Fluke has fused the power of a digital multimeter and a mA loop calibrator to produce their 787B ProcessMeter. The Fluke 787B allows users to multiply their troubleshooting abilities with a single compact device. Users benefit from the capabilities of the Fluke 87 DMM, with addition of the capacity for measuring, sourcing and simulating mA with the utmost precision and accuracy.
Fluke Connect® with ShareLive™ video call is the most comprehensive software and wireless test tool family that keeps you in contact with your entire team without leaving the field. The Fluke Connect family offers the largest system of connected test tools in the world. The software is compatible with the HTC One and One M8 running Android 4.4.x or higher, LG G3 and Nexus 5 running Android 4.4.x or higher, Samsung Galaxy S4 running Android 4.3.x or higher, Samsung Galaxy S5 running Android 4.4.x or higher. It also works with the iPhone models 4S and higher running iOS 8.0 or higher, iPad Air and iPad Mini (2nd generation) in an iPhone frame on iPad and iPod Touch (5th generation) and comes with over 30 different Fluke products. Please see our Fluke Connect category for the complete collection of Fluke Connect Products. Take me to Fluke Connect (Click Here).
Measurement function | Range and resolution | Best accuracy (% of reading + LSD) | |
V dc | 400.0 mV, 4.000 V, 40.00 V, 400.0 V, 1000 V | 0.1% + 1 | |
V ac (true-rms) | 400.0 mV, 4.000 V, 40.00 V, 400.0 V, 1000 V | 0.7% + 2 | |
mA dc | 30.000 mA | .05% + 2 | |
A dc | 1.000 A (0.440 A continuous) | 0.2% + 2 | |
A ac | 1.000 A (0.440 A continuous) | 1% + 2 | |
Resistance | 400.0 Ohms, 4.000 k, 40.00 k, 400.0 k, 4.0 M, 40 M | 0.2% + 1 | |
Frequency (0.5 Hz to 20 kHz) | 199.99 Hz, 1999.9 Hz, 19.999 kHz | .005% + 1 | |
Diode test | 2.000 V (shows diode voltage drop) | 2% + 1 | |
Continuity | Beeps for resistance < approx. 100 ohms | ||
Output function | Range and resolution | Drive capability | Accuracy (% of span) |
DC current output (Internal battery operation) | 0.000 to 20.000 mA or 4.000 to 20.000 mA, (selectable at power-up) Over-range to 24.000 mA | 24 V compliance or, 1,200 ohms, @ 20 mA | 0.05% |
DC current simulate (Ext. 15 V to 48 V loop supply) | 0.000 to 20.000 mA or 4.000 to 20.000 mA, (selectable at power-up) Over-range to 24.000 mA | 1000 ohms, @ 20 mA | 0.05% |
Current adjustment modes | Manual: Coarse, Fine, 25% and 100% step | ||
Automatic: Slow Ramp, Fast Ramp, 25% step | |||
Temperature range of 18 °C to 28 °C, for one year after calibration | |||
General Specifications | |||
Maximum voltage applied between any jack and earth ground | 1000 V RMS | ||
Storage temperature | -40 °C to 60 °C | ||
Operating temperature | -20 °C to 55 °C | ||
Temperature coefficient | 0.05 x (specified accuracy) per °C (for temperatures < 18 °C or > 28 °C) | ||
Relative humidity | 95% up to 30 °C; 75% up to 40 °C; 45% up to 50 °C; 35% up to 55 °C | ||
Vibration | Random, 2 g, 5-500 Hz | ||
Shock | 1 meter drop test | ||
Safety | IEC61010-1, Pollution Degree 2/IEC61010-2-033, CAT IV 600 V/CAT III 1000 V | ||
Size (HxWxL) | 50 x 100 x 203 mm (1.97 x 3.94 x 8.00 in) | ||
Weight | 600 g (1.3 lbs) | ||
Battery | Four AA alkaline batteries | ||
Battery life | 140 hours typical (measurement), 10 hours typical (sourcing 12 mA) | ||
Warranty | Three years |