Main Unit Specifications
Item | Description | |||
---|---|---|---|---|
Number of analog input channels | 10 channels | |||
External Input/Output | Trigger input, Logic input 4 ch or Pulse input 4 ch, Alarm output 4 ch | |||
Sampling interval | 100 ms to 1 h (Max:100 ms/10ch) | |||
TIME/DIV | 1 sec/DIV to 72 h/DIV | |||
Trigger function | Type | Start | Data capture starts when a trigger is generated | |
Stop | Data capture starts when a trigger is generated | |||
Condition | Start | Off, Level, Alarm, Scheduled time, External, Day | ||
Stop | Off, Level, Alarm, Scheduled time, Time, Day, External | |||
Condition | Analog | H, L, Window In, Window Out, Rate of change | ||
Logic | H, L | |||
Pulse | H, L, Window In, Window Out, Rate of change | |||
Alarm output | No. of channels | 4 | ||
Output format | Relay contact output (NO/NC) | |||
Output conditions | Level judgement, Window judgement, Logic pattern judgement, Pulse judgement | |||
Output terminal | M4 screw terminal | |||
Rating | 250 VAC/2A, 30 VDC/2A | |||
Pulse input range | Count mode | 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M C/F.S. (max. 50 k/sampling interval) | ||
Inst. Mode | 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M C/F.S. (max. 50 k/sampling interval) | |||
RPM mode | 50, 500, 5000, 50 k, 500 k, 5 M, 50 M, 500 M RPM/F.S. (max. 50 k/sec) | |||
Calculation function | Statistics calculation *1 | Average, Peak, maximum, Minimum, RMS (2 calculations can be set simultaneously) | ||
Calculation between channels | Addition, Subtraction, Multiplication, Division (for analog input ch1 to ch10) | |||
Wet-and dry-bulb temperature conversion | Humidity is calculated from the measured temperature of Wet-bulb and Dry-bulb (Dry bulb : fixed to ch1, Wet bulb : fixed to ch2, Range 0 to 100 %) | |||
Other function | Searching, Annotation input, Message, Marker function | |||
Interface to PC | Ethernet (10BASE-T / 100BASE-TX), USB (Compatible with full speed) | |||
Ethernet function | Web server function, FTP server function, FTP client function, NTP client function | |||
USB function | USB drive mode (File transfer/delete of internal memory) | |||
Internal storage device | Internal flash memory : 14 Mbyte, USB memory slot (compatible with full speed) *2 | |||
Display | Size | 5.7 inch TFT color LCD | ||
Displayed items | Waveforms + Digital values (vertical, horizontal), Digital + Calculation, Bar graph (vertical) | |||
Clock accuracy (23°C environment) | ±0.002 % (within approx. 50 sec per month) | |||
Operating environment | Temperature : 0 to 50 °C, Humidity : 5 to 85 % R.H. | |||
Withstand voltage | Between each input channel and GND: 1 minute at 350 Vp-p | |||
Between input terminals: 1 minute at 350 Vp-p | ||||
Power supply | 100 to 240 V AC / 50 to 60 Hz, M4 screw terminal (Power cord is not attached) | |||
Power consumption | 38 VA (when LCD is ON) | |||
External dimensions (W x H x D)(approx.) | 144 mm x 144 mm x 200 mm | |||
Weight (approx.) | 2.1 kg | |||
Vibration resistance | Compatible with JIS Vibration testing methods for automobile Type 1 Class A-equivalent | |||
Dust proof, Drip proof: IP65 compatible (for front panel only) |
Analog input Specifications
Item | Description | |||
---|---|---|---|---|
Type of input terminal | M4 screw type terminal | |||
Measurement range | Voltage | 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20, 50, V, 1-5 V F.S. | ||
Temperature | Thermocouple : K, J, E, T, R, S, B, N, W (WRe5-26) | |||
R.T.D. : Pt100, JPt100, Pt1000 (IEC751) | ||||
Humidity | 0 to 100 % RH (Voltage 0 to 1 V conversion, when using optional B-530 humidity sensor is used) | |||
Input filter | Off, 2, 5, 10, 20, 40 (Moving average) | |||
Measurement accuracy *3 | Voltage | ±0.1 % of F.S. | ||
Thermocouple | Measurement range | Accuracy | ||
R/S | 0 TS 100 °C 100 < TS 300 °C R : 300 < TS 1600 °C S : 300 < TS 1760 °C | ± 5.2 °C ± 3.0 °C ± (0.05 % of rdg + 2.0 °C) ± (0.05 % of rdg + 2.0 °C) | ||
B | 400 TS 600 °C 600 < TS 1820 °C | ± 3.5 °C ± (0.05 % of rdg + 2.0 °C) | ||
K | -200 TS -100 °C -100 < TS 1370 °C |
| ||
E | -200 TS -100 °C -100 < TS 800 °C |
| ||
T | -200 TS -100 °C -100 < TS 400 °C |
| ||
J | -200 TS -100 °C -100 < TS 100 °C 100 < TS 1100 °C |
| ||
N | 0 TS 1300 °C |
| ||
W | 0 TS 2000 °C |
| ||
R.T.D. | Pt100 | -200 to 850 °C (FS = 1050 °C) |
| |
JPt100 | -200 to 500 °C (FS = 700 °C) |
| ||
Pt1000 | -200 to 500 °C (FS = 700 °C) |
| ||
* If the reference junction compensation is internal, add ± 0.5 °C to each of the above values. | ||||
A/D converter | 16 bit (out of which 14 are internally acknowledged) | |||
Maximum input voltage | 60 Vp-p (Between ± terminals) | |||
60 Vp-p (Between input terminals) | ||||
60 Vp-p (Between input terminal and GND) | ||||
Withstand voltage | Between each input channel and GND: 1 minute at 350 Vp-p | |||
Between input terminals: 1 minute at 350 Vp-p |
Remarks
*1 for realtime and between cursors (when data reviewing)
*2 1 file 2 GB (it depends on used USB memory)
*3 Thermocouple: 0.32φ for T type, 0.65φ for other types
PC software specifications
Item | Description |
---|---|
Compatible OS | Windows 2000/XP/Vista (32 bit and 64 bit version) |
Function | Control of main unit, Realtime data capture, File conversion |
Max. number of controlable units | 10 units |
Max. number of controlable channels | 100 channels |
Settable items | AMP settings, Capture settings, Trigger/Alarm settings, Report settings, Other |
Captured data | Realtime data, Internal memory data, USB memory data (CSV and Binary data) |
Display | Analog waveforms, Logic waveforms, Pulse waveforms, Digital values |
File conversion | Data conversion between cursors/All data |
Monitor functions | Alarm monitor enables sending of email to the specified address |
Dual display function | Current/previous data display |
Statistics calculation, report | Maximum, minimum, average data during measurement |
Report function | Automatic creation of daily or monthly files |