidity Response Time in Air | t90 - 2 minutes to 90% of change |
Humidity Calibrated Accuracy | ±3 %RH maximum |
Humidity Specified Accuracy Range | 25 %RH to 75 %RH +20 °C to +40 °C Hysteresis Error 1 % Typical, 3% Maximum |
- EtO Sterilization
- Environmental Studies
- Hostile Environment Monitoring
- Medical & Pharmaceutical