We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies. Learn more.
The DPI 104 is a microprocessor-controlled digital pressure gauge that combines precision and functionality in a compact, robust and simple-to-use package. The DPI 104 matches advanced silicon sensor technology with several convenient design features resulting in an accurate, versatile yet affordable digital test gauge. Supplied as a standalone process indicator or in a kit with the widely proven hand pumps, the DPI 104 digital pressure gauge provides a reliable and economic solution for a wide range of pressure sensing applications.
The DPI 104 can be used as a stand-alone process indicator or as a pressure reference for use with our widely popular hand pumps.
Burst Pressure | Burst pressure is 2x working pressure (exception, 20,000 psi (1400 bar); burst pressure 29,000 psi (2000 bar). |
---|---|
Selectable Pressure Units | kPa, MPa, kg/cm2, psi, mbar, bar, inHg, inH2O, mH2O, mmH2O and mmHg |
Display | Pressure reading: 5 digits with 0.5 in (12.7 mm) character height Full scale indicator: 2.5 digits with 0.25 in (6.35 mm) character height |
Pressure Bar Graph | In addition to the numeric pressure indication, the DPI 104 LCD display contains a circular twenty segment bar graph to provide the user with a quick visual estimation of pressure from 0 to 100% full scale output. The bar graph increments represent 5% of the user-selectable range. |
Display Update | Two times per second |
Minimum/Maximum Values | The minimum and maximum pressure values can be displayed on the DPI 104. This function can be enabled/disabled, and reset by the user. |
Switch Test | The DPI 104 features a switch test function that will capture and display the open and closed values from an external pressure switch. Maximum switch impedance 200 O. |
Voltage Output | The DPI 104 can be programmed to provide a 0 to 5 V output signal that can be configured as directly proportional to the pressure shown in the display or set to a fixed value. The voltage output mode provides 0.1% FS accuracy from 50 mV to 5 V. |
Alarm Output | The alarm output consists of an open drain field effect transistor (maximum current 250 mA, maximum voltage 24 VDC) |
Adjustable Mounting Positions | For added convenience, the DPI 104 housing will rotate 320° around the pressure fitting and the faceplate can be rotated in any orientation for optimum visibility. |
Menu Lock | To guard against unauthorized menu access, the DPI 104 features a menu and tare lock function |
RS232 Interface | Serial communications are provided to allow transfer of data to a PC with the optional serial lead (IA4090-2- V0). Using this link all menu commands and display data are available via an ASCII command set or the optional SiCal Pro Software. |
Network Capability | Up to 99 DPI 104 instruments can be connected together in a daisy chain-configured network. |
Universal Pressure Module (UPM) Capability | UPM modules can be connected to the DPI 104 to enhance the accuracy and extend the pressure range. |
Battery | The DPI 104 is supplied with a 9V alkaline battery, type MN1604. Battery life is approximately six months when used daily for one hour per day. For increased performance a 9V lithium battery (not supplied by GE) is suggested. |
Pressure Port | 1/4 NPT or BSP male for units to 10,000 psi (700 bar) 9/16 x 18 UNF male cone connection 15,000 psi (1000 bar) and 20,000 psi (1400 bar) units |
Enclosure | Case material: ABS/PC plastic sealed to Type 4/IP65 Wetted parts: All stainless steel (316) or Inconel welded pressure cavity for compatibility with aggressive media. |
European Compliance | CE marked |
SiCalPro Software | This software package allows the user to control the DPI 104 remotely through a virtual instrument panel on the computer screen. The calibration data can be logged, viewed and printed in graphical format or as a calibration certificate. |
System Requirements | Minimum Intel Pentium with Windows® 95 or higher |
General | |
Storage Temperature | -4°F to 158°F (-20°C to 70°C) |
Dimensions | Diameter: 3.74 in (95 mm) excluding pressure fittings Depth: 2.17 in (55 mm) |
Weight | 13 oz (350 g) approximate |
Mechanical Vibration | To Def Stan 66-31, 8.4 Cat III |
Mechanical Shock | To BS EN 61010:2001 |
Electrical Environments | EMC: BS EN61326-1:1998 + A2:2001 Electrical Safety: BS EN 61010:2001 Mechanical Safety Pressure Equipment Directive_Class: Sound Engineering Practice (SEP) |
Options | DPI 104 external power supply unit with lead p/n 191-350 PC serial lead for connection to PC p/n IA4090-2-V0. (Note: Option A is recommended with PC serial lead.) SiCalPro software p/n 1S-SiCALPRO-DPI104 (includes IA4090-2-V0) Mating plug for eight-pin socket for switch test, voltage output, external power and alarm function (customers wishing to make their own connections/leads) p/n 1S-04-0027 9/16 x 18 UNF to 3/8 BSP adapter for connection to PV 212 p/n 182-190 DPI 104 UPM power lead for use with UPM IDOS Sensor p/n IA4101-1-V0. Requires Option H. Universal 12V external power supply for use with the UPM IDOS remote sensor p/n 191-129 |
Pressure Ranges
Range | Resolution | Maximum Working Pressure | ||||
psi | bar | type | psi | mbar | psi | bar |
0-10** | 0-0.7 | G* | 0.001 | 0.01 | 11.2 | 0.77 |
0-30 | 0-2 | G* or A | 0.001 | 0.1 | 32 | 2.2 |
0-100 | 0-7 | G* or A | 0.01 | 0.1 | 111.7 | 7.7 |
0-300 | 0-20 | G* or A | 0.01 | 1 | 319 | 22 |
0-1000 | 0-70 | G* or A | 0.1 | 1 | 1117 | 77 |
0-3000 | 0-200 | SG | 0.1 | 10 | 3190 | 220 |
0-5000 | 0-350 | SG | 0.1 | 10 | 5583 | 385 |
0-10,000 | 0-700 | SG | 1 | 10 | 11,165 | 770 |
0-15,000 | 0-1000 | SG | 1 | 100 | 15,950 | 1100 |
0-20,000 | 0-1400 | SG | 1 | 100 | 22,330 | 1540 |
*All gauge models will respond to negative pressures.
**0.15% FS for 10 psi (700 mbar) range.
- Process monitoring and control
- Test and calibration