Define Instruments Zen Data Logger

SKU
ZEN-LOG
Manufacturer:
Define Instruments
Define Instruments

Key Features:

  • 16 universal isolated inputs
    • TC, RTD, mA, mV, V, frequency, & counter inputs
  • <±0.1% FSO accuracy (unless input states otherwise)
  • 4 digital inputs, 2 analog ouputs, 2 relay outputs
  • Log 7,858,683 samples for all channels to 1GB SD card
    • 31,774 samples to 32MB on-board memory
  • Optional Modbus for integration with SCADAs and PLCs
  • Requires a USB Bridge Key in order to program
  • CE approved & 5-year guarantee
Product Options:
$1,474.00

Secured and trusted checkout with

The Define Instruments Zen data logger is perfect for connecting to a combination of analog and digital sensors. The inputs are all individually isolated and EMC hardened, making it fit for the most demanding industrial environments. It can be set up as an analog multiplexer, utilizing the on-board dual analog outputs, or as an alarm and control station, using the 16 built-in setpoints. These setpoints can trigger up to 16 relay contacts if connected to the optional ZEN-RIO I/O expansion (sold separately).

A free download of the WorkBench software is available; use it in conjunction with the Zen data logger to plot graphs and visualize your data for a better understanding of your results. You can also export your data to CSV format for further scrutiny using Excel.

 

 

The WorkBench Software can be found here

Input

Number of input channels

16

Input configuration

Isolated universal inputs: Current, Voltage, RTD, Thermocouple, Potentiometer, Digital Pulse, Current Sensor or Attenuator input
See below for input specifications for each input type

Input isolation

2,500V AC for 1 minute between all input channels

Isolation test voltage

1000V DC for 1min between analog input channels, and between analog inputs and digital outputs

Resolution

16 bits

Accurate to

<±0.1% fso (unless otherwise stated for specific input dd>

 

Power

Power supply

HV= 85–265V AC / 95–370V DC, OR
MV= 24–48V AC / 17–72V DC

Supply frequency

50/60Hz, 10VA

Mains isolation

250V AC

Isolation test voltage

Mains isolation test voltage 3000V AC 50Hz for 1min to all inputs and outputs

 

General Specifications

Noise immunity (CMRR)

160dB tested at 300V RMS 50Hz

RF immunity

<±1% effect fso dd>

Permanent memory

(E²ROM) 100,000 writes per input parameter

Linearity & repeatability

<±0.1% dd>

Channel Separation

125db minimum

 

Thermocouple Input

Thermocouple types

B= 32 to 3272°F (0 to 1800°C)
E= -328 to 1292°F (-200 to 700°C)
J= -328 to 1832°F (-200 to 1000°C)
K= -328 to 2300°F (-200 to 1260°C)
N= -328 to 2372°F (-200 to 1300°C)
R= 32 to 3092°F (0 to 1700°C)
S= 32 to 3092°F (0 to 1700°C)
T= -328 to 752°F (-200 to 400°C)

Input impedance

>500KΩ

T/C lead resistance

100Ω max

Cold junction compensation

14 to 140°F (–10 to 60°C)

CJC drift

<0.02°c °c typical for all dd>

Accuracy

0.1% of FSO ±1°C typical

Sensor open

Upscale

 

RTD Input

RTD input type

Pt100 3 wire RTD DIN 43760: 1980
Pt1000 3 wire RTD standard

Range

-328 to 572°F (-200 to 300°C), 0.02°F (0.01°C) resolution
-328 to 1472°F (-200 to 800°C), 0.1°F (0.1°C) resolution

Lead wire resistance

10Ω/lead max recommended

Sensor current

0.6mA continuous

Sensor fail

Upscale

Accuracy

-328–572°F (-200–300°C) = ±0.1°C
-328–1472°F (-200–800°C) = ±0.3°C

Ambient drift

0.003°C/°C typical

 

Current Input

Range

0–20mA, 4–20mA

Input impedance

45Ω

Max over-range

Protected by PTC to 24V DC

Linearity & repeatability

0.1% FSO max

Accuracy

0.1% FSO max

Channel separation

0.001% max

Ambient drift

0.003%/°C FSO typical

RF immunity

1% effect FSO typical

Fail safe micro switch

Prevents external loops from being interrupted in the event of a power loss to the Zen

 

Voltage Input

Ranges

±200mV, –200mV to 1V, 0–10V, 0–18V

Input impedance

>500KΩ on all ranges

Maximum over voltage

24V DC

Linearity & repeatability

0.1% FSO max

Accuracy

0.1% FSO max

Channel separation

0.001% max

Ambient drift

0.003%/°C FSO typical

RF immunity

1% effect FSO typical

 

