The IDT 600 is an effective barrier for protecting Druck portable pressure calibrators against the ingress of dirt and moisture which may be present in the gaseous pressure media or pneumatic devices under test. Installed directly onto the standard calibrator pressure port, regular use of the IDT 600 will prevent contamination of the calibrator pneumatic system, improving it’s long term reliability and reducing maintenance costs.
In addition it prevents cross contamination of the devices under test. The IDT600-2 has no diaphragm or other internal components which could create a pressure drop and does not affect calibration accuracy. The unit is simple to dismantle and easy to clean. It should be periodically cleaned to prevent excessive contamination buildup which may overflow into the calibrator. The top and bottom sections are simply unscrewed, releasing the transparent chamber which is then wiped clean.
**Used with all Druck Portable Calibrator Products**
Pressure Connections | 1/8 NPT or BSP male thread for calibrator, 1/8 NPT or BSP female for the user connection. |
---|---|
Maximum Pressure | The system should not be pressurized above 500 psig (35 bar). |
Orientation | The IDT 600 is designed for use in the vertical position and should be removed from the calibrator during transit. |
Pressure Media | Gases compatible with stainless steel, acrylic and nitrile o-ring seals. |
Simple to use | The IDT 600 is fitted to the calibrator and the device under test is connected to the upper 1/8 NPT or BSP female port. Any dirt or moisture is then trapped inside the transparent chamber giving the operator an opportunity to clean the trap before contamination occurs. |