GW Instek GCT-9040 40A AC Ground Bond Tester

SKU
GCT-9040
Manufacturer:
GW Instek
GW Instek

Key Features:

  • AC 40A Ground Bond Tester
  • Measuring Resistance from 1mΩ~650mΩ
  • Connect with the GPT-9800/9900 series to become a multifunctional test or simultaneous test system
  • 240x64 Ice Blue Dot Matrix LCD
  • Function Key for quick selecting
  • High intensity flash for Caution & Status indication
  • PWM switching amplifier to enhance the power efficiency and reliable testing
  • Max. 100 memory block for test condition setting.
  • Remote terminal on the front panel for “start” and “stop” remote active
  • Interface : USB device, Signal I/O and GPIB (optional)
Product Options:
$1,804.00

Secured and trusted checkout with

The GW Instek GCT-9040 delivers the greatest AC test current of 40A ensuring test proficiency and dependability. Its PWM design, vast LCD display, 100 memory blocks from setting criteria and programmable communications interface provides users with a greater readability and convenience.

General

NOTE

The specifications apply when the GCT-9040 is warmed up for at least 30 minutes and works in operation temperature : 15˚C ~35˚C.

Display

240 x 64 Ice Blue Dot Matrix LCD

Memory

100 sets for panel setting save and recall

Power Source

AC 100 V / 120 V / 220 V / 230 V ±10%

Line Frequency

50/60Hz

Comsumption

700VA (max)

Dimensions

330(W) x 148(H) x 460(D) mm

Weight

Approx. 17kg (max)

 

AC Ground Bond

GROUND BOND

 

Output-Current

03.00A~ 40.00A ac

Output-Current Resolution

0.01A

Output-Current Accuracy

3A≦I≦8A : ± (1% of setting + 0.2A)
8A< I≦40A : ± (1% of setting + 0.05A)

Test-Voltage

8Vac max (open circuit)

Test-Voltage Frequency

50Hz/60Hz selectable

Resistance Measurement Range

1.0mΩ ~ 650.0mΩ

Resistance Measurement Resolution

0.1mΩ

Resistance Measurement Accuracy

± (1% of reading + 2mΩ)

Window Comparator Method

Yes

TIMER (Test Time)

0.5s~999.9s

GND

OFF (fix)

Test Method

Four Terminal

MEMORY

 

Single Step Memory

MANU : 100 blocks

INTERFACE

 

LINK

For system connection

USB

Standard

GPIB

Option

Remote Terminal (Front)

Standard

Signal I/O

Standard

Display

240 x 64 Ice Blue Dot matrix LCD

POWER SOURCE & COMSUMPTION

 

Source

AC 100 V / 120 V / 220 V / 230 V ±10%, 50/60Hz

Comsumption

Max. 700VA

DIMENSIONS & WEIGHT

 

 

330(W) x 148(H) x 460(D) mm; Approx. 17kg max.

Output Limitation in AC Ground Bond Testing as below

 

Output Range

Pause

Output Time (max)

Ground Bond

20A<I≦40A

At least as long as the output time

999.9s

3A≦I≦20A

No Necessary

999.9s

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account
  • Safety Testing of Ground Bond of Electrical Product in Manufacturing
  • Quality Assurance Verification
  • Safety Standard Compliance Pre-qualification in R&D
  • Performing a Sequential or Simultaneous Test After Connect with GPT-9800/9900 Series