Digital Pulse Input

Frequency range

0–2500.0Hz

Fast counter range

0–2500.0Hz

Sensors

Open collector (NPN, PNP), TTL or Clean Contact

Frequency resolution

0.1Hz

Debounce counter range

0–50Hz max

Counter register output

32 bit

Accuracy

±0.5%

 

AC Current Sensor Input

Sensor type

Current transformer ACCS-420, ACCS-420-L and ACCS-010

Amperage range Header selectable

ACCS-420/010= 100/150/200A
ACCS-420-L= 10/20/50A

Overload (continuous)

ACCS-420/010= 175/300/400A respectively
ACCS-420-L = 80/120/200A respectively

Output (Representing 0–100% of full scale input range)

ACCS-420(-L)= 4–20mA DC loop powered
ACCS-010= 0–10V DC

Power supply

ACCS-420(-L)= Loop powered, 15–36V DC
ACCS-010= Self powered

Accuracy

1% of full scale

Response time

250ms (10–90%)

Isolation voltage

2,000V

Frequency

50–60Hz

 

Features

Expansion interface

Add 16 relay outputs and 16 digital control inputs with a ZEN-RIO Relay/IO Expansion (sold separately)

HMI interface

For FM1602 Remote Display (sold separately)

 

Potentiometer Input

Potentiometer input

3-wire

Excitation voltage

Variable

Potentiometer resistance

<2kω low pot;>2kΩ high pot

Field programmable zero

0–90% of span

Field programmable span

0.1–100%

Linearity & repeatability

<±0.05% fso dd>

Response time

100msec

Ambient drift

<50ppm dd>

 

Attenuator Input

Attenuator type

Define Instruments HVA-1000

Max input voltage

1000V DC

Attenuation factor

1000 ±0.1%

Input impedance

3.8MΩ

Output impedance

3.8kΩ

Attenuator type

Differential resistive

Ambient drift

50ppm/°C max

 

Relay Output

Number of relay outputs

2

LED indication

of each output

Isolation

Isolated

Contact rating

5A 30V DC

 

Analog Output (Retransmission)

Number of analog outputs

2

Analog output type

Isolated 4-20mA analog outputs

Isolation test voltage

1000V DC for 1min between analog output and digital output

Resolution

15 bits, 16,000 steps

Compliance voltage

10V 500Ω

 

Intelligent Multiplexer

Channel selection

Binary (digital inputs)

Cycle time

4sec minimum to cycle through all 16 inputs

Resolution

12 bits, 4000 steps typical

 

Comm Port(s)

Protocols

Modbus RTU, RS232, RS422 or RS485

Port 1

Select either:
RS= RS422 / RS485 OR
EMOD= Ethernet Modbus (10/100/Auto)

Port 2

RS232 / RS485 auto-select
Selectable baud rate 2400–230000 baud. Format: 8 bit, no parity, 1 stop

Isolation test voltage

1000V DC for 1min between comms and analog input/output, and between comms and digital input/output

 

Digital Inputs

Number of digital inputs

4 x Opto isolated inputs

LED indication

of each digital input

Functions

Status, up counter, up/down counter with direction, debounced counter, frequency, gated frequency

Counter register output

32 bit

Frequency range

0–10,000Hz

Input voltage

5–30V DC

Threshold

4.6V typical

Debounce counter range

0–100Hz

Load

At 5V DC= 1.1mA; At 24V DC= 7mA

Isolation test voltage

1000V DC for 1min between digital input and analog input/output, and between digital input and digital output

 

Data Logging

Real-time clock

Data logging

to Micro SD card (supplied). Logging to onboard memory when Micro SD card removed

Micro SD card capacity

1GB (7,858,683 samples for all channels)
Not compatible with higher-capacity Micro SD cards

Onboard memory capacity

32MB (31,774 samples for all channels)
Used when Micro SD card is removed

Datalog retrieval

Fast, simple datalog retrieval and visualization using Define WorkBench

 

Programming

USB programmable

Via USB prog port using Bridge Key USB programmer (sold separately)

Define WorkBench

Simple configuration and data log retrieval using Define WorkBench

 

Construction

Casing

DIN 35 rail mounting; Material: ABS inflammability V0 (UL94)

Dimensions (H x W x D)

2.32 x 10.04 x 5.67" (59 x 255 x 144mm), with plugs in

Weight

ZEN-LOG-RS = 27.2oz (770g)
ZEN-LOG-EMOD = 27.7oz (784g)

 

Environmental Conditions

Operating temperature

14 to 140°F (–10 to 60°C)

Storage temperature

–4 to 176°F (–20 to 80°C)

Operating humidity

5–85% RH max, non-condensing

 

Certifications & Compliances

EN 61326-1

Complies with EN-61326-1:2006

EMC

Emissions EN 558022-A; Immunity EN 50082-1; Safety EN 60950

Write Your Own Review
Only registered users can write reviews. Please Sign in or create an account

Products In This Series

As low as

Now $195.00

As low as

Now $604.00

As low as