We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
GW Instek has introduced their GOM-804 as the latest installment to their new series of DC milliohm meters. The GOM-804 offers a vast 3.5” TFT color display with 50,000 counts maximum measurement display for superior measurement clarity. It is suited with a 60 readings per second fast sampling rate, 0.05% precision and a four-wire measurement method in addition to the temperature measurement and temperature compensation measurement function.
The meter delivers a D-sub 25-pin combined interface to carry out, contingent on the functionalities, Handler, Scan or EXT IO for respectively connecting to a sorting machine; connecting to an external on-off switch, and directly conducting external trigger control. Its functionality keys and direction keys allow for ease of operation, enabling users to complete their tasks with efficiency and convenience.
General
NOTE | Specifications are ensured only for GOM-804/805 at least 30 minute of warm-up time and Operating Temperature Specification: 18~28˚C |
Display | 3.5'(320 x 240) TFT LCD |
Memory | 20 sets for panel setting |
Power Source | AC 100 - 240 V, 50/60Hz |
Comsumption | 25VA (max.) |
Dimension | 223(W) x 102(H) x 283(D) mm |
Weight | Approx.3kg(max.) |
Specifications
MODEL | GOM-804 | GOM-805 | |||||||||
DISPLAY | |||||||||||
| 50,000 counts | ||||||||||
SAMPLING RATE | |||||||||||
Slow | 10 readings / s | ||||||||||
Fast | 60 readings / s | ||||||||||
RESISTANCE MEASUREMENT | |||||||||||
Range / Test current (fixed) |
| ||||||||||
Accuracy | 50mΩ : ± (0.1% of reading + 0.02% of range) | ||||||||||
| 500mΩ ~ 50Ω : ± (0.05% of reading + 0.02% of range) | ||||||||||
| 500Ω ~ 500kΩ : ± (0.05% of reading + 0.008% of range) | ||||||||||
| 5MΩ : ± (0.2% of reading + 0.008% of range) | ||||||||||
Resolution | 1µΩ, 10µΩ, 100µΩ, 1mΩ, 10mΩ, 100mΩ, 1Ω, 10Ω, 100Ω | ||||||||||
TEMPERATURE | |||||||||||
Range | -50˚C ~399.9˚C | ||||||||||
Accuracy | -10˚C ~40˚C : 0.3% ± 0.5˚C ; Other : 0.3% ± 1.0˚C | ||||||||||
Resolution | 0.1˚C | ||||||||||
DRY CIRCUIT | |||||||||||
| ----- | Open circuit less than 20mV; For 500mΩ, 5Ω, 50Ω range only | |||||||||
DRIVE MODE | |||||||||||
DC+/DC- | ----- | Yes | |||||||||
Pulsed | ----- | Yes | |||||||||
PWM | ----- | Yes | |||||||||
Zero | ----- | Yes | |||||||||
OTHER FUNCTIONS | |||||||||||
| Trigger – Internal, Manual, External、Math – ABS, REL, %, TC、Hi / Lo | ||||||||||
| Average 2 ~ 10 times、Measured delay、Go / No-Go、TC for Transformer | ||||||||||
| Diode、Continuity beeper、Sorting (only for GOM-805) | ||||||||||
INTERFACE | |||||||||||
USB | Standard | Standard | |||||||||
RS-232C | Standard | Standard | |||||||||
HANDLER/SCAN/EXT I/O | Standard | Standard | |||||||||
GPIB | Option | Standard |